PXIe-5111 Trigger Sources

Version:
    Last Modified: June 13, 2019

    The following diagram illustrates the trigger sources on the PXIe-5111.

    1. Available triggers are Start (Acquisition Arm), Arm Reference, Reference (Stop), and Advance.
    2. Available output triggers are Start Trigger Out, Reference (Stop) Trigger Out, and Advance Out.
    3. Analog waveform interpolated for sub-sample trigger location. Other triggers are not sub-sample accurate.