The following items are known issues in NI-HSDIO 2.0 sorted by Date.
ID | Known Issue | |||||
---|---|---|---|---|---|---|
306800 Return | Cannot perform self-calibration in MAX on NI 6555/6556. If you try to self-calibrate an NI 6555/6556 device in MAX it returns error -200445. Self-calibration of NI 6555/6556 devices takes between 15 and 20 minutes, which is longer than the allotted time period for self-calibration in MAX. Workaround: Self-calibrate your NI 6555/6556 device using the NI-HSDIO self-calibration VIs or functions instead.
| |||||
306802 Return | Building NI-HSDIO-linked 64-bit user applications requires an additional linker library directory and dependency. For building NI-HSDIO examples in 64-bit user mode applications that link to NI-HSDIO, you must add $(NIIVIPATH32)\Lib_x64 as an additional linker library directory and add niHSDIO.lib as an additional linker dependency. Workaround: Add $(NIIVIPATH32)\Lib_x64 as an additional linker library directory and add niHSDIO.lib as an additional linker dependency.
| |||||
306803 Return | When using a MXI connection, some NI PXI Express devices may not be recognized after a shutdown. When using a MXI connection, some NI PXI Express devices may not be recognized after a shutdown. Workaround: When you turn off your computer, either restart or power cycle your chassis before powering up your computer again.
| |||||
306807 Return | Error -219001 may occur on NI PXIe-6544/6545/6547/6548 devices when scynchronizing acquisitions using NI-TClk. For the NI PXIe-6544/6545/6547/6548, internal error -219001 may occur intermittently when synchronizing multiple acquisitions using NI-TClk at rates below 5 MHz. Workaround: If this error occurs, consider oversampling by configuring your acquisition with a higher sample rate. For more information, contact NI Technical Support at ni.com/support.
| |||||
306808 Return | The NI PXIe-6544/6545/6547/6548 may not utilize the full bandwidth of the PCI Express bus. In rare cases, the NI PXIe-6544/6545/6547/6548 may not utilize the full bandwidth of the PXI Express bus. These modules are x4 PXI Express devices, which are capable of using up to four 2.5 Gbps bus data lanes. Start-time negotiation between the PXI Express controller and the device may result in the device using only one or two data lanes. The result is reduced performance when reading or writing large amounts of waveform data. Workaround: This situation is normally corrected the next time you restart your system. For more information, contact NI Technical Support at ni.com/support.
| |||||
306810 Return | NI-HSDIO generates an error with an unknown status code when allocating streaming waveforms that are not 128-byte aligned. NI-HSDIO generates an error with an unknown status code when allocating streaming waveforms that are not 128-byte aligned. Workaround: When allocating streaming waveforms, ensure that they are 128-byte aligned.
| |||||
306813 Return | MAX might not display NI-HSDIO devices when deploying to a LabVIEW Real-Time target. NI-HSDIO devices might not be visible in Measurement & Automation Explorer (MAX) when deploying to a LabVIEW Real-Time target. Workaround: For more information, search ni.com/kb using the code 4RCAMBD4 or contact NI Technical Support at ni.com/support.
| |||||
306816 Return | Some Windows power management modes terminate in-progress NI-HSDIO operations. NI recommends that you disable Windows power management modes, such as sleep, standby, and hibernate when using this product. These modes terminate any NI-HSDIO operation in progress. Notice that in Windows 7/Vista or on a laptop computer, the default power management settings are likely to enable sleep mode. Workaround: Disable Windows power management modes. For more information, contact NI Technical Support at ni.com/support.
| |||||
306817 Return | When installing NI-HSDIO in silent mode from the command line, your computer automatically trusts NI software When you run setup.exe to install the software from the DVD, you are prompted to choose whether you want to always trust software from National Instruments. If you do not select this option, your installation may be interrupted by one or more Microsoft Windows security dialog boxes. However, if you run the installer in silent mode from the command line, you will not receive this prompt, and the installer will automatically install NI's Software Publisher Certificate to the your Windows Trusted Publisher Store. Workaround: If you do not want to make this change to your Windows Trusted Publisher Store, do not choose this installation method. For more information, contact NI Technical Support at ni.com/support.
| |||||
399545 Return | Cannot perform self-calibration in MAX on NI 6555/6556. If you try to self-calibrate an NI 6555/6556 device in MAX it returns error -200445. Self-calibration of NI 6555/6556 devices takes between 15 and 20 minutes, which is longer than the allotted time period for self-calibration in MAX. Workaround: Self-calibrate your NI 6555/6556 device using the NI-HSDIO self-calibration VIs or functions instead.
| |||||
307658 Return | In NI 6555/6556 devices, if both acquisition and generation sessions are open when you conduct self-calibration in one session, the data from the other session is invalid. Self-calibration of NI 6555/6556 devices uses both acquisition and generation sessions. Therefore, if you have both sessions open when you conduct self-calibration in one session, the data from the other session is invalid Workaround: Workaround: Complete all measurements before or after self-calibrating your NI 6555/6556.
| |||||
183465 Return | Using script triggers with Hardware Compare results in uncertain HW compare error locations Any script statement that relied on external events will add an unknown delay to the locations of all the HW compare error samples decteted. Examples of such statements are: WAIT UNTIL ScriptTriggerN REPEAT UNTIL ScriptTriggerN Because the HW compare engine is running while the script operation waits, it is impossible to HW compare errors to digital pattern locations. Workaround: If possible, do not rely on external triggers when the HW Compare engine is active.
| |||||
346946 Return | While using Scripts, reloading a different waveform with the same name as an existing waveform may generate an incorrect waveform To optimize performance and reduce memory usage, use scripting. If the waveforms are known before the script is written to memory, the behavior is as advertised. However, if a script is loaded, and then a waveform is modified (overwritten, deleted and then loaded with same name), the HSDIO card may or may not generate the correct waveform and never throws an error. Workaround: After changing the waveform, reload the script.
| |||||
393626 Return | When using scripting on the PXIe-6548, if you nest a repeat forever loop immediately inside another loop, the device may throw a Device Data Underflow error. For example, the following script will occasionally error: script script1 repeat until scripttrigger0 repeat forever generate wfm1 marker0(0) end repeat end repeat end script Workaround: To work around this issue, program a waveform prior to the inner loop. If you insert "generate wfm1 marker0(0)" before "repeat forever", this will be functionally identical and will prevent the issue from occurring.
| |||||
407117 Return | Multiple CVI examples will not compile on Windows XP becuase of file path length When using Windows XP, several NI-HSDIO CVI examples will fail to compile because the file path length is too long. Workaround: There are several options you can take if you want to run any examples that exhibit this issue: - Manually copy the example folder to a shorter file path location. - Rename the folder to a shorter name.
| |||||
407892 Return | Passing an invalid value for Number of Records to niHSDIO_ReadMultiRecord and niHSDIO_FetchMultiRecord entry points produces an error message that says to use -1, which is not supported. When you use an invalid value for the Number of Records parameter when calling any of the niHSDIO_ReadMultiRecord or niHSDIO_FetchMultiRecord functions, the error message suggests that -1 (negative 1) can be used to obtain all available records. Workaround: Passing -1 to these functions is only supported when using LabVIEW, because the LabVIEW entry points automatically allocate memory for the arrays, whereas the array must be allocated outside of the function when using other programming languages..
| |||||
409366 Return | C Examples that use NI-TClk fail in Visual Studio 2012 because they expect /SAFESEH to be enabled by default When attempting to compile an NI-HSDIO C example that uses NI-TClk using Visual Studio 2012, you may receive an error similar to the following: "niTClk.lib(niTClk.dll) : error LNK2026: module unsafe for SAFESEH image." This error is cause by Visual Studio 2012 having /SAFESEH (Safe Exception Handling) enabled by default, which niTClk.dll is not compiled with. For more information, consult the Microsoft documentation here: http://msdn.microsoft.com/en-us/library/9a89h429.aspx Workaround: To work around this issue, follow the steps described below. 1. Right-click on the Project and select Properties 2. Expand Configuration Properties -> Linker -> Advanced 3. Change "Image Has Safe Exception Handlers" to "No (/SAFESEH:NO)" 4. Click OK. - Try compiling again
|
Document last updated on 6/12/2013