PXIe-4154 Pinout
- Updated2025-11-19
- 1 minute(s) read
The following figure shows the terminals on the CH 0 and CH 1 connectors on the PXIe-4154.
| Item | Signal Description |
|---|---|
| Sense LO | Voltage remote sense input terminals. Used to compensate for I*R voltage drops in cable leads, connectors, and switches. |
| Output LO | LO force terminal connected to channel power stage (generates and/or dissipates power). Positive polarity is defined as voltage measured on HI > LO. |
| Output HI | HI force terminal connected to channel power stage (generates and/or dissipates power). Positive polarity is defined as voltage measured on HI > LO. |
| Sense HI | Voltage remote sense input terminals. Used to compensate for I*R voltage drops in cable leads, connectors, and switches. |
| Chassis ground | Tied to chassis ground through module front panel. Use for connections to cable shields or grounding the LO force terminal. |
Note Front panel connectors can accept wire gauges from 12 AWG to 24
AWG.