Digital filtering, enabled by default, prevents the switch module from being triggered by pulses that are less than 150 ns on PXI trigger lines 0–7. Some devices used with a switch may only be able to send triggers that are less than 150 ns. For the switch module to recognize these pulse widths, digital filtering must be disabled.
You can disable digital filtering by using the niSwitch Property
Node
or the
niSwitch_SetAttributeViBoolean
function. Complete the following steps to disable digital filtering.
Caution When digital filtering is disabled, it is possible for the switch
module to be triggered by noise on the PXI trigger lines. If the device triggering
the switch is capable of sending pulses greater than 150 ns, you should not disable
digital filtering.
- Add the niSwitch Property Node to the block diagram.
- Click the property node and select Module Characteristics»Digital Filter Enable.
- Right-click the Digital Filter Enable property and select Change to Write.
- Right-click the Digital Filter Enable property and create a constant.
- Set the constant to False.