1. Reset the PXIe-4150 through Hardware Configuration Utility or MAX and then perform the self-test again.
  2. Confirm that 48 V auxiliary power is connected to the PXIe-4150 and powered on, reset the PXIe-4150, and then perform the self-test again.
  3. Restart the system, and then perform the self-test again.
  4. Power off the chassis.
  5. Reinstall the failed module in a different slot.
  6. Power on the chassis.
  7. Perform the self-test again.