RFmxVNA Calibration Save VI
- Updated2025-10-13
- 3 minute(s) read
Computes error-correction terms and saves the calset in memory. Call this VI after all calibration standards have been measured using RFmxVNA Calibration Acquire VI.
If you do not specify Calset Name input parameter, the calset is saved as default calset for the specified signal and is not available for use in other signals. If you specify non-empty Calset Name string, calset is saved as a named calset available for use across all signals and is also selected as the active calset for the current signal.

Inputs/Outputs
Selector String
—
Selector String specifies a selector string comprising of the signal name. If you do not specify the signal name, the default signal instance is used. The default value is "" (empty string). Example: "" "signal::mysignal1" You can use the RFmxVNA Build Signal String VI to build the selector string.
Instrument Handle In
—
Instrument Handle In specifies the RFmx session refnum.
Calset Name
—
Calset Name specifies the name of the calset. If you do not specify this parameter, the calset is saved as default calset for the specified signal and is not available for use in other signals. If you specify non-empty Calset Name string, calset is saved as a named calset available for use across all signals and is also selected as the active calset for the current signal.
error in (no error)
—
error in describes error conditions that occur before this node runs.
Instrument Handle Out
—
Instrument Handle Out passes a reference of your RFmx session to the next VI.
error out
—
error out returns error information. |
Selector String
—
Instrument Handle In
—
error in (no error)
—
Instrument Handle Out
—
error out
—