| ID | Known Issue |
|---|
| 689967 | WTS Software may fail to start up if the WTS Instrument Software License for Qualcomm Connectivity Interface Support is activated without an NI-RFmx WLAN license.
The WTS Instrument Software License for Qualcomm Connectivity Interface Support also requires a valid NI-RFmx WLAN license. You must purchase and activate this license separately. Failure to activate the license can cause WTS Software to report errors at startup and not operate as expected.
Workaround: Ensure you purchase and activate the NI-RFmx WLAN license.
| Reported Version: 1.4.1 | | Resolved Version: N/A | | Added: 4/30/2018 |
|
| 675405 | In WTS Software, the instrument is unresponsive after sending a misformatted string using raw sockets.
The string must be sent using one of the following ways:
If you use a pling in the beginning of the string, you must also use the same pling in the end of the string. If you do not use the same string format and use raw sockets as communication method, the client can no longer communicate with the instrument.
Workaround: Use the same pling at the beginning and end of the strings. For example:- CONFigure:REFerence "INTERNAL",1
- CONFigure:REFerence INTERNAL,1
- CONFigure:REFerence'INTERNAL',1
| Reported Version: 1.4 | | Resolved Version: N/A | | Added: 12/18/2017 |
|
| 677407 | In WTS Software, the PXIe-5840 has a high LO leakage that causes measurement degradation when using Bluetooth.
When using Bluetooth with the PXIe-5840, the LO leakage is strong enough to degrade the measurements. The numbers may be fine for production test, but for characterization they may not be suitable.
Workaround: Contact your NI representative to request a fix.
| Reported Version: 1.4 | | Resolved Version: N/A | | Added: 12/18/2017 |
|
| 679212 | In WTS Software, auto broadcasting might fail during the first few signal generations.
When executing scenarios with auto broadcasting enabled, clients might not be joined or might throw an error during the first one to two signal generations, depending on how the signal is generated. This is related to the first time execution speed of some commands after the clients are initially joined by calling the SYSTem:RFSG:CONCurrency:TESTrun Start command, and also depends on the configured join time.
Workaround: Perform either of the following options as a workaround.- Execute the test plan for each virtual instrument alone once before using the auto broadcasting functionality. The auto broadcasting commands may be in the test plan during this single execution but they should not find another client to join with.
| Reported Version: 1.4 | | Resolved Version: N/A | | Added: 12/18/2017 |
|
| 677464 | In WTS Software, the WTS Instrument Interface terminates when the software restarts or is shutdown.
The WTS Instrument Interface requires a data connection to the WTS Software. When the connection to the WTS Software is lost, either because the instrument restarts or the software is terminated, all connected local and remote instances of the WTS Instrument Interface are terminated.
Workaround: Disconnect all instances of the WTS Instrument Interface before issuing a restart command.
| Reported Version: 1.4 | | Resolved Version: N/A | | Added: 12/18/2017 |
|
| 651101 | WTS Software does not work with other regional settings other than English-US.
SCPI and VISA support only using a period (.) as a decimal point and comma (,) to separate parameters and lists. WTS Software malfunctions and becomes unusable if you use any other settings.
Workaround: Ensure you are using (.) as the decimal symbol value, and (,) as the list separator value in Regional Settings»Additional Settings.
| Reported Version: 1.3 | | Resolved Version: N/A | | Added: 6/20/2017 |
|
| 583918 | In WTS Software, the WTS scheduling chart displays the signal analyzer as active when it waits for a trigger.
The signal analyzer is active when waiting for a trigger, and you cannot use it for anything else during that time.
Workaround: N/A
| Reported Version: 1.1 | | Resolved Version: N/A | | Added: 05/04/2016 |
|