Creating Calkits for Calibration Methods
- Updated2025-10-08
- 1 minute(s) read
Creating Calkits for Calibration Methods
Depending on the calibration method, RFmx VNA calibration expects a combination of calibration elements having a specific Usage type defined in the calkit file.
The usage type of a calibration element is in the Type(s) column of the Calibration Element tab. Define the usage type by enabling one or more checkboxes in the Usage section of the calibration element configuration dialog.
The Usage area includes the following options.
The following table summarizes which calibration element usage types are expected by a particular RFmx VNA calibration method.
| Usage | SOLT | Overdetermined SOL | TRL / LRL | ||
|---|---|---|---|---|---|
| Known Thru | Unknown Thru (SOLR) | Known Thru | Unknown Thru (SOLR) | ||
| Short | 1 | 1 | |||
| Open | 1 | 1 | |||
| Load | 1 | 1 | |||
| Thru | 1 | 1 | 1 | ||
| Line | N (N≥1) | ||||
| Reflect | 1 | ||||
| Termination | N (N≥3) | N (N≥3) | |||
Note Although SOLR (SOLT using unknown Thru) does not require the definition of a Thru
calibration element in the Calkit, the connection of a physical Thru (the
unknown Thru) is required during calibration procedure.