BTS-16101 Specifications
- 업데이트 날짜:2024-12-11
- 25분 (읽기 시간)
Definitions
Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty. Warranted specifications account for measurement uncertainties, temperature drift, and aging. Warranted specifications are ensured by design or verified during production and calibration.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
- Typical specifications describe the performance met by a majority of models.
- Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
- Measured specifications describe the measured performance of a representative model.
Specifications are Typical unless otherwise noted.
Conditions
Specifications are valid for the system and all included instruments under the following conditions unless otherwise noted.
- Battery Test System environmental characteristics are met
- Battery Test System Software Suite 2.0 or later
- Connection to a computer with a monitor, keyboard, and mouse that satisfies the minimum requirements of the Battery Test System Software Suite
- Instrument-level conditions are met
Instrument performance is defined in the instrument specification document. Refer to the instrument specification document for the full set of specifications and information.
Hardware Reference
This section describes the location and function of hardware contained within the BTS-16101 measurement rack.
Measurement Rack Component Locations
Refer to the following figures to locate the BTS-16101 measurement rack components.
- Fan Panel
- Emergency Power Off (EPO) Panel
- 1U Open-Bottom Brush Panel Cable Entry
- Primary Drawer
- Secondary Drawer
- 1U Brush Cable Entry
- GPIB-RS232
- RMX-4101 Power Supplies
- Uninterruptible Power Supply (UPS)
- Industrial Casters
- Power Distribution Unit (PDU)
- Drawer Handles
- Auxiliary PDU
- PDU
- Remote EPO Connector Panel
- 3U Power Entry Panel (PEP)
- 4U Air Inlet Panel
- Drawer Slides
- Ethernet Switch
- RMX-4101 Power Supplies
- Transceiver Cable (TRC) Mounting Brackets
- Articulating Cable Arm
- Articulating Cable Arm
- Transceiver Cable (TRC) Mounting Brackets
- Drawer Slides
- Ethernet Switch
- RMX-4101 Power Supplies
Measurement Rack Component Features
Refer to the following topics for information on function of the components in the BTS-16101 measurement rack.
Power Entry Panel
| Figure Callout | Description | Use |
|---|---|---|
| 1 | Protective Earthing Ground Stud | Provides internal grounding plate connections with the external grounding lug for
all rack mount equipment. High levels of leakage current may be present on the
measurement rack. Connect the measurement rack to the protective earth terminal before
connecting to AC power. Refer to the
BTS-16101 Battery Test System Measurement Rack Safety,
Environmental, and Regulatory Information
document
for specifications for protective earth terminal wiring. Note The facility installation shall provide
a means for connection to protective earth, and qualified personnel shall install a
protective earthing conductor from the BTS-16101 protective earthing terminal to the protective earth wire in
the facility. |
| 2 | Power Inlet Connector | Provides a low power (IEC 60320 C20) power entry. |
| 3 | Main Breaker (CB Outlet) | Provides circuit protection as a main breaker by controlling whether AC power will be allowed into the rack. If the Main Breaker switch is off, the rack cannot be powered on. |
| 4 | Ethernet Port | Facilitates network connectivity to internal equipment. The Gigabit Ethernet port connects to the cRIO-9047 controller. |
| 5 | USB 3.0 Ports (x2) | Provides access to extension cables within the rack. The internal USB extension cables ship unterminated for user-customizable connection to internal equipment. Use the USB cables that are included in the kit for external connection. |
| 6 | DC Output to EPO Panel | Provides DC power and inhibit control to/from the EPO panel. |
| 7 | Output Connector | Provides AC output to the uninterruptible power supply (UPS). |
| 8 | Grounding Plate | Provides a connection point for the measurement rack ground. |
The PEP also includes a line filter for EMI protection within the system.
Emergency Power Off Panel
When a test system encounters a serious issue or an emergency is taking place in the facility, operators need the ability to quickly and cleanly power off the test system. The emergency power off (EPO) panel mechanisms are included on the measurement rack to simplify connectivity and inhibit power switching.
Operators can use the EPO to reset a system in an error state, prevent damage to a DUT, or prevent harm to themselves.
| Figure Callout | Description | Use |
|---|---|---|
| 1 | Main Power Switch | Provides the primary way of controlling whether the system is active. It toggles
between two positions—Standby ( |
| 2 | Temperature Controller (AutomationDirect SL4824-RR-D) | Measures temperature based on the location of the attached thermocouple. Notice When temperature setpoint
is reached, the temperature controller does not move the system into a shutdown
state. |
| 3 | Emergency Power Off Button | Provides a way to control whether the system is powered. The following behavior
occurs when you press the EPO button:
To release the EPO button after it is pressed, rotate it clockwise. |
Remote EPO Connector
The remote emergency power off (EPO) connector provides access to the power distribution unit (PDU) Disable loop. It also provides dry contacts for remote shutdown of external equipment. This connection allows immediate action to complete the following:
- When the EPO button on the measurement rack is depressed, the remote EPO connector switches all external equipment on the external safety loop to the EPO state.
- When an external instrument on the external safety loop enters an EPO state, the remote EPO connector switches the measurement rack into the Remote EPO state. For more information about this state, refer to Measurement Rack Power States.
When the external instrument exits its EPO state, the remote EPO connector switches the measurement rack out of the Remote EPO state into its previous state.
The Remote EPO connector has connections for the EPO button and the external safety loop.
| Terminal | Connection | Normal State | EPO State |
|---|---|---|---|
| 1 | External safety loop | Closed | Open |
| 2 | External safety loop | Closed | Open |
| 3 | Dry contact, EPO button | Closed (default) | Open |
| 4 | Dry contact, EPO button | Closed (default) | Open |
Use the following mating components to create interlock cabling for an external safety loop.
| Component | Manufacturer/Part Number |
|---|---|
| Plug assembly, size 11 reverse sex series 1, CPC (1 ea) | TE connectivity 206429-1 |
| Kit, cable clamp, standard shell sixe 11, CPC (1 ea) | TE connectivity 1-206062-4 |
| Strip, III+SKT, 18-16, 30AU/FL (4 ea) | TE connectivity 66099-4 |
Measurement Rack Drawers
The measurement rack features up to two sliding drawers which allow for easier cable routing and improves the utilization of space within the rack.
- Sliding Drawer
- PS-16 Power Supply
- cRIO-9805 Ethernet Switch
- Transceiver Cable Mounting Brackets (Optional)
- cRIO-9047 Controller
- C Series Modules (Optional)
- cDAQ1 (Optional)
- cDAQ2 (Optional)
- cDAQ3 (Optional)
- Terminal Block
- DC Power Fuse
- Articulating Cable Arm
The drawers contain the following features:
- Articulating arm that enables the full range of motion of the drawer and organizes cables from the drawer components to other components within the rack
- Wire ducts for easier cable routing and organization
- Optional mounting brackets for TRC-8543/8546 NI-XNET CAN/LIN transceiver cables that can be mounted in either drawer
- DC power fuse and terminal block
Cable Management
To reduce strain on the instrumentation connectors, NI recommends that you provide extra cable length or a service loop on the side of the instrument or at the tie-down points on the drawer front panel, as shown in the following figure.
Measurement Rack Power
This section covers the following power-related components and topics:
Measurement Rack Power States
| Power State | PEP | PDU | UPS | Temperature Controller |
|---|---|---|---|---|
| Off | Off | Off | Off | Off |
| Standby | On (AC) | Off | Off | On |
| EPO | DC only | Off | On | On |
| Remote EPO | On (AC) | Off | On | On |
| On | On (AC) | On | On | On |
| Power State | Description | Action |
|---|---|---|
| Off | The system is entirely disabled with no power passing through the line filter or any internal test system components. | To enter Off state:
|
| Standby | Power to the power distribution unit (PDU) and uninterruptible power supply (UPS) is shut off. Power is passing out of the PEP and into the thermostat emergency power off (EPO) panel. | To enter Standby state:
|
| EPO | Immediately cuts AC power from the PEP and PDU. The UPS remains powered so that the
measurement instruments in the rack continue logging data. DC power
passes out of the PEP and into the thermostat EPO panel. Entering this state also triggers the remote EPO connector, which, when connected to an external safety loop, automatically shuts off external equipment. |
To enter EPO state, press the EPO button. To exit EPO state, turn clockwise to release the EPO button. |
| Remote EPO | Immediately cuts power from the PDU. The UPS remains powered so that the measurement instruments in the rack continue logging data. Power passes out of the PEP and into the thermostat EPO panel. | The system enters Remote EPO state when other equipment
connected to an external safety loop through the remote EPO connector enter an EPO
state. The system exits Remote EPO state after the external instrument exits its EPO state. The rack switches into its previous state. |
| On | A change to this state begins the main power on sequence of the test system. The PDU receives AC power through the UPS from the PEP and enables outlets to other system equipment. The EPO panel is powered. | To enter the On state, follow the process in Powering on the System. |
Connecting Power and Powering Up
Power Distribution Unit
A power distribution unit (PDU) takes an input power signal and distributes it to several outlets that can power components of the system. These internal power outlets from the PDU have a rated voltage and current for both alternating and direct current.
The measurement rack features a single-phase PDU that supports global voltages (100 V to 240 V, 50 Hz to 60 Hz) and has a 20 A (IEC C19) input connector which cables directly to the uninterruptible power supply (UPS).
Refer to the MPD 41145X Family Product Manual, which ships with the Battery Test System, for more information about the PDU.
Uninterruptible Power Supply
Use the uninterruptible power supply (UPS) to power critical components in your system during power loss, brownouts, EPO state, and during normal operation.
The UPS delivers power with a dependable voltage and current supply. It acts as a battery power supply after a power outage or significant brownout. The UPS is available as the following model options:
- APC SRT 2200XLA—120 V
- APC SRT 2200XLI—240 V
Refer to the APC UPS Operation Manual, which ships on disc with the Battery Test System, for more information about the UPS.
Powering On/Off the UPS
Press the POWER ON/OFF button to power on the UPS. The button LED color indicates power status.
| LED State/Color | Status |
|---|---|
| Off | The UPS and the output power are off. |
| White | The UPS and the output power are on. |
| Red | The UPS is on and the output power is off. |
PS-16 Power Supply
The PS-16 is located in the primary drawer in the measurement rack and powers the following components:
- cRIO-9047 controller
- Measurement rack internal Ethernet switch
- cDAQ-9189 chassis
- cRIO-9805 Ethernet switches
- NI-XNET (CAN/LIN) C Series modules
- Serial C Series modules
The PS-16 can also power remote temperature input devices in your system.
DC Power Fuse
The DC power fuse in the primary and secondary drawer provide protection for your system components from short circuits. When a fuse opens, all instruments connected to the terminal block in a drawer—cRIO-9047, cDAQ-9189, all C Series modules, cRIO-9805, GPIB-RS232, and FieldDAQ devices—lose power.
The DC power fuse is a Schurter 8020.50755 A, 500 V AC(6.3×32 mm)fast-acting fuse.
Measurement Rack Ethernet Networking
The measurement rack contains up to three Ethernet switches to network the components in your system:
- Ethernet Switch—The 16-port Moxa Ethernet switch provides networking connection between the cRIO-9047 controller and RMX-4101 power supplies. Refer to the EDS-205A/208A Series Quick Installation Guide, which ships with the Battery Test System, for more information.
- cRIO-9805—Up to two, four-port TSN-enabled Ethernet switches provide time-sensitive networking (802.1AS) connection between the cRIO-9047 controller, cDAQ-9189 chassis, and FieldDAQ devices enabling sub-microsecond-level synchronization between measurement I/O devices.
Troubleshooting Network Issues
cRIO-9047 is Not Communicating with the Network
- Ensure that the Ethernet connections between the cRIO-9047 and the host computer, and between the host computer and the router are secure.
- Configure the IP and other network settings by completing the following
steps.
- Use the USB-to-Type-A cable (included in the Battery Test System kit) to connect the cRIO-9047 USB device port to a host computer. The USB driver creates a virtual network interface card and assigns an IP address to the cRIO-9047 in the format of 172.22.11.x.
- In Measurement & Automation Explorer (MAX), expand your system under Remote Systems.Note If you do not see the cRIO-9047 under Remote Systems, right-clickRemote Systems, and select Troubleshoot Remote System Discovery to walk through troubleshooting steps.
- Select the Network Settings tab to configure the IP and other network settings.
- (Optional) Use the standard measurement rack Ethernet port to reconnect
the cRIO-9047 to the host computer. The cRIO-9047 attempts to initiate a
DHCP network connection at power up.Note If the cRIO-9047 cannot obtain an IP address, it connects to the network with a link-local IP address with the form 169.254.x.x. The host computer communicates with the cRIO-9047 over a standard Ethernet connection.
- Temporarily disable any network firewalls or other security software.
Verify the cRIO-9047 controller IP configuration by completing the following steps:
- Put the cRIO-9047 in safe mode and enable the RS-232 serial port by holding down
the RESET button on the controller front panel for 5
seconds.
The STATUS LED starts blinking three times every few seconds. Refer to the cRIO-904x User Manual for more information about safe mode.
- Connect a monitor to the mini DisplayPort to view the IP address.Note The controller also broadcasts the IP address through the RS-232 serial port. Refer to the cRIO-904x User Manual for more information about serial port configuration on the cRIO-9047 controller.
- Set a new DHCP connection by holding the RESET button on the controller front
panel down for 5 seconds. The STATUS LED repeats the same behavior from Step
1.
If the cRIO-9047 fails to set a new DHCP address, it assigns itself a link-local IP address. If the DHCP connection is successful and appropriate for your application, skip to Step 6.
- In MAX, expand your system under Remote Systems.
- Select the Network Settings tab to configure the IP and other network settings.
- Reboot the cRIO-9047 by pressing the RESET button on the controller front panel.
cDAQ-9189 Chassis Disconnects from Network
If your cDAQ-9189 chassis becomes disconnected from the network, try the following solutions:
- After moving the chassis to a new network, NI-DAQmx may lose connection to the chassis. In this case, click Reconnect to provide NI-DAQmx with the new hostname or IP address.
- The cDAQ chassis icon indicates whether it is recognized and present on the network. If a connected chassis appears as disconnected in the configuration tree in MAX, select Self-Test or Reset Chassis. If successful, the chassis icon changes to blue/grey.
For additional troubleshooting resources for the cDAQ-9189 chassis, refer to Finding a Network DAQ Device in MAX in the Measurement & Automation Explorer Help for NI-DAQmx.
Measurement Rack Characteristics
Physical
| |||||||||
| |||||||||
The main breaker (CB Outlet) on the power entry panel PEP-116 is rated for an impact energy level of IK06 (1J), when tested with a direct vertical impact per IEC 61010-1, 3rd Ed., Table 15 and Clause 8.2. The main breaker should be guarded against impacts exceeding 1J.
AC Input
| |||||||
Maximum current | 16 A | ||||||
Frequency | 50/60 Hz | ||||||
Internal Power Distribution
| |||||||||||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||||||||||
Remote EPO Pins
During remote EPO installation, pin 1 and pin 2 must be shorted with dry contacts to enable power distribution unit outputs.
| |||||||
Environment
| |||||||
| |||||||
Pollution Degree | 2 | ||||||
Maximum altitude | 2,000 m (800 mbar) | ||||||
| |||||||
Protective Earthing
| |||||||||||||||
Measurement Rack I/O
The instrumentation and hardware inside the measurement rack provide additional accessible I/O. The available I/O depends on the measurement rack configuration.
DC Power Supply
The RMX-4101 power supply is installed in the measurement rack and provides power to the DUT or other external devices. The measurement rack can contain up to 12 RMX-4101 power supplies. Refer to the RMX-410x User Manual on ni.com/manuals for more information.
RMX-4101 Specifications
Maximum output voltage | 60 VDC |
Output current | 3.5 A |
Output power | 210 W |
Overvoltage trip point | 5 V to 66 V |
High Voltage Interlock Loop
Digital Input
The NI-9422 is a digital input module that processes high voltage interlock loop (HVIL) digital signals or external ECU digital signals sent to the measurement rack. Refer to the NI-9422 Operating Instructions and Specifications on ni.com/manuals for more information and specifications.
NI-9422 Specifications
Number of channels | 8 digital input channels | ||||||||||||||||||
Input type | Sinking/sourcing | ||||||||||||||||||
| |||||||||||||||||||
| |||||||||||||||||||
| |||||||||||||||||||
| |||||||||||||||||||
Pulse Width Modulation
The NI-9423 and NI-9437 are digital input modules that process HVIL pulse width modulation signals or external ECU digital signals sent to the measurement rack. Refer to the NI-9423 Datasheet and the NI-9437 Datasheet on ni.com/manuals for more information and specifications.
NI-9423 Specifications
Number of channels | 8 digital input channels | ||||||||||||||||||
Input type | Sinking | ||||||||||||||||||
| |||||||||||||||||||
Maximum input delay time | 1 μs | ||||||||||||||||||
| |||||||||||||||||||
NI-9437 Specifications
Number of channels | 8 digital input channels | ||||||||||||||||||||||||||||||
Input type | Sinking | ||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||
Maximum input delay time | 1 μs | ||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||
Current Input
The NI-9227 is an analog input module that processes HVIL current input sent to the measurement rack. Refer to the NI-9227 Operating Instructions and Specifications on ni.com/manuals for more information and specifications.
NI-9227 Specifications
Number of channels | 4 analog input channels | ||||||||||||||||||
ADC resolution | 24 bits | ||||||||||||||||||
Type of ADC | Delta-Sigma (with analog prefiltering) | ||||||||||||||||||
Sampling mode | Simultaneous | ||||||||||||||||||
| |||||||||||||||||||
| |||||||||||||||||||
| |||||||||||||||||||
Safe operating input range | 5 Arms | ||||||||||||||||||
| |||||||||||||||||||
High Voltage Interlock Loop Control
External Relay
The NI-9482 is a single pole single throw (SPST) electromechanical relay that controls the continuity of the high voltage interlock loop (HVIL) during test execution. Refer to the NI-9482 User Manual and Specifications on ni.com/manuals for more information and specifications.
NI-9482 Specifications
Number of channels | 4 electromechanical relay channels | ||||||||||||||||||||||||||||||||||
Relay type | SPST | ||||||||||||||||||||||||||||||||||
Power-on output state | Channels off | ||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||
Resistance per channel, channel on | 0.2 Ω | ||||||||||||||||||||||||||||||||||
Switching rate | 1 operation per second | ||||||||||||||||||||||||||||||||||
Relay release time | 10 ms maximum | ||||||||||||||||||||||||||||||||||
Relay operate time | 15 ms maximum | ||||||||||||||||||||||||||||||||||
Relay bounce time | 3 ms | ||||||||||||||||||||||||||||||||||
Channel A-to-channel B safety voltage | 250 Vrms maximum, Measurement Category II | ||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||
DUT Voltage Measurement
Module Voltage Measurement
The NI-9225 and NI-9228 are analog input modules that measure battery module voltage output by the DUT. Refer to the NI-9225 Operating Instructions and Specifications and the NI-9228 Datasheet on ni.com/manuals for more information and specifications.
NI-9225 Specifications
Number of channels | 3 analog input channels | ||||||||||||||||||
ADC resolution | 24 bits | ||||||||||||||||||
Type of ADC | Delta-Sigma (with analog prefiltering) | ||||||||||||||||||
Sampling mode | Simultaneous | ||||||||||||||||||
| |||||||||||||||||||
| |||||||||||||||||||
Typical operating voltage range | 300 Vrms | ||||||||||||||||||
| |||||||||||||||||||
NI-9228 Specifications
Number of channels | 8 analog input channels | ||||||||||||||||||||||||||||||||||
ADC resolution | 24 bits | ||||||||||||||||||||||||||||||||||
Type of ADC | Delta-Sigma (with analog prefiltering) | ||||||||||||||||||||||||||||||||||
Sampling mode | Simultaneous | ||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||
Cell Voltage Measurement
The NI-9224 is an analog input module that measures cell voltage output by the DUT up to 250 V. Refer to the NI-9224 Datasheet on ni.com/manuals for more information and specifications.
NI-9224 Specifications
Number of channels | 8 analog input channels | ||||||||||||||||||||||||||||||||||
ADC resolution | 24 bits | ||||||||||||||||||||||||||||||||||
Type of ADC | Delta-Sigma | ||||||||||||||||||||||||||||||||||
Sampling mode | Simultaneous | ||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||
Overvoltage protection, AI+ to AI- | 250 Vrms | ||||||||||||||||||||||||||||||||||
Input coupling | DC | ||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||
Digital Signal Input and Output
The NI-9422 is a digital input module that processes high voltage interlock loop (HVIL) digital signals or external ECU digital signals sent to the measurement rack. Refer to the NI-9422 Operating Instructions and Specifications on ni.com/manuals for more information and specifications.
NI-9422 Specifications
Number of channels | 8 digital input channels | ||||||||||||||||||
Input type | Sinking/sourcing | ||||||||||||||||||
| |||||||||||||||||||
| |||||||||||||||||||
| |||||||||||||||||||
| |||||||||||||||||||
The NI-9375 is a combination digital input, digital output module that processes external ECU signal inputs and ignition signal outputs during test execution. Refer to the NI-9375 Datasheet on ni.com/manuals for more information and specifications.
NI-9375 Specifications
Number of channels | 16 digital input channels 16 digital output channels | ||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||
The NI-9475 is a digital output module that generates static digital signals and pulse width modulated signals for DUT inputs. Refer to the NI-9475 Datasheet on ni.com/manuals for more information and specifications.
NI-9475 Specifications
Number of channels | 8 digital input channels | ||||||||||||
Output type | Sourcing | ||||||||||||
Power-on output state | Channels off | ||||||||||||
External power supply voltage range (VSUP) | 0 VDC to 60 VDC | ||||||||||||
Output impedance (R0) | 0.14 Ω maximum | ||||||||||||
Continuous output current (I0), per channel | 1 A maximum | ||||||||||||
Output voltage | VSUP - (I0 * R0) | ||||||||||||
| |||||||||||||
Output delay time (full load) | 1 μs maximum | ||||||||||||
VSUP-to-COM safety voltage | 60 VDC maximum, Measurement Category I | ||||||||||||
| |||||||||||||
Temperature Measurement
Cell Temperature Inputs
The NI-9212 is a channel-to-channel isolated thermocouple input module that measures the temperature of various points on the DUT during test execution. Refer to the NI-9212 and TB-9212 Datasheet on ni.com/manuals for more information and specifications.
NI-9212 Specifications
Number of channels | 8 isolated thermocouple channels 2 internal cold-junction compensation channels | ||||||||||||||||||||||||||||||||||
ADC resolution | 24 bits | ||||||||||||||||||||||||||||||||||
Type of ADC | Delta-Sigma | ||||||||||||||||||||||||||||||||||
Sampling mode | Simultaneous | ||||||||||||||||||||||||||||||||||
Voltage measurement range | 78.125 mV | ||||||||||||||||||||||||||||||||||
Temperature measurement ranges | Works over temperature ranges defined by NIST4 National Institute of Standards and Technology[4] (J, K, T, E, N, B, R, S thermocouple types) | ||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||
The NI-9213 is a high-density thermocouple input module that measures the temperature of various points on the DUT during test execution. Refer to the NI-9213 Datasheet on ni.com/manuals for more information and specifications.
NI-9213 Specifications
Number of channels | 16 thermocouple channels 1 internal autozero channel 1 internal cold-junction compensation channel | ||||||||||||||||||
ADC resolution | 24 bits | ||||||||||||||||||
Type of ADC | Delta-Sigma | ||||||||||||||||||
Sampling mode | Scanned | ||||||||||||||||||
Voltage measurement range | 78.125 mV | ||||||||||||||||||
Temperature measurement ranges | Works over temperature ranges defined by NIST[4] (J, K, T, E, N, B, R, S thermocouple types) | ||||||||||||||||||
| |||||||||||||||||||
| |||||||||||||||||||
Remote Temperature Measurement
The FD-11613 and FD-11614 are thermocouple devices that are installed inside the environmental chamber to measure temperatures within the chamber during test execution. Refer to the FD-11613 Specifications and FD-11614 Specifications on ni.com/manuals for more information and specifications.
FD-11613 Specifications
Number of channels | 8 isolated thermocouple channels 2 cold-junction compensation channels | ||||||||||||||||||||||||||||||||||
ADC resolution | 24 bits | ||||||||||||||||||||||||||||||||||
Type of ADC | Delta-Sigma | ||||||||||||||||||||||||||||||||||
Sampling mode | Simultaneous | ||||||||||||||||||||||||||||||||||
Voltage measurement range | 78.125 mV | ||||||||||||||||||||||||||||||||||
Temperature measurement ranges | Works over temperature ranges defined by NIST[] (J, K, T, E, N, B, R, S thermocouple types) | ||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||
Operating temperature | -40 °C to 85 °C (-40 °F to 185 °F) | ||||||||||||||||||||||||||||||||||
FD-11614 Specifications
Number of channels | 16 isolated thermocouple channels 4 cold-junction compensation channels | ||||||||||||||||||||||||||||||||||
ADC resolution | 24 bits | ||||||||||||||||||||||||||||||||||
Type of ADC | Delta-Sigma | ||||||||||||||||||||||||||||||||||
Sampling mode | Simultaneous | ||||||||||||||||||||||||||||||||||
Voltage measurement range | 78.125 mV | ||||||||||||||||||||||||||||||||||
Temperature measurement ranges | Works over temperature ranges defined by NIST[] (J, K, T, E, N, B, R, S thermocouple types) | ||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||
Operating temperature | -40 °C to 85 °C (-40 °F to 185 °F) | ||||||||||||||||||||||||||||||||||
DUT and Instrument Communication
The NI-9860 and NI-9870 facilitate communication of information between the DUT and instrumentation inside the measurement rack. Refer to the NI-9860 Getting Started Guide and NI-9870 Getting Started Guide on ni.com/manuals for more information and specifications.
NI-9860 Specifications
Communication protocol | CAN LIN | ||||||
Number of ports | 2 | ||||||
Connector type | NI-XNET hardware selectable interface port | ||||||
| |||||||
NI-9870 Specifications
Communication protocols | RS-232 RS-485 | ||||||
Number of ports | 4 | ||||||
Maximum baud rate | 921.6 kb/s | ||||||
Maximum cable length | 250 pF equivalent | ||||||
UART buffers per port | 64 B FIFO | ||||||
Data bits | 5, 6, 7, 8 | ||||||
Stop bits | 1, 1.5, 2 | ||||||
Flow control | XON/OFF, RTS/CTS, None | ||||||
| |||||||
The GPIB-RS232 is an instrument control device that provides GPIB-to-RS232 and RS232-to-GPIB protocol conversions. The GPIB-RS232 uses an RS-232 port on the cRIO-9047 controller. Refer to the NI GPIB-Serial Converter Specifications on ni.com/manuals for more information and specifications.
GPIB-RS232 Specifications
Voltage range | 9 VDC to 28 VDC |
Current consumption at 12 VDC | 300 mA typical, 800 mA maximum |
1 Refer to the NI PS-16 Power Supply User Manual and Specifications on ni.com/manuals for more information and specifications.
2 The combined current output from the AP6020 PDU is limited to 10 A.
3 The USB ports on the power distribution panel are not connected to internal hardware and are available for expansion.
4 National Institute of Standards and Technology