SET-1240 Specifications
- Updated2026-05-04
- 7 minute(s) read
SET-1240 Specifications
These specifications apply to the SET-1240.
Revision History
| Version | Date changed | Description |
|---|---|---|
| 379273A-01 | April 2026 | Initial release. |
Looking For Something Else?
For information not found in the specifications for your product, such as operating instructions, browse Related Information.
Definitions
Warranted Specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
Values are Typical unless otherwise noted.
Conditions
- All voltages are relative to GND
SET-1240 Pinouts
Access detailed pinout information for connecting to J1, J2, XJ2, and XJ3 connectors on the SET-1240.
J1 and J2 Pinout
J1 and J2 are female 44-pin high-density D-SUB connectors with a 4-40 UNC screw lock.
| Signal | Description (6-Wire Mode) | Description (4/5-Wire Mode) |
|---|---|---|
| CHx_A+ | Emulated A+ signal output for Channel x | Emulated + signal output for Channel x |
| CHx_A- | Emulated A- signal output for Channel x | Common line. Shorted to Chx_B+ |
| CHx_B+ | Emulated B+ signal output for Channel x | Common line. Shorted to Chx_A- |
| CHx_B- | Emulated B- signal output for Channel x | Emulated - signal output for Channel x |
| CHx_EXC+/- | Excitation input | Excitation input |
| NC | No connection | No connection |
XJ2 Pinout
XJ2 is a 110-pin Hard Metric type A connector, per IEC 61076-101. XJ2 is in the standard position with pinout for fully compatible rear I/O.
| Row | E | D | C | B | A |
|---|---|---|---|---|---|
| 1 | CH1_CTRL_B | CH1_CTRL_A | NC | CH0_CTRL_B | CH0_CTRL_A |
| 2 | CH3_CTRL_B | CH3_CTRL_A | NC | CH2_CTRL_B | CH2_CTRL_A |
| 3 | GND | GND | GND | GND | GND |
| 4 | CH5_CTRL_B | CH5_CTRL_A | NC | CH4_CTRL_B | CH4_CTRL_A |
| 5 | CH7_CTRL_B | CH7_CTRL_A | NC | CH6_CTRL_B | CH6_CTRL_A |
| 6 | GND | GND | GND | GND | GND |
| 7 | NC | NC | NC | NC | NC |
| 8 | NC | NC | NC | NC | NC |
| 9 | GND | GND | GND | GND | GND |
| 10 | NC | NC | NC | NC | NC |
| 11 | NC | NC | NC | NC | NC |
| 15 | NC | NC | NC | NC | NC |
| 16 | NC | NC | NC | NC | NC |
| 17 | GND | GND | GND | GND | GND |
| 18 | NC | NC | NC | NC | NC |
| 19 | NC | NC | NC | NC | NC |
| 20 | GND | GND | GND | GND | GND |
| 21 | NC | NC | NC | NC | NC |
| 22 | NC | NC | NC | NC | NC |
| 23 | GND | GND | GND | GND | GND |
| 24 | NC | NC | NC | NC | NC |
| 25 | NC | NC | NC | NC | NC |
| Signal | Description |
|---|---|
| CHx_CTRL_A | Gain control for signal A output generation |
| CHx_CTRL_B | Gain control for signal B output generation |
| GND | Ground connection |
| NC | No connection |
XJ3 Pinout
XJ3 is an 8-blade Universal Power Module (UPM). XJ3 is in the standard position with pinout for fully compatible rear I/O.
| Pins | SLSC Signal Name | Instrument Bus Signal Name | Usage Description |
|---|---|---|---|
| H | V1+ | Instrument_0+ | Input: Test Excitation+ |
| G | V1- | Instrument_0- | Input: Test Excitation- |
| F | V2+ | Instrument_1+ | Output: Test Signal AB+ |
| E | V2- | Instrument_1- | Output: Test Signal AB- |
| D | NC | NC | — |
| C | NC | NC | — |
| B | NC | NC | — |
| A | NC | NC | — |
SET-1240 Design Standards and Compatibility
The SET-1240 complies with the following specifications in the Switch Load and Signal Conditioning Module Design Specifications document.
| Switch Load and Signal Conditioning (SLSC) Module Design Specifications Version | 1.6 |
|---|---|
| SLSC Compliance Level | 1 |
| Rear I/O Compatibility Category | Fully Compatible Rear I/O using XJ2 and XJ3 in standard locations. |
| RTI Compatibility Category Glyph | [N3] |
SET-1240 Physical Characteristics
Explore the dimensions and weight specifications of the device to understand its physical characteristics.
| Dimensions | 175 mm × 31 mm × 336 mm (6.89 in. × 1.19 in. × 13.21 in.) |
|---|---|
| Weight | 480 g (16.9 oz) |
SET-1240 Electrical Characteristics
| 1 A, 30 V DC Resistive | 105 operations |
|---|
| DC Resistance of Transformers | 115 Ω ±15% |
|---|
| Recommended Operating Range | 400 Hz to 7 kHz |
|---|
SET-1240 Power Requirements
| Voltage/Current Rating | 3.3 V at 65 mA 24 V at 0.5 A |
|---|---|
| Power Rating | 12.2 W |
SET-1240 Rated Voltages
Rated voltages are the voltage value or range at which the specified circuit’s operation and performance characteristics have been evaluated and tested.
| Maximum Input Voltage, Any Pin on J1, J2, and XJ3 (Channel-to-Channel, Channel-to-Ground) | ±60 V DC |
|---|---|
| Maximum CTRL Voltage on XJ2 | ±10 V DC |
| Maximum Switching Power | <1 W |
| Maximum Voltage Difference (Channel-to-Channel, Channel-to-Ground) | 60 V DC, 42.4 V pk |
| Maximum Excitation AC Voltage (400 Hz to 1 kHz) | 3 V RMS |
| Maximum Excitation AC Voltage (1 kHz to 7 kHz) | 7 V RMS |
| Maximum Common Mode Excitation Voltage Offset | ±45 V DC |
Measurement Category I Guidelines
Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated low-voltage sources, and electronics.
SET-1240 Fuse Ratings
| Current Rating | 3 A |
|---|---|
| Voltage Rating | 125 V AC, 125 V DC |
| Size and Dimensions | 7.4 mm × 3.1 mm |
| Type | Time-Lag T |
| Breaking Capacity | 100 A |
| Recommended Replacement | Schurter 3404.0118.22 |
Environmental Guidelines
SET-1240 Environmental Characteristics
| Operating Temperature | 0 °C to 40 °C, chassis intake |
|---|---|
| Storage Temperature | -40 °C to 85 °C |
| Operating Humidity | 10% RH to 90% RH, non-condensing |
|---|---|
| Storage Humidity | 5% RH to 95% RH, non-condensing |
| Pollution Degree | 2 |
|---|
| Maximum Altitude | 2,000 m |
|---|
| Operating Vibration | 5 Hz to 500 Hz, 0.3 g RMS |
|---|---|
| Non-Operating Vibration | 5 Hz to 500 Hz, 2.4 g RMS |
| Operating Shock | 30 g, half-sine, 11 ms pulse |
| Non-Operating Shock | 50 g, half-sine, 11 ms pulse |