PXIe-5633 Specifications

PXIe-5633 Specifications

These specifications apply to the PXIe-5633 when used as a standalone instrument.

Revision History

Version Date changed Description
379153A-01 April 2025 Initial release of document.

Looking For Something Else?

For information not found in the specifications for your product, such as operating instructions, browse Related Information.

Definitions

Warranted Specifications describe the performance of a model under stated operating conditions and are covered by the model warranty. Specifications account for measurement uncertainties, temperature drift, and aging. Specifications are ensured by design or verified during production and calibration.

Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

  • Typical—describes the performance met by a majority of models.
  • Typical-95—describes the performance met by 95% (≈2σ) of models with a 95% confidence.
  • Nominal—describes an attribute that is based on design, conformance testing, or supplemental testing.
  • Measured—describes the measured performance of a representative model.
  • Values are Warranted Specifications unless otherwise noted.

    Conditions

    All specifications are valid under the following conditions unless otherwise noted.

    • 30 minutes warm-up time; warm-up time begins when the PXI Express chassis has been powered on and the operating system has completely loaded
    • Calibration cycle is maintained
    • Environment temperature is within the ambient range
    • Installed in chassis with 82 W slot cooling capacity with fan mode set to Auto
    • Empty chassis slots contain slot blockers and EMC filler panels to minimize temperature drift and reduce emissions

    Warranted specifications are valid under the following condition unless otherwise noted.

    • Over an ambient temperature range of 0 °C to 40 °C

    Typical and Typical-95 specifications are valid under the following condition unless otherwise noted.

    • Over an ambient temperature range of 23 °C ±5 °C

    Typical specifications do not include measurement uncertainty.

    Measured specifications do not include measurement uncertainty and are measured immediately after a device self-calibration is performed.

    PXIe-5633 Front Panel and LEDs

    Figure 1. PXIe-5633 Front Panel


    Table 1. PXIe-5633 Connector Descriptions
    Connector Connector Type Description
    PORT 1 3.5 mm (f) Port 1 for VNA measurements; input/output terminal for RF signals for VST measurements when integrated with VST
    SWITCH IN 3.5 mm (f) Intermodule connection only (when integrated with VST)
    IF OUT 3.5 mm (f) Reserved
    SWITCH OUT 3.5 mm (f) Intermodule connection only (when integrated with VST)
    IF IN 3.5 mm (f) Reserved
    PORT 2 3.5 mm (f) Port 2 for VNA measurements; input/output terminal for RF signals for VST measurements when integrated with VST
    LO IN MMPX (f) Reserved
    LO OUT MMPX (f) Reserved
    REF IN MMPX (f) Reserved
    REF OUT MMPX (f) Reserved
    PFI 0 MMPX (f) Reserved
    PFI 1 MMPX (f) Reserved
    Table 2. PXIe-5633 LED Indicators
    LED Indications
    ACC

    Access: Indicates the basic hardware status of the module.

  • Off—The module is not yet functional or has detected a problem with a PXI Express power rail
  • Amber—The module is being accessed; accessed means that you are writing to the module setup registers to control the module, reading from the module to monitor its status, or transferring data to or from the module
  • Green—The module is controllable through the software
  • ACTIVE

    Indicates the state of the module.

  • Off—The module is idle
  • Dim amber—The module is armed and waiting for an acquisition trigger
  • Solid green—The module has received an acquisition trigger and is making a measurement
  • Solid red—The module has detected an error; the LED remains red until the error condition is removed

    Error conditions include loss of PLL lock and thermal shutdown.

  • Note Indicators are listed in increasing order of priority.

    PXIe-5633 Physical Characteristics

    Table 3. PXIe-5633 Physical Characteristics
    Dimensions 3U, 1 slot

    21.1 cm × 12.9 cm × 2.03 cm

    (8.3 in. × 5.1 in. × 0.8 in.)

    Note For more information, visit ni.com/dimensions and search by module number.
    Weight 555 g (19.6 oz)

    VNA Frequency Specifications

    VNA Frequency Range

    Table 4. PXIe-5633 Frequency Range
    8 GHz frequency range option 50 MHz to 8 GHz
    12 GHz frequency range option 50 MHz to 12 GHz
    18 GHz frequency range option 50 MHz to 18 GHz
    26.5 GHz frequency range option 50 MHz to 26.5 GHz

    VNA Internal Frequency Reference Accuracy

    Note The internal clock uses the PXIe_CLK100 signal as a reference, typically 25 ppm. Refer to your chassis specifications to determine the reference clock frequency accuracy. When using a PXIe-1095 chassis with the Timing and Synchronization OCXO upgrade option as the external reference, the frequency accuracy is ±80 × 10-9.

    VNA Frequency Resolution

    Table 5. PXIe-5633 Frequency Resolution
    Resolution 1 Hz

    VNA IF Bandwidth and Miscellaneous

    Table 6. PXIe-5633 IF Bandwidth
    Minimum VNA IF bandwidth 1 Hz
    Maximum VNA IF bandwidth 15 MHz
    Available IF bandwidth values include 10 MHz, 15 MHz, and [1, 2, 3, 5, or 7] × 10n Hz for integers n between 0 and 6.
    Table 7. Miscellaneous
    Number of points 1 to 100,001

    VNA System Specifications

    VNA Dynamic Range

    Table 8. PXIe-5633 Dynamic Range
    Frequency Typical (dB)
    100 MHz to 300 MHz 134
    >300 MHz to 6 GHz 146
    >6 GHz to 18 GHz 137
    >18 GHz to 22 GHz 134
    >22 GHz to 26.5 GHz 125
    Dynamic range specifies the difference between the maximum power and the noise floor, excluding crosstalk. It is calculated from the VNA Generation Maximum Output Power specification and the VNA Receiver Noise Floor specification, normalized to a 10 Hz IF Bandwidth.

    Conditions: 10 Hz IF Bandwidth.

    VNA Effective Residual Error Terms

    Table 9. PXIe-5633 Effective Residual Error Terms
    Frequency Directivity (dB) Source Match (dB) Load Match (dB) Reflection Tracking (dB) Transmission Tracking (dB)
    100 MHz to 300 MHz 46 41 40 0.07 0.07
    >300 MHz to 6 GHz 46 41 41 0.05 0.05
    >6 GHz to 8 GHz 45 42 42 0.05 0.06
    >8 GHz to 12 GHz 44 41 40 0.07 0.07
    >12 GHz to 18 GHz 44 40 39 0.09 0.09
    >18 GHz to 20 GHz 44 40 39 0.08 0.08
    >20 GHz to 26.5 GHz 41 35 34 0.12 0.12
    Conditions:
    • PXIe-5633 calibrated using CAL-5501 vector calibration module and instrument temperature is within 1 °C of calibration temperature.
    • Frequency List, Power Level, and IF Bandwidth are set identically across calibration and measurement.
    • Calibration performed with 10 Hz IF Bandwidth and -10 dBm Power Level at 23 ± 5 °C ambient temperature.

    VNA Corrected Measurement Performance

    Table 10. PXIe-5633 Reflection Measurement Uncertainty, Warranted
    Frequency Logarithmic Linear
    Reflection Level (dB) Magnitude (dB) Phase (Degrees) Reflection Range (dB) Magnitude (Linear)
    100 MHz to 12 GHz 0 0.20 1.6 0 to -3 0.023
    100 MHz to 12 GHz -3 0.20 1.6 -3 to -6 0.016
    100 MHz to 12 GHz -6 0.21 1.6 -6 to -15 0.012
    100 MHz to 12 GHz -15 0.41 3.0 -15 to -25 0.009
    100 MHz to 12 GHz -25 1.21 8.3 -25 0.008
    >12 GHz to 20 GHz 0 0.31 2.5 0 to -3 0.035
    >12 GHz to 20 GHz -3 0.31 2.5 -3 to -6 0.023
    >12 GHz to 20 GHz -6 0.31 2.5 -6 to -15 0.017
    >12 GHz to 20 GHz -15 0.52 3.9 -15 to -25 0.011
    >12 GHz to 20 GHz -25 1.45 10.1 -25 0.010
    >20 GHz to 26.5 GHz 0 0.38 2.9 0 to -3 0.043
    >20 GHz to 26.5 GHz -3 0.38 2.9 -3 to -6 0.029
    >20 GHz to 26.5 GHz -6 0.38 2.9 -6 to -15 0.021
    >20 GHz to 26.5 GHz -15 0.61 4.4 -15 to -25 0.013
    >20 GHz to 26.5 GHz -25 1.69 11.3 -25 0.011
    Conditions:
    • This specification is derived using VNA receiver specifications and residual error terms when PXIe-5633 is calibrated using CAL-5501 vector calibration module.
    • Analysis is performed with DUT S11 = S22, S21 = S12 = 0 (lin), 10 Hz IF Bandwidth, 0 dB Test Attenuation, -10 dBm Source Power Level during calibration and measurement.
    • Frequency List is identical between calibration and measurement.
    • Ambient temperature is 23 ± 5 °C and board temperature is ±1 °C from calibration.
    Figure 2. PXIe-5633 Reflection Uncertainty Linear Magnitude, Typical


    Figure 3. PXIe-5633 Reflection Uncertainty Log Magnitude, Typical


    Figure 4. PXIe-5633 Reflection Uncertainty Phase, Typical


    Figure 5. PXIe-5633 Reflection Uncertainty Log Magnitude Phase, Typical


    Table 11. PXIe-5633 Transmission Measurement Uncertainty, Warranted
    Frequency Transmission Coefficient (dB) Magnitude (dB) Phase (Degrees)
    100 MHz to 12 GHz 5 to 0 0.12 1.1
    100 MHz to 12 GHz 0 to -10 0.12 1.1
    100 MHz to 12 GHz -10 to -20 0.14 1.2
    100 MHz to 12 GHz -20 to -30 0.27 2.1
    100 MHz to 12 GHz -30 to -40 0.72 4.9
    >12 GHz to 20 GHz 5 to 0 0.16 1.6
    >12 GHz to 20 GHz 0 to -10 0.16 1.6
    >12 GHz to 20 GHz -10 to -20 0.16 1.6
    >12 GHz to 20 GHz -20 to -30 0.29 2.3
    >12 GHz to 20 GHz -30 to -40 0.73 5.1
    >20 GHz to 23 GHz 0 to -10 0.21 1.8
    >20 GHz to 23 GHz -10 to -20 0.22 1.9
    >20 GHz to 23 GHz -20 to -30 0.35 2.6
    >20 GHz to 23 GHz -30 to -40 0.84 5.7
    >23 GHz to 26.5 GHz 0 to -10 0.21 1.8
    >23 GHz to 26.5 GHz -10 to -20 0.28 2.1
    >23 GHz to 26.5 GHz -20 to -30 0.66 4.5
    Conditions:
    • This specification is derived using VNA receiver specifications and residual error terms when a PXIe-5633 is calibrated using CAL-5501 vector calibration module.
    • Analysis is performed with DUT S12 = S21 (for Transmission Coefficient ≤ 0 dB), S12 = 1/S21 (for Transmission Coefficient > 0 dB, S11 = S22 = 0 (lin), 10 Hz IF Bandwidth, 0 dB Test Attenuation, -10 dBm Source Power Level during calibration and measurement.
    • Frequency List is identical between calibration and measurement.
    • Ambient temperature is 23 ± 5 °C and board temperature is ±1 °C from calibration.
    Figure 6. PXIe-5633 Transmission Uncertainty Magnitude, Typical


    Figure 7. PXIe-5633 Transmission Uncertainty Phase, Typical


    VNA Measurement Speed

    Table 12. PXIe-5633 Measurement Speed (ms), Measured
    Frequency Sweep Range IF Bandwidth Number of Points
    Corrected Two-Port Measurement Uncorrected One-Port Measurement
    51 201 401 1601 51 201 401 1601
    50 MHz to 26.5 GHz 10 kHz 21.1 78.6 155 605 10.9 39.5 77.5 306
    50 MHz to 26.5 GHz 100 kHz 9.9 34.2 66.5 252 5.2 17.2 33.1 129
    50 MHz to 26.5 GHz 1 MHz 7.7 25.4 48.9 182 4.0 12.7 24.3 93.9
    50 MHz to 26.5 GHz 15 MHz 6.5 20.4 38.4 138 3.5 10.1 19.0 71.9
    Conditions:
    • Measured using a PXIe-1095 chassis and a PXIe-8881 embedded controller with Windows 10 64-bit, Xeon 8-core processor, and 16 GB DDR4 RAM configuration.
    • Measurement speed is defined as the time between when RFmx VNA Initiate is called and when the specified measurements have been returned with corrections applied by an RFmx VNA fetch function.
    • Each sweep is run n times and the times for n - 1 iterations are averaged (the first iteration is dropped).
    • For two-port corrected measurements, S21 is enabled and fetched.
    • One-port uncorrected measurements are for one sweep direction.
    • Auto IF Bandwidth Scaling Enabled is set to False.
    • InstrumentStudio debug mode is disabled.

    VNA Receiver Specifications

    VNA Noise Floor

    Table 13. PXIe-5633 Noise Floor (dBm/Hz)
    Frequency Warranted Typical
    100 MHz to 300 MHz -107 -126
    >300 MHz to 6 GHz -131 -138
    >6 GHz to 12 GHz -125 -132
    >12 GHz to 22 GHz -127 -134
    >22 GHz to 26.5 GHz -125 -130
    Conditions: Test receiver power measured with VNA source turned off, high-reflection terminations at the port, 30 kHz IF Bandwidth, 0 dB Test Receiver Attenuation, 100 averages, excluding spurs, normalized to 1 Hz bandwidth, ambient temperature 23 ± 5 °C.

    VNA Trace Noise

    Table 14. PXIe-5633 Trace Noise Magnitude (dB RMS)
    Frequency Warranted Typical
    50 MHz to <6 GHz 0.003 0.0015
    6 GHz to 26.5 GHz 0.006 0.0020
    Conditions:
    • 0 dBm Power Level, 10 kHz IF Bandwidth, 0 dB Test Receiver Attenuation, ambient temperature 23 °C ± 5 °C.
    • Measured using a high-reflection termination at the port and then calculating the uncorrected S11 magnitude stability across 201 iterations in 100 MHz increments.
    Table 15. PXIe-5633 Trace Noise Phase (Degrees RMS)
    Frequency Warranted Typical
    50 MHz to 300 MHz 0.03 0.007
    >300 MHz to 1 GHz 0.03 0.006
    >1 GHz to 3 GHz 0.03 0.005
    >3 GHz to <6 GHz 0.03 0.004
    6 GHz to 26.5 GHz 0.06 0.008
    Conditions:
    • 0 dBm Power Level, 10 kHz IF Bandwidth, 0 dB Test Receiver Attenuation, ambient temperature 23 °C ± 5 °C.
    • Measured using a high-reflection termination at the port and then calculating the uncorrected S11 phase stability across 201 iterations in 100 MHz increments.
    Table 16. PXIe-5633 Measurement Stability, Typical
    Frequency (Hz) Magnitude (dB/°C) Phase (deg/°C)
    100 MHz to 6 GHz 0.015 0.10
    >6 GHz to 8 GHz 0.025 0.40
    >8 GHz to 15 GHz 0.015 0.35
    >15 GHz to 22 GHz 0.030 0.55
    >22 GHz to 26.5 GHz 0.050 0.50
    Conditions:
    • 0 dBm Power Level, 100 Hz IF Bandwidth, 0 dB Test Receiver Attenuation, ambient temperature 23 °C ± 5 °C.
    • Ratioed measurement stability when measuring a low loss through or a high reflect termination.
    • Firmware version: 1.0.2f4

    VNA Non-Linearity

    Table 17. PXIe-5633 Receiver Compression Level for 0.1 dB Compression, Measured
    Frequency Compression Level (dBm)
    >100 MHz to 1 GHz 9
    >1 GHz to 3 GHz 8
    >3 GHz to 6 GHz 10
    >6 GHz to 22 GHz 10
    >22 GHz to 26.5 GHz 10
    Conditions: 0 dB Test Receiver Attenuation. Measured P0.1 dB or ADC saturation.

    Test Receiver Attenuation is configurable from 0 dB to 22 dB and is recommended for use when input signal power is above the 0.1 dB compression level.

    Table 18. PXIe-5633 Linearity
    Frequency Power Level (dBm) Magnitude (dB) Phase (Degrees)
    Warranted Typical Warranted Typical
    100 MHz to 1 GHz -5 to -22.5 0.12 0.04 0.80 0.27
    100 MHz to 1 GHz -30 0.16 0.08 1.06 0.53
    100 MHz to 1 GHz -35 0.20 0.10 1.32 0.66
    100 MHz to 1 GHz -40 0.34 0.16 2.25 1.06
    >1 GHz to 20 GHz -5 to -22.5 0.11 0.03 0.73 0.20
    >1 GHz to 20 GHz -30 0.13 0.06 0.86 0.40
    >1 GHz to 20 GHz -35 0.23 0.11 1.52 0.73
    >1 GHz to 20 GHz -40 0.35 0.16 2.31 1.06
    >20 GHz to 23 GHz -10 to -22.5 0.11 0.06 0.73 0.40
    >20 GHz to 23 GHz -30 0.14 0.07 0.93 0.47
    >20 GHz to 23 GHz -35 0.25 0.13 1.65 0.86
    >20 GHz to 23 GHz -40 0.38 0.20 2.51 1.32
    >23 GHz to 26.5 GHz -10 to -22.5 0.12 0.06 0.80 0.40
    >23 GHz to 26.5 GHz -30 0.28 0.17 1.85 1.13
    >23 GHz to 26.5 GHz -35 0.50 0.27 3.30 1.79
    >23 GHz to 26.5 GHz -40 0.87 0.55 5.75 3.63
    Conditions: Accuracy of test port receiver measured relative to -10 dBm input port power with 0 dB Test Receiver Attenuation, 10 Hz IF Bandwidth at 23 ± 5 °C ambient temperature.

    VNA Uncorrected Error Terms

    Table 19. PXIe-5633 Uncorrected Error Terms, Typical
    Frequency Directivity (dB) Source Match (dB) Load Match (dB)
    50 MHz to 300 MHz 24 23 16
    >300 MHz to 1 GHz 20 18 18
    >1 GHz to 3 GHz 22 16 14
    >3 GHz to 6 GHz 21 17 15
    >6 GHz to 8 GHz 17 19 18
    >8 GHz to 12 GHz 13 14 12
    >12 GHz to 18 GHz 12 10 9
    >18 GHz to 22 GHz 12 11 9
    >22 GHz to 26.5 GHz 10 9 8
    Conditions: Measured at VNA ports, -10 dBm Power Level, 10 kHz IF Bandwidth, 0 dB Test Receiver Attenuation, ambient temperature 23 ± 5 °C.

    VNA Source Specifications

    VNA Output Amplitude Range

    Table 20. PXIe-5633 Maximum and Minimum Output Power (dBm), Typical
    Frequency Maximum Minimum
    50 MHz to 6 GHz 18 -36
    >6 GHz to 8 GHz 16 -38
    >8 GHz to 12 GHz 15 -39
    >12 GHz to 18 GHz 13 -39
    >18 GHz to 22 GHz 10 -43
    >22 GHz to 26.5 GHz 5 -44
    Conditions: Measured at the VNA ports, 23 ± 5 °C ambient temperature.
    Table 21. PXIe-5633 Amplitude Resolution
    Amplitude resolution ≤0.05 dB, nominal

    VNA Output Amplitude Accuracy

    Table 22. PXIe-5633 Output Amplitude Accuracy (dB), Typical
    Frequency Amplitude Accuracy
    50 MHz to 1 GHz 1.10
    >1 GHz to 12 GHz 0.90
    >12 GHz to 22 GHz 0.90
    >22 GHz to 26.5 GHz 1.10
    Conditions:
    • Measured at VNA ports from -30 dBm to Maximum Output Power defined in VNA Output Amplitude Range specification.
    • Ambient temperature 23 ± 5 °C.
    Figure 8. PXIe-5633 Output Amplitude Accuracy


    Conditions: Factory calibration; ambient temperature 23 ± 5 °C.

    VNA Output Harmonics

    Table 23. PXIe-5633 Output Harmonics (dBc), Measured
    Frequency Harmonic Level
    >300 MHz to 1 GHz -34
    >1 GHz to 3 GHz -37
    >3 GHz to 6 GHz -26
    >6 GHz to 8 GHz -26
    >8 GHz to 12 GHz -26
    Conditions:
    • 0 dBm power level.
    • Includes second and third harmonic content up to 26.5 GHz.
    • Excludes sub-harmonic and non-harmonic spurs.

    VNA Pulse Generation

    Table 24. VNA Pulse Timing
    Minimum pulse width 200 ns, nominal
    Minimum pulse period 400 ns, nominal
    Maximum pulse period 10 s, nominal
    Table 25. VNA Pulse On/Off Ratio
    Frequency On/Off Ratio
    50 MHz to 26.5 GHz 90 dB, measured
    Conditions: -20 dBm Source Power, 10 averages.
    Figure 9. VNA Pulse Modulation Shape, Measured


    Conditions: 500 kHz IF Bandwidth, 100 µs Pulse Width, 50 µs Pulse Delay, 16 averages, measured in 500 MHz steps between 1 GHz and 26.5 GHz.

    Safety Voltages

    Connect only voltages that are below these limits.

    PORT 1, PORT 2

    Table 26. Absolute Maximum Input Power for VNA Load Port
    Test Receiver Attenuation < 8 dB +20 dBm
    Test Receiver Attenuation8 dB +27 dBm
    Table 27. Absolute Maximum Reverse Power for VNA Reflection Port
    Notice In certain conditions, such as generating into a short, reflections of signals that otherwise comply with these limits may exceed these values and damage the instrument.
    RF Output Power+10 dBm, Test Receiver Attenuation < 8 dB +15 dBm
    RF Output Power+20 dBm, Test Receiver Attenuation8 dB +20 dBm
    Table 28. Absolute Maximum Input Power for VST Passthrough* When Operating as Vector Signal Analyzer

    *When integrated with a supported VST.

    Reference Level20 dBm Reference Level + 5 dB
    Reference Level > 20 dBm
    Note Derate by 2 dB when source match is worse than -6 dB.
    • 0.3 GHz to 2 GHz: +25 dBm (CW RMS)
    • >2 GHz to 8 GHz: +26 dBm (CW RMS)
    • >8 GHz to 26.5 GHz: +27 dBm (CW RMS)
    Table 29. Absolute Maximum Reverse Power for VST Passthrough* When Operating as Vector Signal Generator

    *When integrated with a supported VST.

    Output Power+16 dBm +25 dBm
    Output Power+20 dBm +20 dBm

    Other Ports

    Table 30. Other Ports
    IF IN absolute maximum input power +18 dBm
    IF OUT absolute maximum reverse power +18 dBm
    SWITCH IN absolute maximum input power +14 dBm
    SWITCH OUT absolute maximum reverse power +10 dBm
    LO IN absolute maximum input power +12 dBm
    LO OUT absolute maximum reverse power +10 dBm
    REF IN maximum input voltage 5 Vpk-pk
    REF OUT absolute maximum reverse voltage 2 Vpk-pk
    PFI 0, PFI 1 absolute maximum input range -0.5 V to 5 V
    Table 31. Measurement Category
    Measurement Category CAT I/O

    Measurement Category

    Caution Do not connect the product to signals or use for measurements within Measurement Categories II, III, or IV.
    Attention Ne pas connecter le produit à des signaux dans les catégories de mesure II, III ou IV et ne pas l'utiliser pour effectuer des mesures dans ces catégories.

    Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated low-voltage sources, and electronics.

    Note Measurement Categories CAT I and CAT O are equivalent. These test and measurement circuits are for other circuits not intended for direct connection to the MAINS building installations of Measurement Categories CAT II, CAT III, or CAT IV.

    Current Ratings

    Table 32. Maximum Continuous Sinking Current, PFI 0 and PFI 1
    Per channel 45 mA
    Sum of all channels 90 mA
    Table 33. Maximum Continuous Sourcing Current, PFI 0 and PFI 1
    Per channel 66 mA
    Sum of all channels 132 mA

    PXIe-5633 Power Requirements

    Table 34. Power Requirements, Nominal
    +3.3 V DC 0.66 A (2.19 W)
    +12 V DC 4.99 A (59.88 W)
    Total power 62.07 W

    Environmental Guidelines

    Notice Failure to follow the mounting instructions in the product documentation can cause temperature derating.
    Notice This product is intended for use in indoor applications only.

    Environmental Characteristics

    Table 35. Temperature
    Operating0 °C to 40 °C
    Storage-40 °C to 71 °C
    Table 36. Humidity
    Operating10% RH to 90% RH, noncondensing
    Storage5% RH to 95% RH, noncondensing
    Table 37. Pollution Degree
    Pollution degree2
    Table 38. Maximum Altitude
    Maximum altitude2,000 m (800 mbar) (at 25 °C ambient temperature)
    Table 39. Shock and Vibration
    Operating vibration 5 Hz to 500 Hz, 0.3 g RMS
    Non-operating vibration 5 Hz to 500 Hz, 2.4 g RMS
    Operating shock30 g, half-sine, 11 ms pulse

    Calibration

    Table 40. PXIe-5633 External Calibration Interval
    Interval 1 year