SHDB13W6-4BNCF-LL Cable
- Updated2025-09-13
- 2 minute(s) read
SHDB13W6-4BNCF-LL Cable
The SHDB13W6-4BNCF-LL is a D-SUB to female BNC cable for connecting the PXIe-4190.
SHDB13W6-4BNCF-LL Cable Pinout
| Left (D-SUB) | Right (BNC/Ring Terminal) | LCR Mode Functionality | SMU Mode Functionality | Cable Color |
|---|---|---|---|---|
| PIN 1 to PIN 7 | — | General purpose input/output | — | |
| A6 (Outer Conductor) | BNC 4 (Outer Conductor) | Isolated Shield* | Red | |
| A6 (Inner Conductor) | BNC 4 (Inner Conductor) | HI CUR | LO | |
| A5 (Outer Conductor) | BNC 3 (Outer Conductor) | Isolated Shield* | Orange | |
| A5 (Inner Conductor) | BNC 3 (Inner Conductor) | HI POT | Sense LO | |
| A2 (Outer Conductor) | BNC 2 (Outer Conductor) | Isolated Shield* | GUARD* | Gray |
| A2 (Inner Conductor) | BNC 2 (Inner Conductor) | LO POT | Sense HI | |
| A1 (Outer Conductor) | BNC 1 (Outer Conductor) | Isolated Shield* | GUARD* | Black |
| A1 (Inner Conductor) | BNC 1 (Inner Conductor) | LO CUR | HI | |
| Outer Shell | GND | Chassis Ground | Green | |
| * The female BNC mounting plate is conductive and ties GUARD/Isolated Shield terminals together, terminating the 4TP connection. As a result, additional considerations apply when the DUT is not located near the mounting plate, such as when using the cable as a test port extension. | ||||
Note Isolated Shield and GUARD are electrically equivalent and are allowed to be tied
together in SMU mode.