Self-Calibrating the PXIe-5108 in MAX
- Updated2025-11-26
- 1 minute(s) read
Performing self-calibration adjusts the PXIe-5108 for variations in the environment. The PXIe-5108 devices are externally calibrated at the factory, however you should perform a complete self-calibration after installation.
-
Install the PXIe-5108 and let it warm up for the
recommended warm-up time listed in the .
Note Warm up begins when the PXI chassis has been powered on and the operating system has completely loaded.
-
Self-calibrate the PXIe-5108 by clicking the
Self-Calibrate button in MAX or calling niScope Cal Self
Calibrate (niScope_CalSelfCalibrate).
Note Low energy transients can appear at the output terminals of your PXIe-5108 during certain situations, such as power-up, power-down, device driver loading, and self-calibration.