Each issue appears as a row in the table and includes the following fields:
ID | Known Issue | |||||
---|---|---|---|---|---|---|
681944 Return | Certain programming sequences can result in internal bus contention that, over time, could damage the instrument. Contention can occur after using source memory on the PXIe-6570 internal bus. Over time this can cause damage to the instrument. Workaround: Install NI-Digital Patch 16.0.2 or 17.0.2 to resolve this issue. Additionally, any future version of NI-Digital post NI-Digital 17.5 will resolve this issue.
| |||||
673746 Return | HRAM may return an incomplete set of failures if HRAM is configured to use pretrigger samples and capture failures only If the HRAM settings within the Digital Pattern Editor are configured to have the Cycles to Capture to Failures Only and the Pretrigger Cycles is greater than 0, the failures captured may be different than the requested failures. All returned failures are actual failures, but some of the failures that were requested may not be returned. Workaround: Setting the Cycles to Capture to All will guarantee the correct samples are captured..
| |||||
649771 Return | HRAM may trigger on the wrong vector If you receive a digital scope error stating that niDigital_FetchHRAM failed to return data for requested cycle or that not enough cycle results exist in HRAM, HRAM might be triggering on the wrong vector. Workaround: If you use a repeat opcode three vectors after a label, four vectors after a call opcode, or between 160 and 180 vectors after a label, move the opcode down at least one vector or remove it.
| |||||
648748 Return | Calling the select function on any channel right after initialize sets all channels to the input value of the select function This situation causes the software state to be out of sync with the hardware state, and the PPMU source or burst pattern functions do not automatically put the channels in the correct state. Workaround: You must use the select function to manually select the state of each channel before using the PPMU source or burst pattern functions.
| |||||
640885 Return | Calling PPMU Source or Burst Pattern with a channel list could check the selected mode of the incorrect channel The PPMU Source or Burst Pattern functions may not select the correct mode for the specified channels when you call these functions with a channel list. Workaround: You can explicitly set the mode of each channel before you call the PPMU Source or Burst Pattern functions by using the Select function.
| |||||
577343 Return | Loading large patterns in a 32-bit process runs out of memory As a result of limited system memory in a 32-bit process, loading a large pattern in a 32-bit process might fail. Workaround: Split large patterns loaded in a 32-bit process into patterns that are small enough to fit into the available system memory.
| |||||
588214 Return | No LED activity during firmware update During a firmware update, there is no LED activity to indicate that there is an operation in progress. Workaround: The indicator ribbon in Measurement & Automation Explorer displays during an ongoing firmware update. The upgrade might take several minutes to complete.
| |||||
608891 Return | After updating firmware, the completion indicator disappears After you select another instrument in Measurement & Automation Explorer, the indicator ribbon that shows a successful firmware update disappears and the notification to power cycle the chassis is no longer present. Workaround: Turn off the chassis after all firmware updates are complete to finalize the updates.
| |||||
610028 Return | niDigital Export Signal does not export the start trigger The niDigital Export Signal VI and NIDigital.ExportSignal method do not allow exporting the start trigger. Workaround: Use the Start Trigger Output Terminal property to export the start trigger.
| |||||
610031 Return | Exporting the sync pulse errors out When attempting to synchronize multiple instruments, exporting the sync pulse returns an error that states the attribute is read only. Workaround: Use the TClk synchronization VIs to configure synchronization for multiple instruments.
| |||||
610148 Return | The Requirements section does not list the PXIe-6570 for NI-Digital examples in the NI Example Finder The PXIe-6570 is required for every NI-Digital example in the NI Example Finder. Workaround: N/A
| |||||
672536 Return | NI Example Finder does not filter for PXIe-6570 hardware Using hardware filtering in the NI Example Finder does not return any examples that support the PXIe-6570. Workaround: To find PXIe-6570 examples, use the NI Example Finder search feature with "6570" as the keyword or browse the task structure under Hardware Input and Output»Modular Instruments»NI-Digital Pattern Driver.
|
Contact NI regarding this document or issues in the document. If you contact NI in regards to a specific issue, reference the ID number given in the document. The ID number contains the current issue ID number as well as the legacy ID number (use the current ID number when contacting NI). You can contact us through any of the normal support channels including phone, email, or the discussion forums. Visit the NI Website to contact us. Also contact us if you find a workaround for an issue that is not listed in the document.