Automatically detects the vCal ports connected to the VNA ports and writes the vCal Orientation property. Call this VI before calling RFmxVNA Calibration Initiate VI.

Auto detection of vCal orientation initially occurs at the midpoint of the frequency range specified for the calibration. If the connection is not detected, it retries at the lowest frequnecy in the configured range.

The auto detection may fail if the power level at the vCal module is too low. If it fails, use the vCal Orientation attribute to configure the orientation manually

Note You must select SParams or Wave measurement before calling this VI.


icon

Inputs/Outputs

  • cstr.png Selector String

    Selector String specifies a selector string comprising of the signal name. If you do not specify the signal name, the default signal instance is used. The default value is "" (empty string).

    Example:

    ""

    "signal::mysignal1"

    You can use the RFmxVNA Build Signal String VI to build the selector string.

  • cgenclassrntag.png Instrument Handle In

    Instrument Handle In specifies the RFmx session refnum.

  • cerrcodeclst.png error in (no error)

    error in describes error conditions that occur before this node runs.

  • igenclassrntag.png Instrument Handle Out

    Instrument Handle Out passes a reference of your RFmx session to the next VI.

  • ierrcodeclst.png error out

    error out returns error information.