Correction:Calibration:Calkit:Electronic:Orientation
- Updated2025-10-13
- 1 minute(s) read
Specifies the orientation of the vCal fixture ports with respect to vCal ports.
Use "portA:portname, portB:portname" format to specify vCal orientation.
When using NI-PXIe 5633 without an NI switch module, if vCal orientation is not specified, it is assumed to be "portA:port1, portB:port2".
When using NI-PXIe 5633 with an NI switch module, you must set vCal orientation for all the specified Calibration Ports.
The default value is an empty string.
Remarks
The following table lists the characteristics of this property.
| Short Name | vCal Orientation |
| Data type | ![]() |
| Permissions | Read/Write |
| High-level VIs | N/A |
| Resettable | Yes |