Site Preparation
- Updated2022-04-20
- 1 minute(s) read
Site Preparation
You must follow all safety and site requirements and instructions to properly set up, configure, and use the ECUTS.
Ensure the facility meets the following environmental, physical, and interface requirements before the ECUTS arrives.
ECUTS-16000 Environmental, Physical, and Interface Requirements
Ventilation Clearances
The following clearances are required to ensure proper airflow to the ECUTS. Clearances are
measured from the ECUTS test rack exterior and do not include the 9025TR platform or monitor arm. Ensure
obstructions do not violate the following clearance requirements:
- Top: 760 mm (30 in.)
- Front and Back: 300 mm (12 in.)
Caution You must follow the ECUTS
power dissipation and ventilation requirements to avoid thermal hazards.
Attention Vous devez suivre les exigences de
ventilation et de dissipation de puissance du ECUTS pour éviter tout risque
thermique.