NI-DAQmx .NET Framework 4.0 API Reference

Table of Contents

ExternalCalibrationSession.Setup9266 Method

  • Updated2023-02-21
  • 1 minute(s) read
ExternalCalibrationSession.Setup9266 Method
Applies the specified DAC value to the AO channels, such that it can then be measured by a DMM

Namespace:  NationalInstruments.DAQmx
Assembly:  NationalInstruments.DAQmx (in NationalInstruments.DAQmx.dll) Version: 20.7.40.49302

Syntax

public void Setup9266(
	string physicalChannels,
	int value
)
Public Sub Setup9266 ( 
	physicalChannels As String,
	value As Integer
)

Parameters

physicalChannels
Type: SystemString
Specifies the name of the physical channel to use to create the virtual channel. The DAQmx physical channel constant lists all physical channels for devices installed in the system. You cannot specify more than one physical channel.
value
Type: SystemInt32
The binary number to write to the DAC.

Exceptions

ExceptionCondition
DaqExceptionThe NI-DAQmx driver returned an error.

Remarks

Refer to the calibration procedure for your device for detailed calibration instructions.
Note Note
After you commit the calibration, you cannot undo it. If any calibration settings are incorrect, you must recalibrate the device.

See Also