How Do I Verify That the Mixed Signal Oscilloscope Probe is Compensated Properly?
- Updated2023-09-07
- 1 minute(s) read
How Do I Verify That the Mixed Signal Oscilloscope Probe is Compensated Properly?
- If you are using a switchable 1x/10x scope probe, set the oscilloscope probe slide switch to 10x.
- Connect the oscilloscope probe to the compensation tabs noting which tab is signal and which is GND.
- In the VirtualBench application, set the probe attenuation to 10x and look at the Mixed Signal Oscilloscope waveform. You should see a square wave. Use Auto Setup if necessary.
- Using the adjustment tool (included in the VirtualBench oscilloscope probe kit), adjust the compensation screw that is in the housing of the oscilloscope probe until you see a perfectly square waveform without overshoot or undershoot.
Adjustment Tool

Your probe is now properly compensated.