Flatness Cal Adjust VI
- Updated2025-10-07
- 2 minute(s) read
Saves the externally measured power of a signal generated by a PXIe-5413/5423/5433, performs any required adjustments, and computes any calibration constants in the onboard memory.
This VI also indicates if flatness calibration is complete. You must repeat the flatness calibration adjustment procedure steps until this VI returns TRUE for flatness cal complete.

Inputs/Outputs
external cal session in
—
External Cal Session In identifies an external calibration session.
measured power (dBm)
—
Measured Power (dBm) specifies the power in dBM of the analog output signal generated by the waveform generator as measured by the power sensor connected to the waveform generator.
error in
—
error in describes error conditions that occur before this node runs. The default is no error. This node executes regardless of an incoming error. This input contains status, code, and source, which provide standard error in functionality.
external cal session out
—
External Cal Session Out returns a reference to your external calibration session to wire to the next VI.
flatness cal complete
—
Flatness Cal Complete returns TRUE if flatness calibration is complete. If this parameter returns FALSE, you must repeat calls to the Flatness Cal Configure and Flatness Cal Adjust VIs until flatness calibration is complete.
error out
—
error out contains error information. This output provides standard error out functionality. |
external cal session in
—
measured power (dBm)
—
error in
—
external cal session out
—
flatness cal complete
—
error out
—