Specifies the minimum voltage, in volts, that the output can produce when generating the desired pulse current on the specified channel(s) during the on phase of a pulse.

This property is applicable only if the Compliance Limit Symmetry property is set to Asymmetric and the Output Function property is set to Pulse Current .

You must also specify a Pulse Voltage Limit High to complete the asymmetric range.

Note The channel must be enabled and the output relay must be connected for the specified limit to take effect. Refer to the Output Enabled property and the Output Connected property for more information about enabling the channel and connecting the output relay respectively.
Note NI-DCPower uses the terms "source" and "output". However, while sinking with electronic loads and SMUs these correspond to "sinking" and "input", respectively.
Note Refer to Supported Properties by Device for information about supported devices.

Valid Values: [- Pulse Voltage Limit Range , -1% of Pulse Voltage Limit Range ]

  • The range bounded by the limit high and limit low must include zero.
Note The limit may be extended beyond the selected limit range if the Overranging Enabled property is set to TRUE or if the Output Function property is set to a pulsing function.

Default Value: Refer to Supported Properties by Device for the default value by device.

Remarks

The following table lists the characteristics of this property.

Short Name Pulse Voltage Limit Low
Data type cdbl.png
Permissions Read/Write
High-level VIs N/A
Channel-based Yes
Instrument-based No
Resettable No