Self-Calibrating the PXIe-5842 with InstrumentStudio
- Updated2025-07-31
- 2 minute(s) read
NI recommends that you perform an initial self-calibration using InstrumentStudio.
NI VSTs are represented in InstrumentStudio as a vector signal analyzer and a vector signal generator. Performing self-calibration with either representation of the PXIe-5842 accomplishes the same self-calibration, so it is not necessary to use both when you self-calibrate the PXIe-5842.
Complete the following steps to self-calibrate your VST with InstrumentStudio.
Note Ensure that any connected external
hardware is in an idle state and not transmitting during self-calibration. NI does not
recommend that you perform a self-calibration while connected in direct loopback
configurations.
-
Before you use the instrument or perform a self-calibration, you must allow the
instrument to warm up.
-
If your instrument does not have power, complete the following steps to provide
power for your specific instrument:
- PXIe-5842 VST with 54 GHz Frequency Extension: Provide power to the RMM-5585 remote measurement module using the power adapter, then power on the PXI Express chassis.
- All other PXIe-5842 instruments: Power on the PXI Express chassis.
- Wait for the operating system to load completely.
-
If your instrument supports multiple device personalities, configure the device
personality in Hardware Configuration Utility or MAX.
Note You must perform an individual self-calibration for each device personality.
- Allow the instrument to warm up for 30 minutes.
-
If your instrument does not have power, complete the following steps to provide
power for your specific instrument:
- Launch InstrumentStudio by navigating to .
- To create an analyzer panel for the PXIe-5842, click .
-
Select the tab for the PXIe-5842 and click the gear icon
(
) in the upper right corner of the right sidebar.
- Hover over and click Self-Calibrate....
- Click Continue.