Configures the function to generate on the specified channel(s).

When DC Voltage is selected, the device generates the desired voltage level on the output as long as the output current is below the current limit. Use the following properties to configure the channel when DC Voltage is selected:

When DC Current is selected, the device generates the desired current level on the output as long as the output voltage is below the voltage limit. Use the following properties to configure the channel when DC Current is selected:

When Pulse Voltage is selected, the device generates pulses at the desired pulse voltage levels on the output as long as the output current is below the pulse current limit. Use the following properties to configure the channel when Pulse Voltage is selected:

When Pulse Current is selected, the device generates pulses at the desired pulse current levels on the output as long as the output voltage is below the pulse voltage limit. Use the following properties to configure the channel when Pulse Current is selected:

When Constant Resistance is selected, the device generates the desired resistance level on the output as long as the output current is below the current limit. Use the following properties to configure the channel when Constant Resistance is selected:

When Constant Power is selected, the device generates the desired power level on the output as long as the output current is below the current limit. Use the following properties to configure the channel when Constant Power is selected:

Note NI-DCPower uses the terms "source" and "output". However, while sinking with electronic loads and SMUs these correspond to "sinking" and "input", respectively.

Default Value : Refer to Supported Properties by Device for the default value by device.

Remarks

The following table lists the characteristics of this property.

Short Name Output Function
Data type ci32.png
Permissions Read/Write
High-level VIs niDCPower Configure Output Function
Channel-based Yes
Instrument-based No
Resettable No
DC Voltage 1006

Sets the output function to DC voltage.

DC Current 1007

Sets the output function to DC current.

Pulse Voltage 1049

Sets the output function to pulse voltage.

Pulse Current 1050

Sets the output function to pulse current.

Constant Resistance 1161

Sets the output function to constant resistance.

Constant Power 1162

Sets the output function to constant power.