Saves the externally measured power of a signal generated by a PXIe-5413/5423/5433, performs any required adjustments, and computes any calibration constants in the onboard memory.

This VI also indicates if flatness calibration is complete. You must repeat the flatness calibration adjustment procedure steps until this VI returns TRUE for flatness cal complete.


icon

Inputs/Outputs

  • cni5433_Ext_Cal_APIlvclass.png external cal session in

    External Cal Session In identifies an external calibration session.

  • cdbl.png measured power (dBm)

    Measured Power (dBm) specifies the power in dBM of the analog output signal generated by the waveform generator as measured by the power sensor connected to the waveform generator.

  • cerrcodeclst.png error in

    error in describes error conditions that occur before this node runs.

    The default is no error. This node executes regardless of an incoming error.

    This input contains status, code, and source, which provide standard error in functionality.

  • ini5433_Ext_Cal_APIlvclass.png external cal session out

    External Cal Session Out returns a reference to your external calibration session to wire to the next VI.

  • ibool.png flatness cal complete

    Flatness Cal Complete returns TRUE if flatness calibration is complete. If this parameter returns FALSE, you must repeat calls to the Flatness Cal Configure and Flatness Cal Adjust VIs until flatness calibration is complete.

  • ierrcodeclst.png error out

    error out contains error information. This output provides standard error out functionality.