Flatness Cal Configure VI
- Updated2025-10-07
- 3 minute(s) read
Configures the PXIe-5413/5423/5433 for an iteration of the flatness calibration step.

Inputs/Outputs
external cal session in
—
External Cal Session In identifies an external calibration session.
error in
—
error in describes error conditions that occur before this node runs. The default is no error. This node executes regardless of an incoming error. This input contains status, code, and source, which provide standard error in functionality.
external cal session out
—
External Cal Session Out returns a reference to your external calibration session to wire to the next VI.
current configuration
—
Current Configuration returns the current configuration of the waveform generator, including the channel being calibrated, the expected voltage range (Vpk-pk), the expected frequency, and whether the configuration is valid in the current iteration of the flatness calibration step. Use this information to configure the DMM to measure the voltage range and frequency of the signal on the specified channel.
next configuration
—
Next Configuration returns the next expected configuration of the waveform generator, including the channel being calibrated, the expected voltage range (Vpk-pk), the expected frequency, and whether the configuration will be valid in the next iteration of the flatness calibration step. Use this information to determine if the next configuration is within the limits of the DMM.
error out
—
error out contains error information. This output provides standard error out functionality. |
external cal session in
—
error in
—
external cal session out
—
current configuration
—
channel
—
expected vpp
—
valid
—
error out
—