Replacing an EMO Button
- Updated2025-12-16
- 2 minute(s) read
- Power off your ATE Core Configurations Generation 2 system by flipping the main disconnect branch breaker to off.
- If your ATE Core Configurations Generation 2 system utilizes a UPS to provide backup power, press the power button on the UPS to power it down.
- Disconnect your ATE Core Configurations Generation 2 system from facility power.
| Shipping Kit Part Number | Description |
|---|---|
| 789091-01 | EMO Button |
The diagram in this procedure illustrates the ATE-38U/44U front I/O panel. If you are replacing an EMO button on the rear I/O panel or on the ATE-15U, the same procedure applies.
Equipment
- 3 mm hex key
- Remove the ATE Core Configurations Generation 2 left side panel using the 3 mm hex key.
- Locate the EMO button that you are replacing.
-
Disconnect the EMO button cable from the I/O signal cable.c
- If you are replacing the EMO button on the front I/O panel, disconnect the EMO button cable from the front I/O signal cable.
- If you are replacing the EMO button on the rear I/O panel, disconnect the EMO button cable from the rear I/O signal cable.
- Connect the new EMO button to the EMO button shroud and install the assembly in the I/O panel.
- Connect the new EMO button cable to the I/O signal cable.
- EMO button cable
- I/O signal cable
- EMO button
- EMO button shroud
- Reinstall the ATE Core Configurations Generation 2 left side panel using the 3 mm hex key.
Once installation of the new EMO button is complete, you can reconnect facility power to your ATE Core Configurations Generation 2 system and power on the rack.