SET-2010 Specifications

These specifications apply to the SET-2010.

Notice To ensure the specified EMC performance, operate this product only with shielded cables and accessories on J1 and J2 ports/connections, HD-44 cable, Trigger in/out, and Ethernet.

Revision History

Version Date Changed Description
379186B-01 October 2025 Added Caution statements to Current Ratings tables.
379186A-01 July 2025 Initial release.

Looking For Something Else?

For information not found in the specifications for your product, such as operating instructions, browse Related Information.

Definitions

Warranted Specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.

Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

  • Typical—describes the performance met by a majority of models.
  • Nominal—describes an attribute that is based on design, conformance testing, or supplemental testing.
  • Measured—describes the measured performance of a representative model.
  • Values are Typical unless otherwise noted.

    Conditions

    Specifications are valid under the following conditions unless otherwise noted.
    • All voltages are relative to GND

    SET-2010 Pinouts

    Use the pinouts to connect to terminals on the SET-2010.

    Figure 1. SET-2010 J1 and J2 Connector Pinout (Front)

    J1 and J2 Connector Pinout (Front)

    Table 1. J1 and J2 Connector Pin Assignments
    Signal Description
    Px.y Line y in Port x
    GND Ground connection
    NC No connection
    ENx Enable
    REFx Reference
    Figure 2. SET-2010 XJ2 Connector Pinout (Rear)

    XJ2 Connector Pinout (Rear)

    Table 2. XJ2 Connector Pin Assignments
    Row E D C B A
    1 P0.1- / P4.1- P0.1+ / P4.1+ NC P0.0- / P4.0- P0.0+ / P4.0+
    2 P0.3- / P4.3- P0.3+ / P4.3+ NC P0.2- / P4.2- P0.2+ / P4.2+
    3 GND GND GND GND GND
    4 P1.1- / P5.1- P1.1+ / P5.1+ NC P1.0- / P5.0- P1.0+ / P5.0+
    5 P1.3- / P5.3- P1.3+ / P5.3+ NC P1.2- / P5.2- P1.2+ / P5.2+
    6 GND GND GND GND GND
    7 P2.1- / P6.1- P2.1+ / P6.1+ NC P2.0- / P6.0- P2.0+ / P6.0+
    8 P2.3- / P6.3- P2.3+ / P6.3+ NC P2.2- / P6.2- P2.2+ / P6.2+
    9 GND GND GND GND GND
    10 P3.1- / P7.1- P3.1+ / P7.1+ NC P3.0- / P7.0- P3.0+ / P7.0+
    11 P3.3- / P7.3- P3.3+ / P7.3+ NC P3.2- / P7.2- P3.2+ / P7.2+
    12 NC NC NC NC NC
    13 NC NC NC NC NC
    14 NC NC NC NC NC
    15 P4.1- / P0.1- P4.1+ / P0.1+ NC P4.0- / P0.0- P4.0+ / P0.0+
    16 P4.3- / P0.3- P4.3+ / P0.3+ NC P4.2- / P0.2- P4.2+ / P0.2+
    17 GND GND GND GND GND
    18 P5.1- / P1.1- P5.1+ / P1.1+ NC P5.0- / P1.0- P5.0+ / P1.0+
    19 P5.3- / P1.3- P5.3+ / P1.3+ NC P5.2- / P1.2- P5.2+ / P1.2+
    20 GND GND GND GND GND
    21 P6.1- / P2.1- P6.1+ / P2.1+ NC P6.0- / P2.0- P6.0+ / P2.0+
    22 P6.3- / P2.3- P6.3+ / P2.3+ NC P6.2- / P2.2- P6.2+ / P2.2+
    23 GND GND GND GND GND
    24 P7.1- / P3.1- P7.1+ / P3.1+ NC P7.0- / P3.0- P7.0+ / P3.0+
    25 P7.3- / P3.3- P7.3+ / P3.3+ NC P7.2- / P3.2- P7.2+ / P3.2+
    Table 3. XJ2 Connector Signal Descriptions
    Signal Description
    Px.y Line y in Port x
    GND Ground connection
    NC No connection
    Figure 3. SET-2010 XJ3 Connector Pinout (Rear)

    XJ3 Connector Pinout (Rear)

    Table 4. XJ3 Connector Pin Assignments
    Pins Signal Slot1/Slot2
    H V1+ Instrument_0+
    G V1- Instrument_0-
    F V2+ Instrument_1+
    E V2- Instrument_1-
    D V3+ Fault_A
    C V3- Fault_B
    B V4+ Fault_C
    A V4- Fault_D

    SET-2010 Physical Characteristics

    Table 5. Physical Characteristics
    Dimensions

    175 mm x 31 mm x 336 mm

    (6.89 in. x 1.19 in. x 13.21 in.)

    Weight

    382 g (13.47 oz)

    With two fault cards installed: 697 g (24.58)

    Front I/O Connector 2x female 44-pin high-density D-SUB with 4-40 UNC screw lock
    RTI Connectors 2 mm Hard Metric, per IEC 61076-101

    SET-2010 Electrical Characteristics

    Table 6. Maximum Initial Contact Resistance
    J1 to XJ2 500 mΩ
    J2 to XJ2 500 mΩ
    J1 to J2 500 mΩ
    Table 7. Expected Electrical Lifetime
    1 A, 30 VDC resistive 105 operations
    1.5 A, 30 VDC resistive 104 operations
    Table 8. Bandwidth
    -3 dB, 50 Ω Termination ≤ 20 MHz

    SET-2010 Power Specifications

    Table 9. SLSC Bus
    Voltage/Current Rating

    3.3 V at 50 mA

    24 V at 1.2 A

    Power Rating 29 W
    Table 10. Continuous Power Rating
    Maximum Allowed Continuous Power Dissipation (Limited by SLSC Chassis Specification) 50 W

    SET-2010 Safety Voltages

    Table 11. Rated Voltages
    Maximum Input Voltage, Any Pin ±60 VDC
    Maximum Switching Power, DC Resistive Load 60 W
    Caution Any external sources must be limited to not exceed these maximum rated currents.
    Attention Les sources externes doivent être limitées pour ne pas dépasser ces tensions nominales maximales.

    SET-2010 Current Ratings

    Table 12. Maximum Current Rating (Including Channels, Instrument Bus, and Signal Reference)
    Per Channel 1.5 A
    Sum of All Channels 48 A
    Caution The above table represents the maximum values for the SET-2010 channel current rating. Depending upon the Rear Transition Interface (RTI) that is in use, these values may be lower. Failure to use an appropriate rating can lead to electrical or fire-related injuries. For a final reference on this current rating, refer to the RTI documentation.
    Attention Le tableau ci-dessus présente les valeurs maximales du courant nominal de la voie SET-2010. Selon l’interface RTI utilisée, ces valeurs peuvent être inférieures. Ne pas utiliser un courant approprié peut entraîner des blessures liées aux chocs électriques ou des brûlures. Pour obtenir une référence finale sur le courant nominal, reportez-vous à la documentation RTI.
    Table 13. Fault Bus Maximum Current Rating (to Slot1 and Slot2 Plug-In Modules)
    Fault Bus Chassis Intake Temperature Maximum Current Rating
    Per fault bus signal (to Slot1 and Slot2 plug-in modules) 0 °C to 35 °C 8 A
    35 °C to 40 °C 6 A
    Per plug-in module (Slot1 or Slot2) 0 °C to 35 °C 16 A
    35 °C to 40 °C 12 A
    Caution The above table represents the default values for the SET-2010 fault bus current rating. Depending upon the plug-in module that is in use, these values may be lower. Failure to use an appropriate rating can lead to electrical or fire-related injuries. For a final reference on this current rating, refer to the plug-in module documentation.
    Attention Le tableau ci-dessus présente les valeurs par défaut du courant nominal du bus de défaillance SET-2010. Selon le module de plug-in utilisé, ces valeurs peuvent être inférieures. Ne pas utiliser un courant approprié peut entraîner des blessures liées aux chocs électriques ou des brûlures. Pour obtenir une référence finale sur le courant nominal, reportez-vous à la documentation du module de plug-in.
    Caution Any external sources must be limited to not exceed these maximum rated currents.
    Attention Les sources externes doivent être limitées pour ne pas dépasser ces tensions nominales maximales.

    SET-2010 Fuse Ratings

    Table 14. Fuse Designation
    Current Rating 3 A
    Voltage Rating 125 VAC, 125 VDC
    Size and Dimensions 7.4 mm x 3.1 mm
    Type Time-Lag T
    Breaking Capacity 100 A
    Recommended Replacement Schurter 3404.0118.22

    Measurement Category

    This product is rated for Measurement Category I (or other non-MAINS circuits).

    Caution Do not connect the product to signals or use for measurements within Measurement Categories II, III, or IV.
    Attention Ne pas connecter le produit à des signaux dans les catégories de mesure II, III ou IV et ne pas l'utiliser pour effectuer des mesures dans ces catégories.

    Measurement Category I (or other non-MAINS circuits) is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated low-voltage sources, and electronics.

    Note Measurement Categories CAT I and CAT O are equivalent. These test and measurement circuits are for other circuits not intended for direct connection to the MAINS building installations of Measurement Categories CAT II, CAT III, or CAT IV.

    Environmental Guidelines

    Notice Failure to follow the mounting instructions in the product documentation can cause temperature derating.
    Notice This product is intended for use in indoor applications only.

    Environmental Characteristics

    Table 15. Temperature
    Operating 0 °C to 40 °C, chassis intake
    Storage -40 °C to 85 °C
    Table 16. Humidity
    Operating 10% RH to 90% RH, noncondensing
    Storage 5% RH to 95% RH, noncondensing
    Table 17. Pollution Degree
    Pollution Degree 2
    Table 18. Maximum Altitude
    Maximum Altitude 2,000 m
    Table 19. Shock and Vibration
    Operating Vibration 5 Hz to 500 Hz, 0.3 g RMS
    Non-Operating Vibration 5 Hz to 500 Hz, 2.4 g RMS
    Operating Shock 30 g, half-sine, 11 ms pulse
    Non-Operating Shock 50 g, half-sine, 11 ms pulse