SET-1240 Specifications

These specifications apply to the SET-1240.

Notice To ensure the specified EMC performance, operate this product only with shielded cables and accessories on J1 and J2 ports/connections, HD-44 cable, Trigger in/out, and Ethernet.

Revision History

Version Date changed Description
379273A-01 April 2026 Initial release.

Looking For Something Else?

For information not found in the specifications for your product, such as operating instructions, browse Related Information.

Definitions

Warranted Specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.

Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

  • Typical—describes the performance met by a majority of models.
  • Nominal—describes an attribute that is based on design, conformance testing, or supplemental testing.
  • Measured—describes the measured performance of a representative model.
  • Values are Typical unless otherwise noted.

    Conditions

    Specifications are valid under the following conditions unless otherwise noted.
    • All voltages are relative to GND

    SET-1240 Pinouts

    Access detailed pinout information for connecting to J1, J2, XJ2, and XJ3 connectors on the SET-1240.

    J1 and J2 Pinout

    J1 and J2 are female 44-pin high-density D-SUB connectors with a 4-40 UNC screw lock.

    Figure 1. SET-1240 J1 and J2 Connector Pinout (Front)


    Table 1. J1 and J2 Connector Signal Descriptions
    Signal Description (6-Wire Mode) Description (4/5-Wire Mode)
    CHx_A+ Emulated A+ signal output for Channel x Emulated + signal output for Channel x
    CHx_A- Emulated A- signal output for Channel x Common line. Shorted to Chx_B+
    CHx_B+ Emulated B+ signal output for Channel x Common line. Shorted to Chx_A-
    CHx_B- Emulated B- signal output for Channel x Emulated - signal output for Channel x
    CHx_EXC+/- Excitation input Excitation input
    NC No connection No connection

    XJ2 Pinout

    XJ2 is a 110-pin Hard Metric type A connector, per IEC 61076-101. XJ2 is in the standard position with pinout for fully compatible rear I/O.

    Figure 2. SET-1240 XJ2 Connector Pinout (Rear)


    Table 2. XJ2 Connector Pin Assignments
    Row E D C B A
    1 CH1_CTRL_B CH1_CTRL_A NC CH0_CTRL_B CH0_CTRL_A
    2 CH3_CTRL_B CH3_CTRL_A NC CH2_CTRL_B CH2_CTRL_A
    3 GND GND GND GND GND
    4 CH5_CTRL_B CH5_CTRL_A NC CH4_CTRL_B CH4_CTRL_A
    5 CH7_CTRL_B CH7_CTRL_A NC CH6_CTRL_B CH6_CTRL_A
    6 GND GND GND GND GND
    7 NC NC NC NC NC
    8 NC NC NC NC NC
    9 GND GND GND GND GND
    10 NC NC NC NC NC
    11 NC NC NC NC NC
    15 NC NC NC NC NC
    16 NC NC NC NC NC
    17 GND GND GND GND GND
    18 NC NC NC NC NC
    19 NC NC NC NC NC
    20 GND GND GND GND GND
    21 NC NC NC NC NC
    22 NC NC NC NC NC
    23 GND GND GND GND GND
    24 NC NC NC NC NC
    25 NC NC NC NC NC
    Table 3. XJ2 Connector Signal Descriptions
    Signal Description
    CHx_CTRL_A Gain control for signal A output generation
    CHx_CTRL_B Gain control for signal B output generation
    GND Ground connection
    NC No connection

    XJ3 Pinout

    XJ3 is an 8-blade Universal Power Module (UPM). XJ3 is in the standard position with pinout for fully compatible rear I/O.

    Figure 3. SET-1240 XJ3 Connector Pinout (Rear)


    Table 4. XJ3 Connector Pin Assignments
    Pins SLSC Signal Name Instrument Bus Signal Name Usage Description
    H V1+ Instrument_0+ Input: Test Excitation+
    G V1- Instrument_0- Input: Test Excitation-
    F V2+ Instrument_1+ Output: Test Signal AB+
    E V2- Instrument_1- Output: Test Signal AB-
    D NC NC
    C NC NC
    B NC NC
    A NC NC

    SET-1240 Design Standards and Compatibility

    The SET-1240 complies with the following specifications in the Switch Load and Signal Conditioning Module Design Specifications document.

    Table 5. Design Standards and Compatibility
    Switch Load and Signal Conditioning (SLSC) Module Design Specifications Version 1.6
    SLSC Compliance Level 1
    Rear I/O Compatibility Category Fully Compatible Rear I/O using XJ2 and XJ3 in standard locations.
    RTI Compatibility Category Glyph [N3]

    SET-1240 Physical Characteristics

    Explore the dimensions and weight specifications of the device to understand its physical characteristics.

    Table 6. Dimensions and Weight
    Dimensions 175 mm × 31 mm × 336 mm

    (6.89 in. × 1.19 in. × 13.21 in.)

    Weight 480 g (16.9 oz)

    SET-1240 Electrical Characteristics

    Table 7. Expected Electrical Relay Lifetime
    1 A, 30 V DC Resistive 105 operations
    Table 8. DC Resistance
    DC Resistance of Transformers 115 Ω ±15%
    Table 9. Excitation Frequency
    Recommended Operating Range 400 Hz to 7 kHz

    SET-1240 Power Requirements

    Table 10. SLSC Bus
    Voltage/Current Rating 3.3 V at 65 mA

    24 V at 0.5 A

    Power Rating 12.2 W

    SET-1240 Rated Voltages

    Rated voltages are the voltage value or range at which the specified circuit’s operation and performance characteristics have been evaluated and tested.

    Table 11. Rated Voltages
    Maximum Input Voltage, Any Pin on J1, J2, and XJ3 (Channel-to-Channel, Channel-to-Ground) ±60 V DC
    Maximum CTRL Voltage on XJ2 ±10 V DC
    Maximum Switching Power <1 W
    Maximum Voltage Difference (Channel-to-Channel, Channel-to-Ground) 60 V DC, 42.4 V pk
    Maximum Excitation AC Voltage (400 Hz to 1 kHz) 3 V RMS
    Maximum Excitation AC Voltage (1 kHz to 7 kHz) 7 V RMS
    Maximum Common Mode Excitation Voltage Offset ±45 V DC
    Hazardous Voltage This product is designed for use with non-hazardous, live signals only. Ensure that all signals are isolated from hazardous, live circuits and that unsafe voltages are not present at the inputs. Exceeding the voltage specifications may cause electric shock or damage the product.
    Tension dangereuse Ce produit est conçu uniquement pour être utilisé avec des signaux en direct non dangereux. Assurez-vous que tous les signaux sont isolés des circuits dangereux sous tension et qu’aucune tension dangereuse n’est présente aux entrées. Le dépassement de la tension spécifiée peut provoquer un choc électrique ou endommager le produit.
    Caution Any external sources must be limited to not exceed these maximum rated currents.
    Attention Les sources externes doivent être limitées pour ne pas dépasser ces tensions nominales maximales.

    Measurement Category I Guidelines

    Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated low-voltage sources, and electronics.

    Note Measurement Categories CAT I and CAT O are equivalent. These test and measurement circuits are for other circuits not intended for direct connection to the MAINS building installations of Measurement Categories CAT II, CAT III, or CAT IV.
    Hazardous Voltage Do not connect the product to signals or use for measurements within Measurement Categories II, III, or IV.
    Tension dangereuse Ne pas connecter le produit à des signaux dans les catégories de mesure II, III ou IV et ne pas l'utiliser pour effectuer des mesures dans ces catégories.
    Hazardous Voltage Do not connect the product to signals or use for measurements within Measurement Categories II, III, or IV, or for measurements on MAINs circuits or on circuits derived from Overvoltage Category II, III, or IV which may have transient overvoltages above what the product can withstand. The product must not be connected to circuits that have a maximum voltage above the continuous working voltage, relative to earth or to other channels, or this could damage and defeat the insulation. The product can only withstand transients up to the transient overvoltage rating without breakdown or damage to the insulation. An analysis of the working voltages, loop impedances, temporary overvoltages, and transient overvoltages in the system must be conducted prior to making measurements.
    Tension dangereuse Ne pas connecter le produit à des signaux dans les catégories de mesure II, III ou IV et ne pas l'utiliser pour des mesures dans ces catégories, ou des mesures sur secteur ou sur des circuits dérivés de surtensions de catégorie II, III ou IV pouvant présenter des surtensions transitoires supérieures à ce que le produit peut supporter. Le produit ne doit pas être raccordé à des circuits ayant une tension maximale supérieure à la tension de fonctionnement continu, par rapport à la terre ou à d'autres voies, sous peine d'endommager et de compromettre l'isolation. Le produit peut tomber en panne et son isolation risque d'être endommagée si les tensions transitoires dépassent la surtension transitoire nominale. Une analyse des tensions de fonctionnement, des impédances de boucle, des surtensions temporaires et des surtensions transitoires dans le système doit être effectuée avant de procéder à des mesures.

    SET-1240 Fuse Ratings

    Table 12. Fuse Designation
    Current Rating 3 A
    Voltage Rating 125 V AC, 125 V DC
    Size and Dimensions 7.4 mm × 3.1 mm
    Type Time-Lag T
    Breaking Capacity 100 A
    Recommended Replacement Schurter 3404.0118.22

    Environmental Guidelines

    Notice Failure to follow the mounting instructions in the SET-1240 User Manual can cause temperature derating.
    Notice Use this product indoors only.

    SET-1240 Environmental Characteristics

    Table 13. Temperature
    Operating Temperature 0 °C to 40 °C, chassis intake
    Storage Temperature -40 °C to 85 °C
    Table 14. Humidity
    Operating Humidity 10% RH to 90% RH, non-condensing
    Storage Humidity 5% RH to 95% RH, non-condensing
    Table 15. Pollution Degree
    Pollution Degree 2
    Table 16. Maximum Altitude
    Maximum Altitude 2,000 m
    Table 17. Shock and Vibration
    Operating Vibration 5 Hz to 500 Hz, 0.3 g RMS
    Non-Operating Vibration 5 Hz to 500 Hz, 2.4 g RMS
    Operating Shock 30 g, half-sine, 11 ms pulse
    Non-Operating Shock 50 g, half-sine, 11 ms pulse