Restore Last External Calibration Constants

Overwrites the current calibration constants with those from the last successful external calibration. Invalidates data from any self-calibrations performed since the last time an external calibration was performed.

1378

Inputs/Outputs

datatype_icon

session in

Handle that identifies your instrument session previously allocated by Initialize External Calibration.

datatype_icon

error in

Error conditions that occur before this node runs.

The node responds to this input according to standard error behavior.

Standard Error Behavior

Default value: No error

datatype_icon

session out

Reference to your instrument session to wire to the next node.

datatype_icon

error out

Error information.

The node produces this output according to standard error behavior.

Standard Error Behavior