PXIe-4081 Overview
- Updated2025-07-08
- 3 minute(s) read
The PXIe-4081 is a 7½-digit, 1,000 V digital multimeter and digitizer for engineers building PXI systems that require high-precision multi-function measurements. Use the PXIe-4081 to make AC/DC voltage, AC/DC current, 2- or 4-wire resistance, and frequency/period measurements, as well as diode tests.
Device Capabilities
The PXIe-4081 has the following features and capabilities.
- 7½ digits at 10 S/s, 5½ digits at 3 kS/s
- 10 to 26-bit flexible resolution
- Voltage measurements up to 1000 V DC (700 VRMS)
- Current measurements up to 3 A
- Resistance measurements up to 5 GΩ
- Four high-voltage, safety-agency approved banana plug connectors
- Waveform acquisitions at sample rates up to 1.8 MS/s.
The following table lists the functions and ranges supported by the PXIe-4081 while in DMM mode:
DMM Function | Ranges |
---|---|
DC Voltage | 100 mV, 1 V, 10 V, 100 V, 1000 V |
AC Voltage | 50 mV, 500 mV, 5 V, 50 V, 700 V |
DC Current | 1 µA, 10 µA, 100 µA, 1 mA, 10 mA, 100 mA, 1 A, 3 A |
AC Current | 100 µA, 1 mA, 10 mA, 100 mA, 1 A, 3 A |
2-Wire Resistance | 100 Ω, 1 kΩ, 10 kΩ, 100 kΩ, 1 MΩ, 10 MΩ, 100 MΩ, 5 GΩ |
4-Wire Resistance | 100 Ω, 1 kΩ, 10 kΩ, 100 kΩ, 1 MΩ, 10 MΩ |
Diode | 10 V at 1 µA / 10 µA / 100 µA; 4.0 V at 1 mA |
Frequency/Period | 15 Hz - 500 kHz / 2 µs - 67 ms |
To measure DC or AC current outside of the normal range on the PXIe-4081, you must use a current shunt module (CSM). Refer to the Measuring Current section for more information. Frequency/period voltage ranges are the same as AC voltage ranges. |
The following table lists the functions and ranges supported for waveform acquisitions:
Waveform Function | Ranges | Coupling |
---|---|---|
Voltage Waveform | 100 mV, 1 V, 10 V, 100 V, 1000 V | AC or DC |
Current Waveform | 100 µA, 1 mA, 10 mA, 100 mA, 1 A, 3 A | DC |
To measure current waveform outside of the normal range on the PXIe-4081, you must use a current shunt module (CSM). Refer to the Measuring Current section for more information. |
The following table lists the available PXIe-4081 input resistance selections. NI-DMM calibrates the PXIe-4081 for full accuracy with all input resistances shown.
DCV and Voltage Waveform | ACV | ||
---|---|---|---|
Range | Input Resistance | Range | Input Resistance |
100 mV | 10 MΩ (default) or 10 GΩ | 50 mV | 10 MΩ |
1 V | 10 MΩ (default) or 10 GΩ | 500 mV | 10 MΩ |
10 V | 10 MΩ (default) or 10 GΩ | 5 V | 10 MΩ |
100 V | 10 MΩ (default) | 50 V | 10 MΩ |
1000 V | 10 MΩ (default) | 700 V | 10 MΩ |
Driver Support
NI recommends that you use the newest version of the driver for your module.
Driver Name | Earliest Version Support |
---|---|
NI-DMM | 15.2 |