Last Modified: September 22, 2017
Front Panel
Table 1. Connector Descriptions
Connector |
Description |
CH 0, CH 1 |
Analog input connection; digitizes data and triggers acquisitions |
TRIG |
External analog trigger connection; signals on the TRIG connector cannot be digitized |
CLK IN |
Input for an external reference clock or sample clock to the device |
AUX I/O |
Provides access to the multipurpose digital timing and triggering lines, PFI 0, PFI 1 (with optional cable) |
Table 2. LED Indicators
LED |
Color |
Description |
ACCESS |
None (LED is off) |
The PXIe-5114 is not yet functional or has detected a problem with a power rail. |
Amber |
The PXIe-5114 is being accessed. |
Green |
The PXIe-5114 is ready to be programmed by NI-SCOPE. |
ACTIVE |
None (LED is off) |
The PXIe-5114 is not armed, triggered, or acquiring a waveform. |
Amber |
The PXIe-5114 is armed and waiting for a trigger. |
Green |
The PXIe-5114 has received a Reference (Stop) trigger or is acquiring a waveform. |
Red |
The PXIe-5114 has detected an error. Access the module with NI-SCOPE to determine the cause of the error. This LED remains red until the error condition is removed. |
AUX I/O Connector Pinout
Table 3. Signal Descriptions
Pin |
Description |
1 |
5 V (Fused) |
2 |
GND |
3 |
Reserved |
4 |
Reserved |
5 |
Reserved |
6 |
PFI 1 |
7 |
Reserved |
8 |
Reserved |
9 |
PFI 0 |
Recently Viewed Topics