ExternalCalibrationSession Methods
- Updated2023-02-21
- 25 minute(s) read
The ExternalCalibrationSession type exposes the following members.
Methods
| Name | Description | |
|---|---|---|
| Adjust1102 | Adjusts the external calibration constants for an SCXI-1102 module. | |
| Adjust1104 | Adjusts the external calibration constants for an SCXI-1104 module. | |
| Adjust1112 | Adjusts the external calibration constants for an SCXI-1112 module. | |
| Adjust1122 | Adjusts the external calibration constants for an SCXI-1122 module. | |
| Adjust1124 | Adjusts the external calibration constants for an SCXI-1124 module. | |
| Adjust1125 | Adjusts the external calibration constants for an SCXI-1125 module. | |
| Adjust1126 | Adjusts the external calibration constants for an SCXI-1126 module. | |
| Adjust1141 | Adjusts the external calibration constants for an SCXI-1141 module. | |
| Adjust1142 | Adjusts the external calibration constants for an SCXI-1142 module. | |
| Adjust1143 | Adjusts the external calibration constants for an SCXI-1143 module. | |
| Adjust11601 | Adjusts theexternal calibrationconstants for an NI FD-11601 device bank. | |
| Adjust11603 | Adjusts theexternal calibrationconstants for an NI FD-11603 device bank. | |
| Adjust11605 | Adjusts theexternal calibrationconstants for an NI FD-11605 device bank. | |
| Adjust11613 | Adjusts theexternal calibrationconstants for an NI FD-11613 device bank. | |
| Adjust11614 | Adjusts theexternal calibrationconstants for an NI FD-11614 device bank. | |
| Adjust11634 | Adjusts theexternal calibrationconstants for an NI FD-11634 device bank. | |
| Adjust11637 | Adjusts theexternal calibrationconstants for an NI FD-11637 device bank. | |
| Adjust1502 | Adjusts the external calibration constants for an SCXI-1502 module. | |
| Adjust1503 | Adjusts the external calibration constants for an SCXI-1503 module. | |
| Adjust1503Current | Adjusts the external calibration constants for an SCXI-1503 module. | |
| Adjust1520 | Adjusts the external calibration constants for an SCXI-1520 module. | |
| Adjust1521 | Adjusts the external calibration constants for an SCXI-1521 module. | |
| Adjust153x | Adjusts the external calibration constants for an SCXI-1530 or SCXI-1531 module. | |
| Adjust1540 | Adjusts the external calibration constants for an SCXI-1540 module. | |
| Adjust4204 | Adjusts the external calibration constants for an NI 4204 device. | |
| Adjust4220 | Adjusts the external calibration constants for an NI 4220 device. | |
| Adjust4224 | Adjusts the external calibration constants for an NI 4224 device. | |
| Adjust4225 | Obsolete. The Adjust4225(String, Double, Double) method has been deprecated. Adjusts the external calibration constants for an NI 4225 device. | |
| Adjust4300 |
Adjusts the
external calibration
constants on an NI 4300 device.
| |
| Adjust4302 |
Adjusts the external calibration constants for an NI 4302. Connect a known voltage to the device and specify that voltage with referenceVoltage.
| |
| Adjust4303 |
Adjusts the external calibration constants for an NI 4303. Connect a known voltage to the device and specify that voltage with referenceVoltage.
| |
| Adjust4304 |
Adjusts the external calibration constants for an NI 4304. Connect a known voltage to the device and specify that voltage with referenceVoltage.
| |
| Adjust4305 |
Adjusts the external calibration constants for an NI 4305. Connect a known voltage to the device and specify that voltage with referenceVoltage.
| |
| Adjust4309 | Adjusts the calibration constants for a PXIe-4309 device. | |
| Adjust4310 | Adjusts the calibration constants for a PXIe-4310 device. | |
| Adjust4322 |
Adjusts the
external calibration
constants on an NI 4322 device.
| |
| Adjust4339 |
Adjusts the external calibration constants for an NI PXIe 4339 device.
| |
| Adjust433x |
Adjusts the
external calibration
constants on an NI PXIe-433x device.
| |
| Adjust4353 |
Adjusts the
external calibration
constants on an NI 4353 device.
| |
| Adjust4357 | Adjusts the external calibration constants for an NI 4357 device.
| |
| Adjust9201 | Adjusts the external calibration constants for an NI 9201 device. | |
| Adjust9202 | Adjusts the external calibration constants for an NI 9202 device. | |
| Adjust9203Gain | Adjusts the external calibration gain constants for an NI 9203 device. | |
| Adjust9203Offset | Adjusts the external calibration offset constants for an NI 9203 device. | |
| Adjust9205 | Adjusts the external calibration constants for an NI 9205 device. | |
| Adjust9206 | Adjusts the external calibration constants for an NI 9206 device. | |
| Adjust9207Gain | Adjusts the external calibration gain constants for an NI 9207 device. | |
| Adjust9207Offset | Adjusts the external calibration offset constants for an NI 9207 device. | |
| Adjust9208Gain | Adjusts the external calibration gain constants for an NI 9208 device. | |
| Adjust9208Offset | Adjusts the external calibration offset constants for an NI 9208 device. | |
| Adjust9209Gain |
Adjusts the external calibration gain constants for an NI 9209 device.
| |
| Adjust9209Offset |
Adjusts the external calibration offset constants for an NI 9209 device.
| |
| Adjust9210 | Adjusts the external calibration constants for an NI 9210 device. | |
| Adjust9211 | Adjusts the external calibration constants for an NI 9211 device. | |
| Adjust9212 | Adjusts the external calibration constants for an NI 9212 device. | |
| Adjust9213 | Adjusts the external calibration constants for an NI 9213 device. | |
| Adjust9214 | Adjusts the external calibration constants for an NI 9214 device. | |
| Adjust9215 | Adjusts the external calibration constants for an NI 9215 device. | |
| Adjust9216 |
Adjusts the external calibration constants for an NI 9216 device.
| |
| Adjust9217 | Adjusts the external calibration constants for an NI 9217 device. | |
| Adjust9218 | Adjusts the external calibration constants for an NI 9218 device. | |
| Adjust9219 | Adjusts the external calibration constants for an NI 9219 device. | |
| Adjust9220 |
Adjusts the
external calibration
constants for an NI 9220 device.
| |
| Adjust9221 | Adjusts the external calibration constants for an NI 9221 device. | |
| Adjust9222 | Adjusts the external calibration constants for an NI 9222 device. | |
| Adjust9223 | Adjusts the external calibration constants for an NI 9223 device. | |
| Adjust9224 | Adjusts the external calibration constants for an NI 9224 device. | |
| Adjust9225 | Adjusts the external calibration constants for an NI 9225 device. | |
| Adjust9226 | Adjusts the external calibration constants for an NI 9226 device. | |
| Adjust9227 | Adjusts the external calibration constants for an NI 9227 device. | |
| Adjust9228 | Adjusts the external calibration constants for an NI 9228 device. | |
| Adjust9229 | Adjusts the external calibration constants for an NI 9229 device. | |
| Adjust9230 | Adjusts the external calibration constants for an NI 9230 device. | |
| Adjust9231 | Adjusts the external calibration constants for an NI 9231 device. | |
| Adjust9232 | Adjusts the external calibration constants for an NI 9232 device.
| |
| Adjust9234Gain | Adjusts the external calibration gain constants for an NI 9234 device. | |
| Adjust9234Offset | Adjusts the external calibration offset constants for an NI 9234 device. | |
| Adjust9238 | Adjusts the external calibration constants for an NI 9238 device. | |
| Adjust9239 | Adjusts the external calibration constants for an NI 9239 device. | |
| Adjust9242 | Adjusts the external calibration constants for an NI 9242 device. | |
| Adjust9244 | Adjusts the external calibration constants for an NI 9244 device. | |
| Adjust9246 | Adjusts the external calibration constants for an NI 9246 device. | |
| Adjust9247 | Adjusts the external calibration constants for an NI 9247 device. | |
| Adjust9250 | Adjusts the external calibration constants for an NI 9250 device. | |
| Adjust9251 | Adjusts the external calibration constants for an NI 9251 device. | |
| Adjust9252 | Adjusts the external calibration constants for an NI 9252 device. | |
| Adjust9253 | Adjusts the external calibration constants for an NI 9253 device. | |
| Adjust9260 | Adjusts the external calibration constants for an NI 9260 device. | |
| Adjust9262 | Adjusts the external calibration constants for an NI 9262 device. | |
| Adjust9263 | Adjusts the external calibration constants for an NI 9263 device. | |
| Adjust9264 | Adjusts the external calibration constants for an NI 9264 device. | |
| Adjust9265 | Adjusts the external calibration constants for an NI 9265 device. | |
| Adjust9266 | Adjusts the external calibration constants for an NI 9266 device. | |
| Adjust9269 | Adjusts the external calibration constants for an NI 9269 device. | |
| Adjust9775 | Adjusts the external calibration constants for an NI 9775 device. | |
| AdjustAOSeries | Adjusts the external calibration constants on an AO Series device. | |
| AdjustDsa4463 | Adjusts the external calibration constants for the analog output section of a NI 4463 device. | |
| AdjustDsa4610 | Adjusts the external calibration gain constants for an NI 4610. | |
| AdjustDsaAI(Double) | Adjusts the external calibration constants for the analog input section of a DSA device without shorting the input terminals. | |
| AdjustDsaAI(Double, Boolean) | Adjusts the external calibration constants for the analog input section of a DSA device. You can specify whether you want to short the input terminals or not. | |
| AdjustDsaAI(AICoupling, Double, Double) | Performs external calibration adjustment on a DSA device with the specified gain and coupling configuration. | |
| AdjustDsaAO | Adjusts the external calibration constants for the analog output section of a DSA device. | |
| AdjustDsaAOTimebase | Adjusts the external calibration constants for the timebase of a DSA device with an adjustible oscillator. | |
| AdjustDsaTimebase | Adjusts the external calibration constants for the timebase of a DSA device with an adjustible oscillator. | |
| AdjustESeries | Adjusts the external calibration constants on an E Series device. | |
| AdjustMSeries | Adjusts the external calibration constants for an M Series device. | |
| AdjustSCBaseboard | Adjusts the external calibration constants for the baseboard of an SC Series device. | |
| AdjustSSeries | Adjusts the external calibration constants for an S Series device. | |
| AdjustTioTimebase | Adjusts the external calibration constants for the timebase of a TIO device. | |
| AdjustXSeries | Adjusts the external calibration constants for an X Series device. | |
| ConnectSCExpressCalibrationAccessoryChannels | Configures a connection on the SC Express accessory for the specified physical channel(s).
| |
| CreateObjRef | Creates an object that contains all the relevant information required to generate a proxy used to communicate with a remote object. (Inherited from MarshalByRefObject.) | |
| CSeriesSetCalibrationTemperature | Specifies the temperature of a C Series device for the current external calibration session. | |
| DisconnectSCExpressCalibrationAccessoryChannels | Disconnects the configured connection on an SC Express accessory and connects the accessory in the default configuration.
| |
| Dispose | Releases all resources used by ExternalCalibrationSession. | |
| Dispose(Boolean) |
Releases the managed and unmanaged resources used by ExternalCalibrationSession or optionally releases only the unmanaged resources.
| |
| DsaSetCalibrationTemperature | Specifies the temperature of a DSA device for the current external calibration session.
| |
| Equals | Determines whether the specified object is equal to the current object. (Inherited from Object.) | |
| FieldDAQSetCalibrationTemperature | Specifies the temperature of a FieldDAQ device for the current external calibration session. | |
| Get11601AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get11603AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get11605AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get11613AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get11614AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get11634AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get4302AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get4303AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get4304AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get4305AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get4322AdjustmentPoints | Returns the reference values to be used by a reference device to create a reference signal. | |
| Get4339AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get9201AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get9202AdjustmentPoints |
Returns the reference voltage values to be used by a reference device to create a reference signal.
| |
| Get9203AdjustmentPoints | Returns the reference current values to be used by a reference device to create a reference signal. | |
| Get9207AdjustmentPoints | Returns the reference voltage or current values to be used by a reference device to create a reference signal. | |
| Get9208AdjustmentPoints | Returns the reference current values to be used by a reference device to create a reference signal. | |
| Get9209AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get9212AdjustmentPoints | Returns the reference values to be used by a reference device to create a reference signal. | |
| Get9213AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get9214AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get9215AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get9216AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get9217AdjustmentPoints | Returns the reference values to be used by a reference device to create a reference signal. | |
| Get9218AdjustmentPoints | Returns the reference values to be used by a reference device to create a reference signal. | |
| Get9219AdjustmentPoints | Returns the reference values to be used by a reference device to create a reference signal. | |
| Get9220AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get9221AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get9222AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get9223AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get9224AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get9225AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get9226AdjustmentPoints | Returns the reference values to be used by a reference device to create a reference signal. | |
| Get9227AdjustmentPoints | Returns the reference current values to be used by a reference device to create a reference signal. | |
| Get9228AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get9229AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get9230AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get9231AdjustmentPoints |
Returns the reference voltage values to be used by a reference device to create a reference signal.
| |
| Get9232AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get9234AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get9238AdjustmentPoints | Returns the reference values to be used by a reference device to create a reference signal. | |
| Get9239AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get9242AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get9244AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get9246AdjustmentPoints | Returns the reference values to be used by a reference device to create a reference signal. | |
| Get9247AdjustmentPoints | Returns the reference values to be used by a reference device to create a reference signal. | |
| Get9250AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get9251AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get9252AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get9253AdjustmentPoints | Returns the reference values to be used by a reference device to create a reference signal. | |
| Get9260AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| Get9262AdjustmentPoints | Returns the reference voltage values to be measured by a reference device. | |
| Get9263AdjustmentPoints | Returns the reference voltage values to be measured by a reference device. | |
| Get9264AdjustmentPoints | Returns the reference voltage values to be measured by a reference device. | |
| Get9265AdjustmentPoints | Returns the reference current values to be measured by a reference device. | |
| Get9266AdjustmentPoints | Returns the reference current values to be measured by a reference device. | |
| Get9269AdjustmentPoints | Returns the reference voltage values to be measured by a reference device. | |
| Get9775AdjustmentPoints | Returns the reference values to be used by a reference device to create a reference signal. | |
| GetDsa4463AdjustmentPoints | Returns the reference voltage values to be used by a reference device to create a reference signal. | |
| GetHashCode | Serves as the default hash function. (Inherited from Object.) | |
| GetLifetimeService | Retrieves the current lifetime service object that controls the lifetime policy for this instance. (Inherited from MarshalByRefObject.) | |
| GetType | Gets the Type of the current instance. (Inherited from Object.) | |
| InitializeLifetimeService | Obtains a lifetime service object to control the lifetime policy for this instance. (Inherited from MarshalByRefObject.) | |
| MemberwiseClone | Creates a shallow copy of the current Object. (Inherited from Object.) | |
| MemberwiseClone(Boolean) | Creates a shallow copy of the current MarshalByRefObject object. (Inherited from MarshalByRefObject.) | |
| SaveChangesAndDispose | Closes an open external calibration session and saves the calibration changes made in the session. | |
| SetSCExpressCalibrationAccessoryBridgeOutput | Configures an SC Express accessory to produce the bridge output specified in volts-per-volt.
| |
| Setup1102 | Configures the calibration task for an SCXI-1102 module. | |
| Setup1104 | Configures the calibration task for an SCXI-1104 module. | |
| Setup1112 | Configures the calibration task for an SCXI-1112 module. | |
| Setup1122 | Configures the calibration task for an SCXI-1122 module. | |
| Setup1124(String, Scxi1124Range, Int64) | Configures the calibration task for an SCXI-1124 module. | |
| Setup1124(String, Scxi1124Range, UInt32) | Configures the calibration task for an SCXI-1124 module. | |
| Setup1125 | Configures the calibration task for an SCXI-1125 module. | |
| Setup1126 | Configures the channel and range for an SCXI-1126 module calibration task. | |
| Setup1141 | Configures the gain value for an SCXI-1141 module calibration task. | |
| Setup1142 | Configures the gain value for an SCXI-1142 module calibration task. | |
| Setup1143 | Configures the gain value for an SCXI-1143 module calibration task. | |
| Setup11605 | Sets up external calibration for an FD 11605 device. | |
| Setup11634 | Sets up external calibration for an FD 11634 device. | |
| Setup11637 | Sets up external calibration for an FD 11637 device. | |
| Setup1502 | Configures the calibration task for an SCXI-1502 module. | |
| Setup1503 | Configures the calibration task for an SCXI-1503 module. | |
| Setup1520 | Configures the calibration task for an SCXI-1520 module. | |
| Setup1521 | Configures the calibration task for an SCXI-1521 module. | |
| Setup153x | Configures the gain value for an SCXI-1530 or SCXI-1531 module calibration task. | |
| Setup1540 | Configures the gain value for an SCXI-1540 module calibration task. | |
| Setup4302 |
Initializes an NI PXIe-4302 device with the specified range for calibration.
| |
| Setup4303 |
Initializes an NI PXIe-4303 device with the specified range for calibration.
| |
| Setup4304 |
Initializes an NI PXIe-4304 device with the specified range for calibration.
| |
| Setup4305 |
Initializes an NI PXIe-4305 device with the specified range for calibration.
| |
| Setup4322 | Configures the calibration task for an NI 4322 device. | |
| Setup4339 |
Initializes an NI PXIe 4339 device with the specified mode, range, and exitation for calibration.
| |
| Setup433x | Configures the calibration task for an NI PXIe-433x device. | |
| Setup9218 | Sets up external calibration for an NI 9218 device. | |
| Setup9219 | Sets up external calibration for an NI 9219 device. | |
| Setup9242 | Reads the specified value from the ADC on the specified physical channel. This method requires a sweep of DAC values obtained from Get9242AdjustmentPoints. | |
| Setup9244 | Reads the specified value from the ADC on the specified physical channel. This method requires a sweep of DAC values obtained from Get9244AdjustmentPoints. | |
| Setup9260 |
Writes the specified binary value to the DAC on the specified physical channel. This modules requires a sweep of DAC values obtained from the Get9260AdjustmentPoints.
| |
| Setup9262 | Writes the specified binary value to the DAC on the specified physical channel. This method requires a sweep of DAC values obtained from Get9262AdjustmentPoints. | |
| Setup9263 | Writes the specified binary value to the DAC on the specified physical channel. This method requires a sweep of DAC values obtained from Get9263AdjustmentPoints. | |
| Setup9264 | Writes the specified binary value to the DAC on the specified physical channel. This method requires a sweep of DAC values obtained from Get9264AdjustmentPoints. | |
| Setup9265 | Writes the specified binary value to the DAC on the specified physical channel. This method requires a sweep of DAC values obtained from Get9265AdjustmentPoints. | |
| Setup9266 | Applies the specified DAC value to the AO channels, such that it can then be measured by a DMM | |
| Setup9269 | Writes the specified binary value to the DAC on the specified physical channel. This method requires a sweep of DAC values obtained from Get9269AdjustmentPoints. | |
| SetupDsa4463 |
Generates a specified voltage from the DSA device. | |
| SetupDsa4480 |
Generates a specified voltage from the DSA device. | |
| SetupDsaAOTimebase | Generates a sine wave of a fixed frequency from the DSA device. You must measure the frequency of the generated sine wave and pass that value into AdjustDsaAOTimebase(Double, Boolean). | |
| ToString | Returns a string representation of the object. (Overrides ObjectToString.) |