This document contains the Calibration Executive known issues that were discovered before and since the release of Calibration Executive 6.4. Known issues are performance issues or technical bugs that NI has acknowledged exist within this version of the product.
Not every issue known to NI appears on this list; it is intended to show the most severe and common issues that you may encounter and provide workarounds when possible. Other technical issues that you may encounter could occur through normal product use or system compatibility issues. You may find more information on these issues in NI’s Product Documentation, Knowledgebase, or Community.
Bug Number |
Legacy ID |
Description |
Details |
---|---|---|---|
208070 | 677002 |
Procedure Throws Error or Terminates When a PXIe-4080/4081/4082 DMM Is UsedPXIe-408x digital multimeters are not compatible with several Calibration Executive procedures for which a PXI-407x digital multimeter is recommended as an instrument standard. For more information and a list of procedures, refer to "DMM Instrument Standard Considerations" in Calibration
Executive Help.
Workaround: Do not substitute a PXIe-4080/4081/4082 in a procedure in which a PXI-4070/4071/4072 DMM is recommended.
|
Reported Version: Calibration Executive 4.2 Resolved Version: N/A Added: Aug 14, 2020 |
1090478 |
NIPM Does Not Enforce Dependencies in Calibration Executive FeedNI Package Manager allows the Calibration Executive feed to be modified to remove core components that are required for certain procedures. Dependencies should not be modified or deselected during installation.
Workaround: There is currently no known workaround for this issue. |
Reported Version: Calibration Executive 6.0 Resolved Version: N/A Added: Aug 14, 2020 |
|
1057234 |
In Read-Only Mode, Uncertainty Database Error Handling Is ImpairedWhen the uncertainty database is read only (for example, the operator logs in without administrator privileges), an initial attempt to manage uncertainties causes Error -2147217911 when launching the Manage Uncertainty UI, as Calibration Executive tries to update the legacy database. After the database is updated to the new format in Calibration Executive 6.0 and later, the initial error when launching the Manage Uncertainty UI is resolved. But any operation that modifies the uncertainties database results in Error -2147467259. Actions that can result in this error include:
Workaround: To resolve these errors, you should remove the read-only attribute of the uncertainty database, and then re-open the Manage Uncertainty UI.
|
Reported Version: Calibration Executive 6.0 Resolved Version: N/A Added: Oct 8, 2020 |
|
1094770 |
Error Occurs Generating a Calibration Report on a Non-English OSWindows OS for languages other than English (Spanish, for example) may use a date format with extended characters. In this case, PDF, Word, and Excel reports cannot be generated using default report options; Calibration Executive displays an error:
Workaround: Configure report options to remove Calibration Date from the report file name. For details about customizing report names, refer to "Configuring Report Options" in Calibration Executive Help.
|
Reported Version: Calibration Executive 6.0 Resolved Version: N/A Added: Aug 14, 2020 |
|
1091448 |
Installing Version 6.0 on Japanese OS with Suggested Components Causes ErrorWhen you install Calibration Executive 6.0 on a Japanese operating system and accept the pre-selected additional items, an error occurs at the NI-HSDIO-Documentation package. Workaround: When you install Calibration Executive on a Japanese OS, deselect all NI-HSDIO components from the additional items list in NI Package Manager. In this case, however, the Japanese OS system cannot be used to calibrate high-speed digital I/O hardware.
|
Reported Version: Calibration Executive 6.0 Resolved Version: N/A Added: Oct 8, 2020 |
|
1090572 |
Error Occurs If User Imports Custom Report Configuration from a Different FolderFrom the Custom Report Configurator, if you try to import a new report configuration from a location other than the default folder, the following error occurs:
Workaround: Copy individual stylesheet
folders to the default report configuration folder location. In the Custom Report Configurator, click Import, and select the main CustomReports folder to import all stylesheets, or select each stylesheet folder individually.
|
Reported Version: Calibration Executive 6.0 Resolved Version: N/A Added: Dec 14, 2020 |
|
1084777 |
Error Occurs If Path to Custom Report Stylesheet or CustomReports Folder ChangesIf the file path for a custom report stylesheet or the CustomReports folder changes (for example, if a user's Documents folder is cleaned up), Calibration Executive displays an error when the report generator is launched:
The report generator then closes abruptly.
Workaround: Do one of the following:
|
Reported Version: Calibration Executive 6.0 Resolved Version: N/A Added: Aug 14, 2020 |
|
1065569 |
Custom Report Features Cannot Be Shared Across User Accounts on a WorkstationAccess to custom reports on a single workstation is limited to the account used to install and use Calibration Executive. By default, the path for custom reports data is:
Workaround: Copy the content of the CustomReports folder (including the style sheet) to a shared folder, and then edit the cex_Config.ini file at C:\Program Files\National Instruments\Calibration Executive\Config\ to change CustomReportsPath to the new location.
|
Reported Version: Calibration Executive 6.0 Resolved Version: N/A Added: Dec 14, 2020 |
|
1730375 |
Error Occurs at Verify RF Output Non-Harmonic Spurs Step for PXIe-564xR ProceduresCalibration Executive throws the following error during the Verify RF Output Non-Harmonic Spurs step for the PXIe-5644R, PXIe-5645R, and PXIe-5646R procedures. This issue was introduced in Calibration Executive 6.4.
Workaround: Use Calibration Executive 6.3 or earlier to run PXIe-5644R, PXIe-5645R, and PXIe-5646R procedures until a fix becomes available for Calibration Executive 6.4
|
Reported Version: Calibration Executive 6.4 Resolved Version: N/A Added: Nov 24, 2021 |
|
1585154 |
NI-FGEN Standards Are Not Available in The Setup WizardNI-FGEN standards are not showing up in the setup wizard. Specifically, the Instrument driver for National Instruments Signal Generator option is missing from the list of available standards. This affects all procedures that require NI-FGEN standards. This issue occurs when you upgrade to Calibration Executive 6.3 or 6.4 from an earlier version. When you perform a clean install of Calibration Executive 6.3 or 6.4, the Instrument driver for National Instruments Signal Generator option is available.
Workaround: Perform the following steps to repair the NI-FGEN driver:
|
Reported Version: Calibration Executive 6.3 Resolved Version: N/A Added: Sep 1, 2021 |
Issues found in this section will not be listed in future known issues documents for this product.
There are currently no issues to list.