Configuring Pneumatics
- Updated2022-04-20
- 1 minute(s) read
Configuring Pneumatics
The ECUTS does not ship ready for use with facility compressed air. To configure the ECUTS pneumatics to route compressed air, use the following guidelines:
- Use 6 mm outer diameter tubing to connect to the pneumatic air port on the back of the test rack.
- Mount pneumatic components such as regulators and valves on the bottom of the test or in empty test rack space.
- Do not remove the ECUTS ballast plates, if installed, from the bottom of the test rack. Removing the ballast plates may cause the ECUTS to become unstable.
- Connect to the pneumatic output ports using 6 mm outer diameter
tubing.Note Available pneumatic output ports on the mass interconnect are J9, J10, or J16 (if populated), in the pin 1 location.
The mating connection on the ITA is a VPC Pneumatic Contact 610132106, 610132110, 610132111, or 610132112 in the VPC Mini Power Module 510108115.