AnalogLevelPauseTrigger Properties

The AnalogLevelPauseTrigger type exposes the following members.

Properties

  NameDescription
Public propertyCondition
Specifies whether the task pauses above or below the threshold you specify with Level.
Public propertyCoupling
Specifies the coupling for the source signal of the trigger if the source is a terminal rather than a virtual channel.
Public propertyDigitalFilterEnable
Specifies whether to apply a digital filter to the digital output of the analog triggering circuitry (the Analog Comparison Event). When enabled, the analog signal must stay above or below the trigger level for the minimum pulse width before being recognized. Use filtering for noisy trigger signals that transition in and out of the hysteresis window rapidly.
Public propertyDigitalFilterMinimumPulseWidth
Specifies in seconds the minimum pulse width the filter recognizes.
Public propertyDigitalFilterTimebaseRate
Specifies in hertz the rate of the digital filter timebase. NI-DAQmx uses this value to compute settings for the filter.
Public propertyDigitalFilterTimebaseSource
Specifies the terminal of the signal to use as the timebase of the digital filter.
Public propertyDigitalSynchronizationEnable
Specifies whether to synchronize recognition of transitions in the signal to the internal timebase of the device.
Public propertyHysteresis
Specifies a hysteresis level in the units of the measurement or generation. If Condition is AboveLevel, the trigger does not deassert until the source signal passes below Level minus the hysteresis. If Condition is BelowLevel, the trigger does not deassert until the source signal passes above Level plus the hysteresis. Hysteresis is always enabled. Set this property to a non-zero value to use hysteresis.
Public propertyLevel
Specifies the threshold at which to pause the task. Specify this value in the units of the measurement or generation. Use Condition to specify whether the task pauses above or below this threshold.
Public propertySource
Specifies the name of a virtual channel or terminal where there is an analog signal to use as the source of the trigger.