Self-Calibrating the PXIe-5842 with the NI-RFSA LabVIEW Example
- Updated2025-07-31
- 1 minute(s) read
Note Ensure that any connected external
hardware is in an idle state and not transmitting during self-calibration. NI does not
recommended that you perform a self-calibration while connected in direct loopback
configurations.
-
Before you use the instrument or perform a self-calibration, you must allow the
instrument to warm up.
-
If your instrument does not have power, complete the following steps to provide
power for your specific instrument:
- PXIe-5842 VST with 54 GHz Frequency Extension: Provide power to the RMM-5585 remote measurement module using the power adapter, then power on the PXI Express chassis.
- All other PXIe-5842 instruments: Power on the PXI Express chassis.
- Wait for the operating system to load completely.
-
If your instrument supports multiple device personalities, configure the device
personality in Hardware Configuration Utility or MAX.
Note You must perform an individual self-calibration for each device personality.
- Allow the instrument to warm up for 30 minutes.
-
If your instrument does not have power, complete the following steps to provide
power for your specific instrument:
- Launch LabVIEW.
-
Launch the Example Finder at and navigate to the example.
- Select .
- Open the example.
-
Complete the following steps to configure the example:
- Select the device identifier assigned to the PXIe-5842 in MAX in the [Resource Name] drop-down menu.
- Set Clock Source to OnboardClock.
- Set Self Calibration Step Operations to Perform All Self Calibration Steps.
- Run the VI.