PXIe-4151 Specifications
- Updated2025-11-20
- 12 minute(s) read
PXIe-4151 Specifications
These specifications apply to the PXIe-4151.
Revision History
| Version | Date changed | Description |
|---|---|---|
| 379000D-01 | October 2025 | Fixed an issue with maximum local sense voltage values. |
| 379000C-01 | May 2025 | Fixed an issue with the pinouts. |
| 379000B-01 | May 2025 | Added pinouts. |
| 379000A-01 | November 2023 | Initial release. |
Looking For Something Else?
For information not found in the specifications for your product, such as operating instructions, browse Related Information.
Definitions
Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
- Typical specifications describe the performance met by a majority of models.
- Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
Specifications are Warranted unless otherwise noted.
Conditions
Specifications are valid under the following conditions unless otherwise noted:
- Ambient temperature[1]1 The ambient temperature of a PXI system is defined as the temperature at the chassis fan inlet (air intake). of 23 °C ± 5 ºC.
- Chassis with ≥38
W slot cooling capacity.[2]2 For increased capability, NI recommends installing the
PXIe-4151 in a chassis with slot
cooling capacity ≥58 W.
- For chassis with slot cooling capacity = 38 W, fan speed set to HIGH
- Calibration interval of 2 years.
- Warm-up time of 30 minutes.
- Self-calibration performed within the last 24 hours.
- NI-DCPower Aperture Time is set to 1 power-line cycle (PLC).
Pinout
Use the pinouts to connect to terminals on the PXIe-4151.
Output Connector
| Pin | Description |
|---|---|
| HI terminal | HI force terminal connected to channel power stage (generates and/or dissipates power). Positive polarity is defined as voltage measured on HI > LO. |
| LO terminal | LO force terminal connected to channel power stage (generates and/or dissipates power). Positive polarity is defined as voltage measured on HI > LO. |
| HI S terminal | Voltage remote sense input terminals. Used to compensate for I x R voltage drops in cable leads, connectors, and switches. |
| LO S terminal | |
| GND ( |
Tied to chassis ground through module front panel. Use for connections to cable shields. Do not use for grounding the LO force terminal due to pin current rating. |
AUX I/O Connector
| Pin | Descriptions | |
|---|---|---|
| 1 | PFI 0 | Programmable Function Interface for digital timing and triggering. |
| 2 | PFI 1 | |
| 3 | FAULT | Device fault status output. |
| 4 | INHIBIT | Forces the power supply output to an inhibited state when not shorted to ground. You can short this pin to ground using the pre-installed jumper to enable normal operation. |
| 5 | GND | Ground reference for AUX I/O. |
| 6 | PFI 2 | Programmable Function Interface for digital timing and triggering. |
| 7 | GND | Ground reference for AUX I/O. |
| 8 | GND | |
AUX Power In Connector
| Signal | Descriptions |
|---|---|
| + | Positive Auxiliary Power Input |
| - | Negative Auxiliary Power Input |
Power Requirements
| PXIe Chassis, +3.3 V | 1 A, maximum |
| PXIe Chassis, +12 V | 1 A, maximum |
| Auxiliary power source input | 45.6 VDC to 50.4 VDC, 9 A, maximum |
Physical Characteristics
| Dimensions | 3U, two-slot, PXI Express/CompactPCI Express module 4.0 cm × 13.0 cm × 21.6 cm (1.6 in. × 5.1 in. × 8.5 in.) |
| Weight | 650 g (22.9 oz) |
| Output | OMNIMATE Hybrid, 7.62 mm (4 position), 2.54 mm (6 position) |
| AUX I/O | MICRO COMBICON - DFMC 0, 5, 2.54 mm (8 position) |
| AUX Power In | MEGA-FIT RECEPTACLE, 5.7 mm, (4 position) |
Instrument Capabilities
| DC voltage ranges | 6 V, 20 V |
| DC current ranges | 100 mA, 1 A, 25 A |
| Available DC output power [3]3 Power limit defined by voltage measured between HI and LO terminals. | 300 W |
Voltage
| Range | Resolution (Noise Limited) | Noise (0.1 Hz to 10 Hz, peak-to-peak, typical) |
Accuracy ± (% of Voltage + Offset)[4]4 Refer to the Remote Sense and Load Regulation sections for additional accuracy derating and conditions. Tambient 23 °C ± 5 °C, Tcal[5]5 Tcal is the internal device temperature recorded by the PXIe-4151 at the completion of the last self-calibration. ± 5 °C |
Tempco[6]6 Temperature coefficient applies beyond 23 °C ± 5 °C ambient within ± 5 °C of Tcal. ± (% of Voltage + Offset)/°C Tambient 0 °C to 55 °C, Tcal ± 5 °C |
|
|---|---|---|---|---|---|
| 6 V | 1 μV | 10 μV | 0.022% + 500 μV | 0.001% + 25 μV | |
| 20 V | 10 μV | 40 μV | 0.024% + 2 mV | 0.001% + 65 μV | |
Current
| Range | Resolution (Noise Limited) | Noise (0.1 Hz to 10 Hz, peak-to-peak, typical) |
Accuracy ± (% of Current + Offset)[7]7 Refer to the Remote Sense and Load Regulation sections for additional accuracy derating and conditions. Tambient 23 °C ± 5 °C, Tcal[8]8 Tcal is the internal device temperature recorded by the PXIe-4151 at the completion of the last self-calibration. ± 5 °C |
Tempco[9]9 Temperature coefficient applies beyond 23 °C ± 5 °C ambient within ± 5 °C of Tcal. ± (% of Current + Offset)/°C Tambient 0 °C to 55 °C, Tcal ± 5 °C |
|
|---|---|---|---|---|---|
| 100 mA | 10 nA | 200 nA | 0.03% + 30 μA | 0.0015% + 1.5 μA | |
| 1 A | 100 nA | 2 μA | 0.03% + 300 μA | 0.0015% + 15 μA | |
| 25 A | 10 μA | 160 μA | 0.065% + 2 mA | 0.0035% + 375 μA | |
Noise
| Wideband source noise[10]10 10 Hz to 20 MHZ bandwidth, measured with 1 µF load capacitance at the end of 1 m shielded cable. PXIe-4151 configured for normal transient response. | <20 mV peak-to-peak, typical |
The following figures illustrate measurement noise as a function of measurement aperture time for the PXIe-4151.
Settling Time and Transient Response
| Voltage mode, ≤18 V step, unloaded | <4.2 ms, typical |
| Current mode, full-scale step, 25 A range | <600 μs, typical |
| Current mode, full-scale step, 1 A range | <320 μs, typical |
| Current mode, full-scale step, 100 mA range | <1.2 ms, typical |
| Current Range | Value |
|---|---|
| 25 A | <250 μs, typical to recover within ±300 mV |
| 1 A | <50 μs, typical to recover within ±10 mV |
Remote Sense
| Voltage accuracy | Add 10 µV per volt of lead drop per ohm of corresponding sense lead resistance to voltage accuracy specifications. |
| Maximum sense lead resistance | 100 Ω |
| Maximum lead drop per lead | 1 V, maximum 20 V between HI and LO terminals |
Load Regulation
Auxiliary Power Input Line Regulation
| Voltage | Error included in accuracy specifications. |
| Current | Error included in accuracy specifications. |
Isolation
| Isolation voltage, any channel I/O pin[12]12 Channel I/O pins include the HI, LO, HI Sense and LO Sense terminals. The Shield GND terminals of the output connector are connected to chassis ground and not considered channel I/O pins. to earth ground | 150 V DC, CAT I |
| Transient overvoltage | 800 V peak |
Protection
General Module Protection
| Between any channel I/O pin[13]13 Channel I/O pins include the HI, LO, HI Sense and LO Sense terminals. The Shield GND terminals of the output connector are connected to chassis ground and not considered channel I/O pins. and LO | 60 V |
| Any AUX I/O pin to ground | 5 V |
| AUX Power In + to AUX Power In – | ±60 V |
| AUX Power In – to ground | ±1 V |
| Overcurrent or overvoltage | Automatic shutdown, output disconnect relay opens |
| Reverse voltage protection | Automatic shutdown, output disconnect relay opens |
| Down programmer overload | Automatic shutdown, output disconnect relay opens |
| Overtemperature | Automatic shutdown, output disconnect relay opens |
| Overvoltage or undervoltage | Automatic shutdown, output disconnect relay opens |
| Overcurrent | Automatic shutdown, output disconnect relay opens, non-accessible fuse as backup protection |
Measurement and Update Timing
| Available sample rates[14]14 When sourcing while measuring, both the Source Delay and Aperture Time affect the sampling rate. When taking a measure record, only the Aperture Time affects the sampling rate. |
(1.8 MS/s)/N, nominal where
|
| Sample rate accuracy | Equal to PXIe_CLK100 accuracy, nominal |
| Maximum measure rate to host | 1.8 MS/s per channel, continuous, nominal |
| Maximum source update rate[15]15 As the Source Delay is adjusted, or if advanced sequencing is used, maximum update rates vary. | 100,000 updates/s, nominal |
| Source event delay | 10 μs, nominal |
| Source event jitter | 2 μs peak-to-peak, nominal |
| Measure event jitter | 2 μs peak-to-peak, nominal |
Triggers
Input Triggers
| Types | Start, Source, Sequence Advance, Measure |
| Polarity | Active high (not configurable) |
| Minimum pulse width | 100 ns |
| Polarity | Active high (not configurable) |
| Pulse width | >200 ns |
Output Triggers (Events)
| Types | Source Complete, Sequence Iteration Complete, Sequence Engine Done, Measure Complete |
| Polarity | Active high (not configurable) |
| Pulse width | >230 ns |
AUX I/O (Fault)
| Direction | Output |
| Logic type | 3.3 V CMOS (open drain, internal pull-up) |
| Polarity | Active low (not configurable) |
| Output voltage range | 0 V to 3.3 V, typical |
| Output impedance (active low) | 50 Ω, nominal |
| Internal pull-up resistance | 10 kΩ, nominal |
Environmental Characteristics
| Characteristic | Specification |
|---|---|
| Operating temperature | 0 °C to 55 °C[16]16 Not all chassis can achieve this ambient temperature range. Refer to PXI chassis specifications to determine the ambient temperature ranges your chassis can achieve., [17]17 Maximum ambient temperature deratings apply when operating the 25 A range above certain current levels. Refer to the Instrument Capabilities section for additional details. |
| Storage temperature | −40 °C to 71 °C |
| Specification | Value |
|---|---|
| Pollution Degree | 2 |
| Specification | Value |
|---|---|
| Maximum altitude | 2000 m |
| Specification | Value |
|---|---|
| Operating Vibration | 5 Hz to 500 Hz, 0.3 g RMS |
| Non-Operating Vibration | 5 Hz to 500 Hz, 2.4 g RMS |
| Operating Shock | 30 g, half-sine, 11 ms pulse |
Calibration Interval
You can obtain the calibration certificate and information about calibration services for the PXIe-4151 at ni.com/calibration.
| Calibration interval | 2 years |
1 The ambient temperature of a PXI system is defined as the temperature at the chassis fan inlet (air intake).
2 For increased capability, NI recommends installing the PXIe-4151 in a chassis with slot cooling capacity ≥58 W.
3 Power limit defined by voltage measured between HI and LO terminals.
4 Refer to the Remote Sense and Load Regulation sections for additional accuracy derating and conditions.
5 Tcal is the internal device temperature recorded by the PXIe-4151 at the completion of the last self-calibration.
6 Temperature coefficient applies beyond 23 °C ± 5 °C ambient within ± 5 °C of Tcal.
7 Refer to the Remote Sense and Load Regulation sections for additional accuracy derating and conditions.
8 Tcal is the internal device temperature recorded by the PXIe-4151 at the completion of the last self-calibration.
9 Temperature coefficient applies beyond 23 °C ± 5 °C ambient within ± 5 °C of Tcal.
10 10 Hz to 20 MHZ bandwidth, measured with 1 µF load capacitance at the end of 1 m shielded cable. PXIe-4151 configured for normal transient response.
11 Measured between HI and LO terminals at the module output connector.
12 Channel I/O pins include the HI, LO, HI Sense and LO Sense terminals. The Shield GND terminals of the output connector are connected to chassis ground and not considered channel I/O pins.
13 Channel I/O pins include the HI, LO, HI Sense and LO Sense terminals. The Shield GND terminals of the output connector are connected to chassis ground and not considered channel I/O pins.
14 When sourcing while measuring, both the Source Delay and Aperture Time affect the sampling rate. When taking a measure record, only the Aperture Time affects the sampling rate.
15 As the Source Delay is adjusted, or if advanced sequencing is used, maximum update rates vary.
16 Not all chassis can achieve this ambient temperature range. Refer to PXI chassis specifications to determine the ambient temperature ranges your chassis can achieve.
17 Maximum ambient temperature deratings apply when operating the 25 A range above certain current levels. Refer to the Instrument Capabilities section for additional details.