PXIe-4154 Ranges
PXIe-4154 Voltage Ranges
For voltage output, the PXIe-4154 uses the following ranges.
| Channel | Voltage Level Range | Voltage Level |
|---|---|---|
| 0 | 6 V | 0 V to +6 V |
| 1 | 8 V | 0 V to +8 V |
| Channel | Current Limit Range | Current Limit |
|---|---|---|
| 0 | 30 mA | +100 µA to +30 mA |
| 3 A | +10 mA to +3 A | |
| 1 | 1.5 A | +10 mA to +1.5 A |
PXIe-4154 Current Ranges
For current output, the PXIe-4154 uses the following ranges.
Notice If you change a current range while the PXIe-4154 is constant current mode, a spike in the output
current could occur for a maximum of 2 ms. This does not
occur if the PXIe-4154 is in constant voltage
mode.
| Channel | Current Level Range | Current Level |
|---|---|---|
| 0 | 30 mA | +100 µA to +30 mA |
| 3 A | +10 mA to +3 A | |
| 1 | 1.5 A | +10 mA to +1.5 A |
| Channel | Voltage Limit Range | Voltage Limit |
|---|---|---|
| 0 | 6 V | 0 V to +6 V |
| 1 | 8 V | 0 V to +8 V |
Overranging
Enabling overranging for a particular channel of this instrument extends voltage output capabilities from 100% to 105% and current capabilities from 100% to 103.5% for the output range.
Overranging is applicable to output ranges only and does not apply to measurement ranges. Measurements in any given range may be made up to 105% of the range by default without enabling overranging.