NI-Digital Pattern Driver 19.0 Known Issues

Overview

This document contains the NI-Digital Pattern Driver known issues that were discovered before and since the release of NI-Digital Pattern Driver and Digital Pattern Editor 19.0. Not every issue known to NI appears on this list; it is intended to show the most severe and common issues that can be encountered.

Known Issues

899451NI-Digital does not validate Bit Order for Capture Engine reconfiguration
577343Loading large patterns in a 32-bit process runs out of memory
672536NI Example Finder does not filter for PXIe-6570 hardware

 

IDKnown Issue
899451

Return

NI-Digital does not validate Bit Order for Capture Engine reconfiguration


If multiple capture waveforms are used within a single session, the NI-Digital driver should validate Sample Width, Pins, Waveform Type (Serial/Parallel), and Bit Order (MSB/LSB) to determine if the Capture Engine must be reconfigured. The Bit Order setting is not being validated. This means that if the two capture waveform configurations differ in Bit Order and the other 3 settings are consistent, then no reconfiguration will take place and the data for one of the waveforms will be returned with incorrect bit order..

Workaround: Configure all waveforms to be a consistent Bit Order and reorder in software as necessary. Alternatively, close and create a new NI-Digital session between waveforms with settings that differ only in Bit Order.

Reported Version: 17.0  Resolved Version: N/A  Added: 11/25/2019
577343

Return
Loading large patterns in a 32-bit process runs out of memory
As a result of limited system memory in a 32-bit process, loading a large pattern in a 32-bit process might fail.

Workaround: Split large patterns loaded in a 32-bit process into patterns that are small enough to fit into the available system memory.

Reported Version: 16.0  Resolved Version: N/A  Added: 10/20/2016
672536

Return
NI Example Finder does not filter for PXIe-6570 hardware
Using hardware filtering in the NI Example Finder does not return any examples that support the PXIe-6570.

Workaround: To find PXIe-6570 examples, use the NI Example Finder search feature with "6570" as the keyword or browse the task structure under Hardware Input and Output»Modular Instruments»NI-Digital Pattern Driver.

Reported Version: 16.0  Resolved Version: N/A  Added: 10/20/2016

 

Contacting NI

Contact NI regarding this document or issues in the document. If you contact NI in regards to a specific issue, reference the ID number given in the document. The ID number contains the current issue ID number as well as the legacy ID number (use the current ID number when contacting NI). You can contact us through any of the normal support channels including phone, email, or the discussion forums. Visit the NI Website to contact us. Also contact us if you find a workaround for an issue that is not listed in the document.

Glossary of Terms

 

  • Bug ID - When an issue is reported to NI, you may be given this ID or find it on ni.com.  You may also find IDs posted by NI on the discussion forums or in KnowledgeBase articles.
  • Legacy ID – An older issue ID that refers to the same issue.  You may instead find this issue ID in older known issues documents.
  • Description - A few sentences which describe the problem. The brief description given does not necessarily describe the problem in full detail.
  • Workaround - Possible ways to work around the problem.
  • Reported Version - The earliest version in which the issue was reported.
  • Resolved Version - Version in which the issue was resolved or was no longer applicable. "N/A" indicates that the issue has not been resolved.
  • Date Added - The date the issue was added to the document (not the reported date).