PXI-2536 Specifications
- Updated2025-05-22
- 6 minute(s) read
PXI-2536 Specifications
PXI-2536 Specifications
Definitions
Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
- Typical specifications describe the performance met by a majority of models.
- Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
Specifications are Warranted unless otherwise noted.
Conditions
Specifications are valid at 23 °C unless otherwise noted.
All voltages are specified in DC, ACpk, or a combination unless otherwise specified.
PXI-2536 Pinout
PXI-2536 Front Panel
- Columns 0:67 connector
- Rows 0:7 IN connector
- Rows 0:7 OUT connector
Columns 0:67 Connector
| Signal | Description |
|---|---|
| Cx | Column connection |
Rows 0:7 IN Connector
| Signal | Description |
|---|---|
| Rx | Row input connection |
| — | No connection |
Rows 0:7 OUT Connector
| Signal | Description |
|---|---|
| Rx | Row output for column expansion |
| RSVD | Reserved, do not connect |
| — | No connection |
Topology
Topology | 1-wire 8 × 68 matrix |
Input Characteristics
Maximum switching voltage (channel-to-ground) | ±12 VDC, 8 VAC | ||||||
Maximum switching power (per channel, resistive) | 1.2 W | ||||||
Maximum switching current | 100 mA | ||||||
| |||||||
Current leakage between column and ground (closed path)[1]1 12 VDC applied at 25° C | 10 nA, typical | ||||||
Offset voltage | 10 µV, typical | ||||||
| |||||||
Total path resistance, row-to-column | 10 Ω, typical | ||||||
Maximum total path resistance, row-to-column | 15.5 Ω, maximum | ||||||
RF Performance Characteristics
Single crosspoint bandwidth (50 Ω system, one row to one column) | >1 MHz, typical | ||||||||
| |||||||||
Dynamic Characteristics
FET operate time[2]2 Operate time is the time from trigger received by hardware to switch output activation.[3]3 Certain applications may require additional time for proper settling. | 12 µs, typical 16 µs, maximum |
Maximum scan rate | 50,000 crosspoints/s |
Simultaneous drive limit | 544 switches |
Expected relay life | Unlimited, when operated within specified limits |
| |||||||
Trigger Characteristics
| |||||||
| |||||||
Physical Characteristics
Relay type | FET switch | ||||||
I/O connector | Three 68-pin receptacle VHDCI | ||||||
| |||||||
Dimensions (L × W × H) | 3U, one slot, PXI/cPCI module 21.6 × 2.0 × 13.0 cm (8.5 × 0.8 × 5.1 in.) | ||||||
Weight | 159 g (5.6 oz) | ||||||
Environment
Operating temperature | 0 °C to 55 °C |
Storage temperature | -40 °C to 70 °C |
Relative humidity | 5% to 85%, noncondensing |
Pollution Degree | 2 |
Maximum altitude | 2,000 m |
Indoor use only.
Shock and Vibration
| Operational Shock | 30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Test profile developed in accordance with MIL-PRF-28800F.) |
| Random vibration, operating | 5 Hz to 500 Hz, 0.3 grms |
| Random vibration, nonoperating | 5 Hz to 500 Hz, 2.4 grms (Tested in accordance with IEC 60068-2-64. Nonoperating test profile exceeds the requirements of MIL-PRF-28800F, Class 3.) |
Compliance and Certifications
Safety Compliance Standards
This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:
- IEC 61010-1, EN 61010-1
- UL 61010-1, CSA C22.2 No. 61010-1
Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for sensitive electrical equipment for measurement, control, and laboratory use:
- EN 61326-2-1 (IEC 61326-2-1): Class A emissions; Basic immunity
- EN 55011 (CISPR 11): Group 1, Class A emissions
- AS/NZS CISPR 11: Group 1, Class A emissions
- FCC 47 CFR Part 15B: Class A emissions
- ICES-001: Class A emissions
CE Compliance

This product meets the essential requirements of applicable European Directives, as follows:
- 2014/35/EU; Low-Voltage Directive (safety)
- 2014/30/EU; Electromagnetic Compatibility Directive (EMC)
- 2011/65/EU; Restriction of Hazardous Substances (RoHS)
- 2014/53/EU; Radio Equipment Directive (RED)
- 2014/34/EU; Potentially Explosive Atmospheres (ATEX)
Product Certifications and Declarations
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, search by model number, and click the appropriate link.
Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.
For additional environmental information, refer to the Engineering a Healthy Planet web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.