PXIe-1085 Specifications
- Updated2025-11-25
- 9 minute(s) read
PXIe-1085 Specifications
PXIe-1085 Specifications
These specifications apply to the PXIe-1085 model.
Revision History
| Version | Date changed | Description |
|---|---|---|
| 376797B-01 | October 2025 | Added front and rear chassis views. |
| 376797A-01 | December 2016 | Initial release. |
| 373712G-01 | October 2015 | Original specifications document: NI PXIe-1085 Series User Manual. |
Looking For Something Else?
For information not found in the specifications for your product, such as operating instructions, browse Related Information.
Chassis Components
Use the diagrams to locate the key features of the PXIe-1085 chassis front and back panels.
- Chassis Carry Handle
- Backplane Connectors
- 10 MHz REF OUT SMA
- 10 MHz REF IN SMA
- PXI Filler Panels
- Chassis Model Name
- Removable Feet
- PXI Express Hybrid Peripheral Slots (16x)
- PXI Express System Timing Slot
- PXI Express System Controller Slot
- Temperature, Fan, and Power LEDs
- Power Inhibit Switch
- System Controller Expansion Slot
- Universal AC Input
- Push-Reset Circuit Breaker
- Remote Inhibit and Voltage Monitoring Connector
- Ethernet Port
- Inhibit Mode Selector Switch
- Fan Speed Selector Switch
- Chassis Ground Screw
- Power Supply Shuttle Mounting Screws (8x)
- Power Supply Shuttle Handle (2x)
- Power Supply Shuttle
- Fan Module (3x)
Electrical
The following section provides information about the PXIe-1085 AC input and DC output.
AC Input
| Input voltage range | 100 V AC to 240 V AC |
| Operating voltage range | 90 V AC to 264 V AC |
| Input current rating | 12 A to 6 A |
| Input frequency | 50/60 Hz |
| Operating frequency range | 47 Hz to 63 Hz |
| Over-current protection | 15 A circuit breaker |
| Efficiency | 70% typical |
| Power disconnect | The AC power cable provides main power disconnect. Do not position the equipment so that it is difficult to disconnect the power cord. The front-panel power switch causes the internal chassis power supply to provide DC power to the CompactPCI/PXI Express backplane. You also can use the rear-panel 8-pin connector and inhibit mode switch to control the internal chassis power supply. |
| 3.3 V | <±0.2% |
| 5 V | <±0.1% |
| ±12 V | <±0.1% |
DC Output
| Voltage | PXIe-1085 12 GB/s Maximum Current | PXIe-1085 24 GB/s Maximum Current |
|---|---|---|
| +3.3 V | 60 A | 60 A |
| +5 V | 44 A | 49 A |
| +12 V | 62 A | 62 A |
| -12 V | 4 A | 4 A |
| 5 VAUX | 2 A | 2 A |
| Slot | +5 V | V (I/O) | +3.3 V | +12 V | -12 V | 5 VAUX |
|---|---|---|---|---|---|---|
| System Controller Slot | 15 A | — | 15 A | 30 A | — | 1 A |
| System Timing Slot | — | — | 6 A | 4 A | — | 1 A |
| Hybrid Peripheral Slot with PXI-1 Peripheral | 6 A | 5 A | 6 A | 1 A | 1 A | — |
| Hybrid Peripheral Slot with PXI-5 Peripheral | — | — | 6 A | 4 A | — | 1 A |
| PXI-1 Peripheral Slot | 6 A | 11 A | 6 A | 1 A | 1 A | — |
| Voltage | Load Regulation |
|---|---|
| +3.3 V | <5% |
| +12 V | <5% |
| +5 V | <5% |
| -12 V | <5% |
| Voltage | Maximum Ripple and Noise |
|---|---|
| +3.3 V | 50 mVpp |
| +12 V | 50 mVpp |
| +5 V | 50 mVpp |
| -12 V | 50 mVpp |
| Over-current protection | All outputs protected from short circuit and overload with automatic recovery |
| Over-voltage protection (3.3 V and 5 V) | Clamped at 20% to 30% above nominal output voltage |
| Power supply shuttle MTTR | Replacement in under 5 minutes |
Remote Inhibit and Voltage Monitoring Connector
| VOH | 3.8 V (IOH = -8 mA) |
| VOL | 0.44 V (IOH = 8 mA) |
| VIH | 3.5 V (minimum) |
| VIL | 1.5 V (maximum) |
Chassis Cooling
| Module cooling system | Forced air circulation (positive pressurization) through three 169 cfm fans with High/Auto speed selector |
| Slot airflow direction | Bottom of module to top of module |
| Module cooling intake | Bottom rear of chassis |
| Module cooling exhaust | Along both sides and top of chassis |
| Power supply cooling system | Forced air circulation through two integrated fans |
| Power supply cooling intake | Right side of chassis |
| Power supply cooling exhaust | Left side of chassis |
Acoustic Emissions
Sound Pressure Level (at Operator Position)
| Auto fan (up to ~30 °C ambient) | 51.2 dBA |
| High fan | 64.1 dBA |
Sound Power
| Auto fan (up to ~30 °C ambient) | 60.8 dBA |
| High fan | 75.9 dBA |
Backplane
| Size | 3U-sized; one system slot (with three system expansion slots) and 17 peripheral slots. Compliant with IEEE 1101.10 mechanical packaging. PXI Express Specification compliant. Accepts both PXI Express and CompactPCI (PICMG 2.0 R 3.0) 3U modules. |
| Backplane bare-board material | UL 94 V-0 Recognized |
| Backplane connectors | Conforms to IEC 917 and IEC 1076-4-101, UL 94 V-0 rated |
10 MHz System Reference Clock: PXI_CLK10
| Maximum slot-to-slot skew | 500 ps |
| Accuracy | ±25 ppm maximum (guaranteed over the operating temperature range) |
| Maximum jitter | 5 ps RMS phase-jitter (10 Hz to 1 MHz range) |
| Duty-factor | 45% to 55% |
| Unloaded signal swing | 3.3 V ±0.3 V |
100 MHz System Reference Clock: PXIe_CLK100 and PXIe_SYNC100
| Maximum slot-to-slot skew | 100 ps |
| Accuracy | ±25 ppm max (guaranteed over the operating temperature range) |
| Maximum jitter | 3 ps RMS phase-jitter (10 Hz to 12 kHz range) 2 ps RMS phase-jitter (12 kHz to 20 MHz range) |
| Duty-factor for PXIe_CLK100 | 45% to 55% |
| Absolute differential voltage (When terminated with a 50 Ω load to 1.30 V or Thévenin equivalent) | 400 mV to 1000 mV |
External 10 MHz Reference Out
| Accuracy | ±25 ppm max (guaranteed over the operating temperature range) |
| Maximum jitter | 5 ps RMS phase-jitter (10 Hz to 1 MHz range) |
| Output amplitude | 1 VPP ±20% square-wave into 50 Ω, 2 VPP unloaded |
| Output impedance | 50 Ω ±5 Ω |
External Clock Source
| Frequency | 10 MHz ±100 ppm |
| Front panel SMA input amplitude | 200 mVPP to 5 VPP square-wave or sine-wave |
| System timing slot PXI_CLK10_IN input amplitude | 5 V or 3.3 V TTL signal |
| Front panel SMA input impedance | 50 Ω ±5 Ω |
| Maximum jitter introduced by backplane | 1 ps RMS phase-jitter (10 Hz to 1 MHz range) |
PXIe_SYNC_CTRL
| VIH | 2.0 V to 5.5 V |
| VIL | 0 V to 0.8 V |
PXI Star Trigger
| Maximum slot-to-slot skew | 250 ps |
| Backplane characteristic impedance | 65 Ω ±10% |
PXI Differential Star Triggers (PXIe-DSTARA, PXIe-DSTARB, PXIe-DSTARC)
| Maximum slot-to-slot skew | 150 ps |
| Maximum differential skew | 25 ps |
| Backplane differential impedance | 100 Ω ±10% |
Mechanical
| Height | 6.97 in. (177.1 mm) |
| Width | 18.30 in. (464.8 mm) |
| Depth | 19.38 in. (492.3 mm) |
| Weight | 40.3 lb (18.28 kg) |
| Chassis materials | Sheet Aluminum (5052-H32, 5754-H22), Extruded Aluminum (6063-T5, 6060-T6), Plate Aluminum (6063-T5, 6061-T6), Cold Rolled Steel, Cold Rolled Stainless Steel, Sheet Copper (C110), Santoprene, Urethane Foam, PC-ABS, Nylon, Polycarbonate, Delrin, Polyethylene, Polyamide (FR-106), Neodymium Magnet |
| Finish | Conductive Clear Iridite on Aluminum, Electroplated Nickel on Cold Rolled Steel, Electroplated Zinc on Cold Rolled Steel, Electroplated Nickel on Copper |
The following figures show the PXIe-1085 chassis dimensions. The holes shown are for the installation of the optional rack mount kits. You can install those kits on the front or rear of the chassis, depending on which end of the chassis you want to face toward the front of the instrument cabinet. Notice that the front and rear chassis mounting holes (size M4) are symmetrical.
- Front Rack Mount Kit
- PXIe-1085 Chassis
- Rear Rack Mount Kit
Environmental
| Maximum altitude | 2,000 m (800 mbar) (at 25 °C ambient) |
| Pollution Degree | 2 |
Indoor use only.
| Ambient temperature range | 0 °C to 55 °C |
| Relative humidity range | 10% to 90%, noncondensing |
| Ambient temperature range | -40 °C to 71 °C |
| Relative humidity range | 5% to 95%, noncondensing |
Shock and Vibration
| Operational shock | 30 g peak, half-sine, 11 ms pulse |
| Random vibration, operating | 5 Hz to 500 Hz, 0.3 grms |
Safety Compliance Standards
This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:
- IEC 61010-1, EN 61010-1
- UL 61010-1, CSA C22.2 No. 61010-1
Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:
- EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
- EN 55011 (CISPR 11): Group 1, Class A emissions
- EN 55022 (CISPR 22): Class A emissions
- EN 55024 (CISPR 24): Immunity
- AS/NZS CISPR 11: Group 1, Class A emissions
- AS/NZS CISPR 22: Class A emissions
- FCC 47 CFR Part 15B: Class A emissions
- ICES-001: Class A emissions
CE Compliance
This product meets the essential requirements of applicable European Directives, as follows:
- 2006/95/EC; Low-Voltage Directive (safety)
- 2014/30/EU; Electromagnetic Compatibility Directive (EMC)
Product Certifications and Declarations
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, search by model number, and click the appropriate link.
Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.
For additional environmental information, refer to the Engineering a Healthy Planet web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.