ExternalCalibrationSession.SetupDsaAOTimebase Method

  • Updated2023-02-21
  • 1 minute(s) read
ExternalCalibrationSession.SetupDsaAOTimebase Method
Generates a sine wave of a fixed frequency from the DSA device. You must measure the frequency of the generated sine wave and pass that value into AdjustDsaAOTimebase(Double, Boolean).

Namespace:  NationalInstruments.DAQmx
Assembly:  NationalInstruments.DAQmx (in NationalInstruments.DAQmx.dll) Version: 20.7.40.49302

Syntax

public double SetupDsaAOTimebase()
Public Function SetupDsaAOTimebase As Double

Return Value

Type: Double
The expected frequency of the sine wave of a calibrated device.

Exceptions

ExceptionCondition
DaqExceptionThe NI-DAQmx driver returned an error.

Remarks

Refer to the calibration procedure for your device for detailed calibration instructions.
Note Note
After you commit the calibration, you cannot undo it. If any calibration settings are incorrect, you must recalibrate the device.