The following items are notable issues fixed between the release of Calibration Executive 6.2 and Calibration Executive 6.3, including additional patches and service packs. If you have an issue ID, you can search this list to validate that the issue has been fixed. This is not an exhaustive list of issues fixed in the current version of Calibration Executive.
Bug Number |
Legacy ID |
Description |
Details |
---|---|---|---|
1066924 |
Fatal Error Occurs at AdjustmentThe following calibration procedures fail to create adjustment sessions and return a fatal error:
When this error occurs, you must terminate the procedure and either use Windows Device Manager to disable and then re-enable the PCI or PXI device or shut down the system completely and power it back on again. Verify Only run mode works correctly for NI 6704, NI 6713, NI 6722, NI 6723 and NI 6733 devices. Workaround: Use a previous version of Calibration Executive to execute the Verify & Adjust run mode on PCI/PXI-4461, -6704, and -6713 devices until a fix becomes available for version 6.0. |
Reported Version: Calibration Executive 6.0 Resolved Version: Calibration Executive 6.3 Added: Oct 8, 2020 |
|
1360252 |
"Verify Main Analog Path Flatness" Step for PXI-5422 is Missing a Test PointThe following test point is missing from the Verify Main Analog Path Flatness step in the PXI-5422 procedure. Workaround: There is currently no known workaround for this issue. |
Reported Version: Calibration Executive 6.1 Resolved Version: Calibration Executive 6.3 Added: N/A |
Installing some patches may require certain additional steps or considerations. Please refer to the following table for more information about patches for this release.
These patches currently do not have any special instructions.