ETX-1640x AC PDU Ratings

The ratings in the following tables are applicable for the RMX-10151 single-phase power distribution unit (PDU) within the ETX-1640x.

Table 28. Input Power Rating
Power connector type IEC 60320 C20
Overvoltage category II
Current rating 16 A
Note The input current rating must be derated to the same value as the upstream outlet it is plugged in to.
Voltage rating

ETX-16400: 240 V AC

ETX-16401: 100 V AC to 240 V AC

Frequency rating 50/60 Hz
Table 29. Output Power Rating
Power connector type C13
Overvoltage category II
Current rating per outlet 10 A
Note The sum of the total current for all C13 outlets cannot exceed the input current rating.
Current rating per bank 10 A
Note There are eight C13 outlets in banks of two.
Total current rating (all outlets) 16 A
Output voltage Same as input voltage
Frequency rating 50/60 Hz
Table 30. AC Outputs for DC PDUs (AUX DC 600 Power Outlets)
Power connector type Molex-style two pin connector, 76825-0002
Output voltage Same as input voltage
Frequency rating 50/60 Hz
Current rating Fused at 6.3 A
Pin 1 L1
Pin 2 L2
Note L2 could be neutral depending on facility power.
Wire gauge 18 AWG to 24 AWG
Table 31. AC Outputs for DC PDUs (AUX DC 1200 Power Outlets)
Power connector type C13
Output voltage Same as input voltage
Current rating 10 A
Frequency rating 50/60 Hz