Use Measurement & Automation Explorer (MAX) to configure your NI hardware.

MAX informs other programs about which NI hardware products are in the system and how they are configured. MAX is automatically installed with FlexRIO.

Note MAX is not available on Linux.
  1. Launch MAX.
  2. In the configuration tree, expand Devices and Interfaces to see the list of installed NI hardware.
    Installed modules appear under the name of their associated chassis.
  3. Expand your Chassis tree item.
    MAX lists all modules installed in the chassis. Your default names may vary.
    Note If you do not see your module listed, press <F5> to refresh the list of installed modules. If the module is still not listed, power off the system, ensure the module is correctly installed, and restart.
  4. Record the name MAX assigns to the hardware. Use this identifier when programming the PXIe-6594.
  5. Self-test the hardware by selecting the item in the configuration tree and clicking Self-Test in the MAX toolbar.
    MAX self-test performs a basic verification of hardware resources.

What Should I Do if the PXIe-6594 Does Not Appear in MAX?

  1. In the MAX configuration tree, expand Devices and Interfaces.
  2. Expand the Chassis tree to see the list of installed hardware, and press <F5> to refresh the list.
  3. If the module is still not listed, power off the system, ensure that all hardware is correctly installed, and restart the system.
  4. Navigate to the Device Manager.
    Operating SystemDescription
    Windows 10/8.1 Right-click the Start button, and select Device Manager.
    Windows 7 Select Start » Control Panel » Device Manager.
  5. Verify the PXIe-6594 appears in the Device Manager.
    1. Under the NI RIO Devices section, confirm that a PXIe-6594 entry appears.
      Note Device Manager identifies the PXIe-6594 as NI FlexRIO PXIe Module (KU15P).
      Note If you are using a PC with a device for PXI remote control system, under System Devices, also confirm that no error conditions appear for the PCI-to-PCI Bridge.
    2. If error conditions appear, reinstall the FlexRIO driver and the PXIe-6594.

What Should I Do If the PXIe-6594 Fails the Self-Test in MAX?

  1. Reset the PXIe-6594 through MAX and then perform the self-test again.
  2. Restart the system, and then perform the self-test again.
  3. Power off the chassis.
  4. Reinstall the failed module in a different slot.
  5. Power on the chassis.
  6. Perform the self-test again.
    Note If the module fails the self-test again, contact NI or visit ni.com/support for further troubleshooting information.