Mounting Requirements
- Updated2025-11-24
- 1 minute(s) read
The minimum spacing requirements for cooling and cabling clearance are listed in the following table. Ensure that your installation meets these requirements.
| Side | Horizontal | Vertical |
|---|---|---|
| Left | 25 mm | 25 mm |
| Right | 25 mm | 12 mm |
| Top | 25 mm | 25 mm |
| Bottom | 12 mm | 25 mm |
| Front | 76 mm | 76 mm |
| Rear | 0 mm | 0 mm |
Measure the ambient temperature at the horizontal and vertical orientation:
- Horizontal orientation: 25.4 mm (1.0 in.) from the center of each side panel.
- Vertical orientation: 25.4 mm (1.0 in.) from the center of the vent on the bottom of the BTS-16110.
Note Avoid operating BTS-16110 in environments with excessive airflow. Operating the
device at low temperatures, low input power voltages and excessive airflow may cause BTS-16110 to lose thermal regulation.
Note Supply BTS-16110 with 22.5 V DC or greater when operating at temperatures
below 20 °C (68 °F).