SWB-2833 Specifications
- Updated2025-06-26
- 6 minute(s) read
SWB-2833 Specifications
About These Specifications
Specifications characterize the warranted performance of the instrument under the stated operating conditions. Data in this document are Specifications unless otherwise noted.
Typical Specifications are specifications met by the majority of the instrument under the stated operating conditions and are tested at 23 °C ambient temperature. Typical specifications are not warranted.
All voltages are specified in DC, ACpk, or a combination unless otherwise specified.
Clean devices and terminal blocks by brushing off light dust with a soft, nonmetallic brush. Remove other contaminants with a soft, lint-free, dampened cloth. Do not use detergent or chemical solvents. The unit must be completely dry and free from contaminants before returning to service.
Cautions
SWB-2833 Pinout
Input Characteristics
| |||||||||
Maximum switching current | 2.0 A (per channel) | ||||||||
Maximum carry current | 2.0 A (per channel) | ||||||||
Maximum switching power | 60 W, 62.5 VA (per channel) | ||||||||
Maximum switching power | 60 W (per crosspoint) | ||||||||
Simultaneous channels at maximum current | 4 | ||||||||
| |||||||||
Thermal EMF, typical | <10 μV | ||||||||
Bandwidth, typical (-3 dB, 50 Ω termination, column-row-column) | ≥10 MHz | ||||||||
| |||||||||
| |||||||||
Minimum switching load[1]1 The minimum switch load is not recommended for 2-wire resistance measurements. | 20 mV/10 mA | ||||||||
Analog bus line connections | AB <0...7> (8 Lines) | ||||||||
Dynamic Characteristics
Relay operate/release time, typical[2]2 Relay operate and release times depend on PC and PXI bus performance and application software. For more information about NI SwitchBlock relay operate times, visit ni.com/info and enter the Info Code exa9ee. | <5 ms |
Expected relay life, mechanical (no load) | 1 × 108 cycles | ||||||||||||||
| |||||||||||||||
Physical Characteristics
Relay type | Electromechanical, latching |
Relay contact material | Palladium-ruthenium, gold covered |
I/O connectors | 160 position, DIN |
Power requirement, carrier | 20 W at 5 V, 5 W at 3.3 V |
Dimensions (L × W × H) | 11.2 cm × 1.2 cm × 17.1 cm(4.4 in. × 0.5 in. × 6.7 in.) |
Weight | 373 g (13.2 oz) |
Environment
| Maximum altitude | 2,000 m (800 mbar) (at 25 °C ambient temperature) |
|---|---|
| Pollution Degree | 2 |
Indoor use only.
Operating Environment
| Ambient temperature range, operating | 0 °C to 55 °C |
|---|---|
| Relative humidity range, operating | 10% to 90%, noncondensing |
Storage Environment
| Ambient temperature range, storage | -20 °C to 71 °C |
|---|---|
| Relative humidity range, storage | 5% to 95%, noncondensing |
Shock and Vibration
| Operating shock | 30 g peak, half-sine, 11 ms pulse |
|---|---|
| Random vibration |
|
Compliance and Certifications
Safety Compliance Standards
This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:
- IEC 61010-1, EN 61010-1
- UL 61010-1, CSA C22.2 No. 61010-1
Electromagnetic Compatibility
- EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
- EN 55011 (CISPR 11): Group 1, Class A emissions
- EN 55022 (CISPR 22): Class A emissions
- EN 55024 (CISPR 24): Immunity
- AS/NZS CISPR 11: Group 1, Class A emissions
- AS/NZS CISPR 22: Class A emissions
- FCC 47 CFR Part 15B: Class A emissions
- ICES-001: Class A emissions
Product Certifications and Declarations
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, search by model number, and click the appropriate link.
Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.
For additional environmental information, refer to the Engineering a Healthy Planet web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.
EU and UK Customers
电子信息产品污染控制管理办法(中国RoHS)
Accessories
Refer to ni.com for more information about the following accessories.
| Accessory | Part number |
|---|---|
| SH160F-160M-NI SwitchBlock Cable | 153028-01 |
| NI TBX-2808 screw terminal accessory for NI SwitchBlock (unshielded) | 781420-08 |
1 The minimum switch load is not recommended for 2-wire resistance measurements.
2 Relay operate and release times depend on PC and PXI bus performance and application software. For more information about NI SwitchBlock relay operate times, visit ni.com/info and enter the Info Code exa9ee.