The following items are notable issues fixed between the release of NI-DCPower 20.5 and NI-DCPower 20.6, including additional patches and service packs. If you have an issue ID, you can search this list to validate that the issue has been fixed. This is not an exhaustive list of issues fixed in the current version of NI-DCPower.
Bug Number |
Legacy ID |
Description |
Details |
---|---|---|---|
1113818 |
NI-DCPower Errors reported in English instead of Japanese on NI Linux Real-Time TargetsOn an NI Linux Real-Time target with the language set to Japanese, the NI-DCPower driver returns errors in English instead of Japanese.
Workaround: Copy and paste the error text information into a translation tool, or for a Japanese LabVIEW installation select Help >> Explain Error and enter the error number.
|
Reported Version: NI-DCPower 20.5 Resolved Version: NI-DCPower 20.6 Added: Sep 7, 2020 |
|
1125517 |
NI-DCPower error -1074097807 on NI Linux Real-Time Target when used with SystemLinkIf self-calibration or self-test runs from SystemLink before executing commands with the driver through MAX or LabVIEW, then driver operations including self-test and self-calibration in MAX or LabVIEW result in error -1074097807 "File was not found at specified path." If the driver executes commands in MAX or LabVIEW before running a self-test or self-calibration in SystemLink, then SystemLink throws error -1074097807.
Workaround: Option 1:
Option 2:
Option 3:
/var/lib/ni-dcpower/sessions/: file.directory: - source: /var/lib/ni-dcpower/sessions/ - user: admin - dir_mode: 777 - file_mode: 777 - recurse: - user - mode system.reboot: module.run |
Reported Version: NI-DCPower 20.5 Resolved Version: NI-DCPower 20.6 Added: Sep 7, 2020 |
|
1125517 |
NI-DCPower error -1074118653 or -1074097807 on LinuxRT when used with Pythonnidcpower.errors.DriverError: -1074118653: Cannot open a session to the instrument.
-1074097807: File was not found at specified path. If Python executes driver commands before MAX or LabVIEW, then driver operations in MAX or LabVIEW produce this error. Workaround:
|
Reported Version: NI-DCPower 20.5 Resolved Version: NI-DCPower 20.6 Added: Sep 7, 2020 |
|
1170689 |
Fetching Live Measurements from Simulated PXIe-4147 Crashes InstrumentStudioFetching live measurements in InstrumentStudio from a simulated PXIe-4147 device created from offline mode causes InstrumentStudio to crash.
Workaround: Update to NI-DCPower 20.6. If you are unable to update, utilize physical hardware (a non-simulated device).
|
Reported Version: NI-DCPower 20.5 Resolved Version: NI-DCPower 20.6 Added: N/A |
Installing some patches may require certain additional steps or considerations. Please refer to the following table for more information about patches for this release.
These patches currently do not have any special instructions.