Benchmarking a System Built from the Static Structural Test Reference Architecture
- 更新时间2023-02-20
- 阅读时长3分钟
Benchmarking a System Built from the Static Structural Test Reference Architecture
To create performance benchmarks for the Static Structural Test Reference Architecture design pattern, NI built a physical test system according to the pattern and tested the system for viability and reliability, using the following tests.
Use the results of the benchmark system tests as reference when building your own system from the Static Structural Test Reference Architecture.
The Static Structural Test Benchmark System
The following figure depicts the composition of the Static Structural Test benchmark system built by NI.
- cDAQ-9189: 8-Slot, TSN-Enabled Ethernet CompactDAQ Chassis
- cRIO-9805: 4-Port, 802.1AS Ethernet Switch Expansion Module for CompactRIO
- Cisco RV340: Router
- Lenovo ThinkStation P330 Tower Gen 2: Server Computer
The 32 cDAQ-9189 8-Slot chassis in the benchmark system are configured with a total of 256 modules and a channel composition that represents a common static structural test application.
| Number and Type of Channels | Number of Modules | Module Models | Sample Rate |
|---|---|---|---|
| 1,944 quarter-bridge strain gauges | 243 | NI-9235 and NI-9236 | 100 S/s |
| 32 voltage | 8 | NI-9215 | 100 S/s |
| 64 temperature | 4 | NI-9213 | 1 S/s |
Conditions
The benchmark test results were found under the following conditions unless otherwise noted.
- Static Test Software Suite 1.2 installed
- Server computer: Lenovo ThinkStation P330 Tower Gen2
- Intel Xeon E-2286G CPU 4.0 GHz, 6-Core
- 64 GB RAM
- 1 TB SSD
- Gigabit Ethernet adapter (x2) (one dedicated for instrumentation network)
- Windows 10 64-bit
- External RAID drive: Oyen Digital 4TB SSD MiniPro RAID v3 (2 TB, RAID 1 configuration)
- External DHCP-enabled router: Cisco RV340 router
- Google Chrome browser
- cDAQ-9189 chassis with cDAQ firmware version 20.0 installed (available through NI-DAQmx cDAQ Firmware 21.0)
- System components connected according to the Static Structural Test Reference Architecture design pattern, as shown in the previous figure
Long-Term Test Results
The Long-Term Test assessed the performance of the benchmark system when continuously logging data without producing errors.
Conditions:
- 2-week test time
- 2,048 channel count
- 100 S/s sample rate for strain and voltage channels
- 1 S/s sample rate for thermocouple channels
| |||||||||||||||||
Channel Count Parametrized Test Results
The Channel Count Parametrized Test found the highest sample rates the benchmark system could achieve at different channel counts. The highest sample rates, reported in the following section, were achieved for 30 minutes without producing errors and within the following parameters:
- < 80% memory usage
- < 95% CPU usage
- < 80% network utility
| |||||||||||||||||||||||||||||||||||||||
Shunt Calibration and Null Offset Test Results
Time to complete shunt calibration on 1,944 channels | 5 minutes |
Time to complete null offset on 1,944 channels | 5 minutes |