Measuring Frequency and Periodic Signal Frequency
- Updated2023-12-14
- 2 minute(s) read
Measuring Frequency and Periodic Signal Frequency
Attach the test probes for measuring frequency and periodic signal frequencies as shown in the following figure:
About Frequency/Period Measurements
The DMM can measure periodic signal frequencies, up to 500 kHz, from a variety of signal sources, ranging from millivolts to hundreds of volts.
The DMM measures frequency by counting the number of zero-crossing rising edges of the input signal, using an onboard 108 MHz timebase.
Frequency (Hz) = timebase (Hz) x Ns / Nt
Period = 1 /Frequency (Hz)
where
timebase = 108 x 106 Hz
Ns = number of rising edges detected in the measurement window
Nt = number of timebase clock periods between the first and the last rising edge
To measure a signal of frequency f in Hz, the DMM needs a minimum aperture of (2/f) s. In other words, the DMM requires an aperture that is greater than or equal to at least two periods of the input signal. Therefore, as a general rule, the aperture should be set to a value greater than or equal to 110% of (2/f) s. The resolution of the frequency/period measurement can be improved by increasing the aperture time enough to allow for an integer increase in the number of periods.
Another way to measure frequency is to perform waveform acquisitions and then to use the signal processing functions in the ADE to extract the frequency.
Related Information
- Device Capabilities
The PXIe-4081 has the following features and capabilities.