Specifies the Start Trigger type. Depending upon the value of this property, more properties may be needed to fully configure the trigger. To set this property, the NI-RFSG device must be in the Configuration state.
PXIe-5654/5654 with PXIe-5696: The Start Trigger is valid only with a timer-based list when RF list mode is enabled.
Supported Devices: PXIe-5644/5645/5646, PXIe-5654/5654 with PXIe-5696, PXI-5670/5671, PXIe-5672/5673/5673E, PXIe-5820/5830/5831/5832/5840/5841/5842/5860
Default Value: None
Related Topics
Remarks
The following table lists the characteristics of this property.
| Short Name | StartTrig.Type |
| Data type | ![]() |
| Permissions | Read/Write |
| High-level VIs | niRFSG Configure Trigger |
| Channel-based | No |
| Resettable | Yes |
| None | 0 | Generation starts immediately. No trigger is configured. |
| Digital Edge | 1 | Data operation does not start until a digital edge is detected. The source of the digital edge is specified in the Digital Edge Start Trigger Source property and the active edge is specified in Digital Edge Start Trigger Edge property. |
| Software | 2 | Data operation does not start until a software event occurs. You can create a software event by calling the niRFSG Send Software Edge Trigger VI with StartTrigger as the trigger name. |
| P2P Endpoint Fullness | 3 | Data operation does not start until the endpoint reaches a threshold specified in the P2P Endpoint Fullness Start Trigger Level property. |