Running a DCIR Test
- Updated2025-10-15
- 1 minute(s) read
To determine the DC internal resistance (DCIR) of a cell, run a DCIR test using the DCIR Test.vi.
Equipment
- PXIe-4051
-
Use the NI Example Finder to open DCIR
Test.vi.
-
In the LabVIEW window, select .
The NI Example Finder window loads.
-
Select .
The front panel of the VI loads.
-
In the LabVIEW window, select .
- On the front panel of the VI, select the electronic load resource from the HW Resource Name control.
- Optional:
To perform an automatic self-calibration, enable
Self-Calibration?
Note Before allowing a DUT self-calibration, NI recommends disconnecting the electronic load. Self-calibrating a connected DUT may damage your equipment or lead to inaccurate data.
- Set Temperature Delta to trigger self-calibration at a specified temperature differential value.
- Set Days to Calibration to trigger self-calibration after a specified number of days since the last calibration.
-
Enter the DCIR Test Parameters.
- Set the Maximum Load Current (A) for the test.
- Set the Power Line Frequency (Hz) for the test environment.
- Click Run.